From: 1140 A new method for measuring through-plane strain using slice following with inversion recovery (SFIR)

Figure 2

Slice following pulse sequence diagram. Two acquisitions are required. First, a thin slice is inverted with a slice selective 180 inversion pulse, while a thicker slice, thick enough to enclose the initially inverted slice, is selected for imaging. Then the second acquisition is done to image the thicker slice without any inversion. The two acquisitions are then subtracted to yield an image of the tussue of inverted slice.