Skip to main content

Table 1 Summary of Quantitative Assessment

From: Evaluation of left ventricular ejection fraction using through-time radial GRAPPA

 

Gold-standard

Real-time

Entire subject population (63/63)

  

EDV

140.7 mL ± 68.1 mL

143.3 mL ± 67.9 mL

ESV

79.2 mL ± 60.0 mL

79.0 mL ± 59.5 mL

EF

48.0% ± 15.6%

48.8% ± 15.0%

Radial GRAPPA—26 cal frames (23/63)

  

EDV

131.9 mL ± 72.5 mL

131.5 mL ± 68.7 mL

ESV

77.3 mL ± 62.3 mL

74.4 mL ± 58.9 mL

EF

46.6% ± 17.1%

48.3% ± 16.6%

Radial GRAPPA—6 cal frames (40/63)

  

EDV

145.8 mL ± 65.8 mL

150.1 mL ± 67.4 mL

ESV

80.4 mL ± 59.4 mL

81.6 mL ± 60.4 mL

EF

48.7% ± 14.7%

49.1% ± 14.2%

Ejection fraction > 50% (33/63)

59.3% ± 8.5%

59.6% ± 8.2%

40% < Ejection fraction < 50% (14/63)

44.5% ± 3.3%

46.0% ± 2.7%

30% < Ejection fraction < 40% (9/63)

34.9% ± 2.6%

36.8% ± 3.5%

Ejection fraction < 30% (7/63)

18.2% ± 6.9%

19.0% ± 5.4%

  1. Summary of average and standard deviations of ESV, EDV, and EF values for the study population as a whole, divided into subcategories with different numbers of calibration frames for the radial GRAPPA reconstruction, and sorted into EF ranges.