Skip to main content

Table 1 Technical parameters of LGE sequences

From: Myocardial late gadolinium enhancement using delayed 3D IR-FLASH in the pediatric population: feasibility and diagnostic performance compared to single-shot PSIR-bSSFP

CMR parameter

Conventional LGE

3D HR-LGE

Base sequence

Single-shot PSIR-bSSFP

3D IR-FLASH

ECG gating

Mid-late diastole

Cardiac quiescence in systole or diastole

Parallel imaging × acceleration factor

GRAPPA × 2

GRAPPA × 2

Field of view

250–350 mm

250–350 mm

Pixel size

1.8 × 1.8 mm

1.1 × 1.1 to 1.4 × 1.4 mm

Slice thickness

6 mm (no gap)

1.1 to 1.4 mm

Flip angle

40˚

18˚

TR/TE

2.57/1.09 ms

330/1.4 ms

Number of averages

1–3a

1

Fat saturation

Yes

Yes

Acquisition plane(s)

Axial, short axis

Coronal

  1. ECG electrocardiogram; FLASH fast low angle shot; GRAPPA generalized autocalibrating partial parallel acquisition; IR inversion recovery; PSIR phase-sensitive inversion recovery; bSSFP balanced steady-state free precession; TE time to echo; TR repetition time
  2. aSmaller patients and those performed under general anesthesia were performed free breathing with more averages to reduce motion artifact and increase signal