Table 2 Global rest myocardial blood flow (MBF) corrected for heart rate (HR)

From: Fully automated, inline quantification of myocardial blood flow with cardiovascular magnetic resonance: repeatability of measurements in healthy subjects

  Test 1 ml/g/min Test 2 ml/g/min Difference in mean ml/g/min p RC RC (%) CV (%) ICC
Rest        < 0.01     
Rest 1 - Rest 2 0.69 ±0.12 0.66 ±0.10 −0.03 ±0.09 0.14 0.18 27.3 9.9 0.66
Rest 1 - Rest 3 0.69 ±0.12 0.61 ±0.10 −0.07 ±0.07 < 0.01 0.20 30.3 10.5 0.62
Rest 2 - Rest 3 0.66 ±0.10 0.61 ±0.10 −0.04 ±0.06 < 0.01 0.15 23.7 8.3 0.72
  1. p – from repeated measures ANOVA and Bonferroni post-hoc analysis
  2. RC repeatability coefficient, RC (%)repeatability coefficient as percentage of the mean, CV within subject coefficient of variation, ICC intraclass correlation coefficient