Fig. 11 | Journal of Cardiovascular Magnetic Resonance

Fig. 11

From: A multi-band double-inversion radial fast spin-echo technique for T2 cardiovascular magnetic resonance mapping of the heart

Fig. 11

Bland Altman Analysis. Correlation of repeated MB-DIR-RADFSE acquisitions and corresponding Bland-Altman plot for a 5 healthy subjects and b 10 patients. The correlation plot indicates a significant correlation between the two sets of measurements for both the normal and patient groups with correlation coefficients of 0.90 and 0.968 respectively (p-value < 0.001). The dotted lines in the Bland-Altman plot represent 1.96 times the standard deviation of the differences in the estimates. The mean difference in T2 between the repetitions is 0.18 ms for the healthy subjects and 0.12 ms for the patients