141 Volumetric cardiac quantification using three dimensional dual phase whole heart MRI
Journal of Cardiovascular Magnetic Resonance volume 10, Article number: A42 (2008)
Introduction
Cardiac ventricular volumes are usually assessed using Simpson's rule over short axis (SA) cines of the heart acquired during breath-holds. This technique is challenging for ill patients and segmentation is difficult due to non isotropic resolution. Recently, a new approach was proposed that calculates the ventricle volumes from two non-angulated isotropic cardiac scans of the whole heart (end-systolic, ES, and end-diastolic, ED) (1). However, this approach results in a relatively long overall scan time. To overcome this limitation we propose acquiring the two volumes in one free-breathing scan using quasi independent navigator beams for each cardiac phase. The aim of this study is to evaluate this new acquisition scheme and compare the EDV, ESV and stroke volumes (SV) obtained using this technique with the traditional multiple 2D (M2D) scans and with flow measurements.
Methods
3D dual phase acquisition scheme
A 3D triggering b-SSFP turbo field echo sequence was modified to enable the acquisition of two cardiac phases at a user defined time (Figure 1). The sequence was implemented on a 1.5 T Philips clinical scanner. A free breathing scan was realized by enabling one navigator beam before data acquisition for each cardiac phase. They were used to prospectively validate or invalidate acquired k-space data for each cardiac phase quasi independently.
Experiments
A prospective study was performed in ten patients and five healthy volunteers. EDV, ESV and SV were obtained and compared from the following imaging acquisitions, M2D SA cine, free breathing flow scans in the aorta and pulmonary artery and the new dual phase scan (imaging parameters are summarized in Table 1). The M2D and flow data were analyzed using commercially available software (Philips View forum). Analysis of the 3D data sets were performed using a semiautomatic segmentation [1, 2]. Bland Altman analyses and T-test analyses were used to assess agreement between the measurements.
Results
3D dual cardiac phase scan
Reformatted slices in one volunteer are shown in Figure 2. The time of the 3D scan was 7:54 ± 1:42 [min:sec], with an scan-efficiency of 47% ± 7% due to respiratory gating.
Statistical analysis
No statistical difference was found for the measured ESV and EDV comparing the 2D and 3D technique (see Bland-Altman plot in Figure 3). Bland-Altman plots of SV comparison of the 3D technique with flow measurements and with the M2D method are shown in Figure 4. It is noticeable that equivalent results were obtained for both ventricles for all approaches. Range and mean data from intra and inter observer variability using the 3D and the M2D method are shown in Table 2.
Conclusion
We have introduced an acquisition scheme that allows precise cardiac volume quantification in a single free breathing scan. The new 3D dual phase technique offers a series of advantages over the traditional M2D approach such as minimal scan planning, isotropic resolution and a good definition of the cardiac valves. Severely ill, sedated patients and those with congenital heart diseases may benefit from the proposed method, since it is a patient friendly scan and provides both morphological and cardiac volume information.
References
Greil GF, et al: J Magn Reson Imaging. 2007
Bottger T, et al: Acad Radiol. 2007
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Arancibia, S.A.U., Tangchaoren, T., Parish, V. et al. 141 Volumetric cardiac quantification using three dimensional dual phase whole heart MRI. J Cardiovasc Magn Reson 10 (Suppl 1), A42 (2008). https://doi.org/10.1186/1532-429X-10-S1-A42
Published:
DOI: https://doi.org/10.1186/1532-429X-10-S1-A42
Keywords
- Congenital Heart Disease
- Dual Phase
- Acquisition Scheme
- Free Breathing
- Cardiac Phase