Table 1 Diagnostic performance of ECV and FIDDLE in canines and patients.

From: Accuracy of ECV imaging for the detection of subendocardial infarction - comparison with black blood delayed enhancement and pathology

  Sensitivity Specificity Accuracy
Overall - Animals
FIDDLE 95% (39/41) 100% (170/170) 99% (209/211)
ECV 27% (11/41) 98% (166/170) 84% (177/211)
Subendocardial (<50% transmurality by pathology)
FIDDLE 93% (28/30) 100% (170/170) 99% (198/200)
ECV 3% (1/30) 98% (166/170) 84% (167/200)
Transmural (>50% transmurality by pathology)
FIDDLE 100% (11/11) 100% (170/170) 100% (181/181)
ECV 91% (10/11) 98% (166/170) 87% (176/181)
Overall - Patients
ECV 53% (25/47) 90% (28/31) 68% (53/78)
Subendocardial (<50% transmurality by FIDDLE)
ECV 26% (5/19) 94% (28/31) 66% (33/50)
Transmural (>50% transmurality by FIDDLE)
ECV 71% (20/28) 94% (28/31) 83% (48/59)