Table 1 T2 estimation accuracy

From: A multi-band double-inversion radial fast spin-echo technique for T2 cardiovascular magnetic resonance mapping of the heart

Reference

T2 (ms)

Single slice DIR-RADFSE

T2 (ms)

MB-DIR-RADFSE

Slice 1

T2 (ms)

MB-DIR-RADFSE

Slice 2

T2 (ms)

T2-Prep bSSFP

T2 (ms)

37.6 ± 0.5

36.± 1.4

36.7 ± 1.3

36.4 ± 1.7

44.7 ± 1.5

51.7 ± 0.8

49.7 ± 1.9

50.5 ± 2.0

50.0 ± 1.4

57.8 ± 2.3

73.5 ± 0.9

72.6 ± 2.0

72.85 ± 2.19

72.3 ± 1.9

85.1 ± 3.0

127.1 ± 2.4

130.4 ± 4.1

130.6 ± 3.5

129.7 ± 3.8

146.1 ± 7.7

  1. Mean and standard deviation of the T2 estimates on agarose gel phantoms are shown. The reference values were obtained from a single-echo spin-echo experiment as described in the Methods section. Note that the estimates from the proposed MB-DIR-RADFSE pulse sequence are closer to those from a single slice excitation scheme while the T2-prep bSSFP over-estimates the T2 values