Fig. 3From: Comparison of different cardiovascular magnetic resonance sequences for native myocardial T1 mapping at 3TError curves for the three sequences on T1MES. Absolute error in MOLLI, SHMOLLI and SASHA T1 values compared to gold standard inversion-recovery spin-echo (IR-SE) in native tissue-like phantom samples (T1s 1000–1500 ms, T2 50–55 ms), Post contrast tissue-like phantom samples (T1s 300–600 ms, T2 60–85 ms) and blood pool like phantom (T1 1880 ms, T2 160 ms). Continuous lines represent averagesBack to article page