Table 2 Diagnostic Accuracy of Ferumoxytol Enhanced Cardiovascular Magnetic Resonance Venography (FE-CMRV)

From: High resolution, 3-dimensional Ferumoxytol-enhanced cardiovascular magnetic resonance venography in central venous occlusion

Reader No. of TP Findings No. of TN Findings No. of FP Findings No. of FN Findings Sensitivity Specificity PPV NPV Accuracy
Consensusa 25 63 2 2 92.6 (75.7–99.1) 96.9 (89.3–99.6) 92.6 (89.2–99.2) 96.9 (89.2–99.2) 95.7 (89.2–98.8)
Ultimate assessmentb 27 65 0 0 100.0 (87.2–100.0) 100.0 (94.5–100.0) 100.0 100.0 100.0 (96.1–100.0)
  1. aConsensus read was determined by consensus of the primary reviewers with a third senior reviewer in cases where there was disagremment between the reviewers A and B for the FE-CMRV
  2. bUltimate assessment was determined by the senior reviewer (J.P.F.) for cases with disagreement between the consensus FE-CMRV and reference standard
  3. Note – Data in parenthesis are 95% CIs. FN = false-negative, FP = false-positive, NPV = negative predictive value, PPV = positive predictive value, TN = true-negative, TP = true-positive