Skip to main content

Table 2 Details of the CMR protocol

From: Feasibility and safety of adenosine cardiovascular magnetic resonance in patients with MR conditional pacemaker systems at 1.5 Tesla

Objective

Sequence

Plane

TR/TE (ms)

Slice thickness (mm)

Slices

Matrix

FOV Phase (mm)

Flip angle

PAT

Anatomical orientation

HASTE

Axial, coronal, sagittal

1000/44

8

27/35

125×256/142×256

290/360

160

1

Cine imaging

True FISP

Long and short axes

66/1.6

8

LA: 3

SA: 10/12

166×256

300/370

79

1

First pass perfusion

GRE

Short axes

176/1.2

8

3/4

96×128

260/300

15

1

Late Gadolinium Enhancement

PSIR

Long and short axes

1024/3.5

8

LA: 3

SA: 10/12

144×256

270

25

0

  1. HASTE Half fourier acquisition single shot turbo spin echo, LA Long axis, SA Short axis, TR Repetition time, TE Echo time, FOV Field of view, PAT Parallel acquisition technique, True FISP, True fast imaging with steady state precession, GRE Gradient echo, PSIR Phase-sensitive inversion recovery