Table 1 Imaging protocol parameters

From: Prospective comparison of novel dark blood late gadolinium enhancement with conventional bright blood imaging for the detection of scar

 

Bright Blood LGE

DB LGE

Preparation

Inversion Preparation

Inversion Preparation & T2 preparation

Readout

Single shot, bSSFP

(FAIR = 50°, FA PD = 8°)

Typical FOV / resolution

360 × 270 mm2

1.4 × 1.9 × 8 mm3

Matrix size

256 × 144 (parallel imaging factor 2)

Number of acquired measurements

8

16

T2 preparation TE

10–40 ms

TE/TR

1.2/2.8 ms

ECG triggering

Inversions every 2 RR (HR < 90 bpm)

Inversions every 3 RR (HR > 90 bpm)

  1. DB dark blood, ECG electrocardiom, FOV field-of-view, LGE late gadolinium enhancement, TE echo time, TR repetition time