Skip to main content

Table 3 Inter-reader agreement for quantitative measurements and qualitative image quality on DANTE-SPACE

From: Surveillance of abdominal aortic aneurysm using accelerated 3D non-contrast black-blood cardiovascular magnetic resonance with compressed sensing (CS-DANTE-SPACE)

 

DANTE-SPACE

Reader 1

DANTE-SPACE

Reader 2

P value

ICC

Bias (LOA)

CV

Maximal Diameter (cm)

4.85 ± 0.87

4.84 ± 0.87

0.92

0.985

0.01(−0.30,0.30)

3.2

ILT/wall Area (cm2)

9.55 ± 5.72

9.44 ± 5.78

0.44

0.983

0.11(−1.90, 2.12)

10.8

Lumen Area (cm2)

7.62 ± 4.46

7.53 ± 4.52

0.31

0.990

0.09(−1.05, 1.22)

7.6

Contrast Ratio

2.50 ± 0.91

2.44 ± 0.94

0.18

0.944

0.06(−0.54,0.66)

12.4

ILT signal ratio

1.05 ± 0.29

0.98 ± 0.26

< 0.001

0.884

0.06(−0.16,0.30)

11.6

Image quality score

3.22 ± 0.58

2.84 ± 0.65

< 0.001

0.577

0.38(−0.73,1.49)

18.7