Skip to main content

Table 1 bSSFP cine imaging sequence parameters

From: A comparison of cine CMR imaging at 0.55 T and 1.5 T

 

0.55 T breath-held bSSFP cine

1.5 T breath-held bSSFP cine

0.55 T free breathing re-binned bSSFP cine

1.5 T free breathing re-binned bSSFP cine

Field of view (mm2)

360 × 270

360 × 270

360 × 270

360 × 270

Slice thickness (mm)

8

8

8

8

Matrix size

256 × 192

256 × 140

192 × 108

192 × 119

TE (ms)

1.67

1.2

1.34

1.06

TR (ms)

4.1

2.79

3.24

2.52

Acquired temporal resolution (ms)

32

28

N/A

N/A

Bandwidth (Hz/Px)

350

1085

501

1085

Parallel imaging acceleration factor

2

2

3

4

Seconds/slice

9

8

18

16

Calculated Phases

30

30

26

30

Flip angle (°)

78

50

80

50

  1. Sequence parameters for breath-held and free breathing re-binned cine acquisitions at 0.55 T and 1.5 T; bSSFP balanced steady statae free precession, TE echo time, TR repetition time