Skip to main content

Table 4 Inter-reader agreement for quantitative measurements and qualitative image quality on CS-DANTE-SPACE

From: Surveillance of abdominal aortic aneurysm using accelerated 3D non-contrast black-blood cardiovascular magnetic resonance with compressed sensing (CS-DANTE-SPACE)

 

CS-SPACE

Reader 1

CS-SPACE

Reader 2

P value

ICC

Bias (LOA)

CV

Maximal Diameter (cm)

4.84 ± 0.88

4.83 ± 0.92

0.76

0.985

0.01(−0.30,0.31)

3.2

ILT/wall Area (cm2)

9.62 ± 5.81

9.39 ± 5.72

0.06

0.989

0.23(−1.38,1.83)

8.6

Lumen Area (cm2)

7.64 ± 4.41

7.72 ± 4.49

0.31

0.993

−0.08(−1.13,0.97)

7.0

Contrast Ratio

2.95 ± 0.98

2.89 ± 0.93

0.50

0.944

0.06(−1.16,1.28)

21.4

ILT signal ratio

1.03 ± 0.32

0.95 ± 0.29

0.01

0.815

0.08(−0.28,0.42)

17.8

Image quality score

3.24 ± 0.62

3.06 ± 0.71

0.05

0.559

0.18(−1.05,1.41)

20.0