Skip to main content
  • Meeting abstract
  • Open access
  • Published:

50 Myocardial signal behaviors of balanced SSFP imaging at 3 T

Introduction

Balanced steady state free precession (bSSFP) imaging has an established role in the CMR assessment of ventricular function and wall motion [1]. However, measured bSSFP signals do not always agree with simple theoretical predictions. More accurate signal models would enable optimization of imaging parameters (e.g. prescribed flip angle and TR) in order to maximize signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). In this work, we examine the effects of (A) non-ideal slice profile, (B) off-resonance, and (C) radiofrequency transmit (B1+) variation, to identify a model that accurately predicts myocardial signal behavior in bSSFP, over the range of imaging parameters routinely used at 3 T.

Materials and methods

Experiments

Experiments were performed on two 3 T GE EXCITE HD systems in five healthy volunteers. Cardiac bSSFP cine loops were acquired using a product 2D FIESTA sequence with prescribed flip-angles from 10° to 90° with 5° increments. The TR and RF pulse duration were fixed, and the imaging flip angle was adjusted by manually changing the RF pulse amplitude. Imaging parameters: FOV = 30 cm, matrix = 224 × 224, slice = 5 mm, 10 cardiac phases, 20 views per segment, TR = 5.0 ms (4 subjects) or 5.9 ms (1 subject due to SAR constraints). RF pulse: duration = 1.28 ms, time-bandwidth product = 2. Off-resonance maps were acquired in a single breath-hold using a cardiac-gated sequence with two different echo times (TEs). The TE difference was set to 2 ms resulting in a frequency range of ± 250 Hz. B1+ maps were acquired in a single breath-hold using the cardiac-gated saturated double angle method [2]. Images from mid-diastole were selected for analysis, and regions of interest containing septal myocardium were manually selected.

Signal models

The steady-state signal from static tissue in bSSFP can be expressed as:

SNR sinα/((T1/T2+1)-cosα·(T1/T2-1)).

Three additional factors were considered: (A) the slice profile α(z) is not ideal, (B) the spins include a range of resonance offsets, and (C) the actual flip angle is not the prescribed flip angle. There are seven possible models that either consider or ignore each factor. All combinations and the conventional method (ignoring all three) were evaluated using a relative error metric defined as:

RelativeError = E(|SI measured -SI expected |/|SI measured |).

Simulations assumed myocardial relaxation times of T1 = 1115 ms and T2 = 41 ms [3]. Bloch simulation and image analysis were performed in MATLAB. It is worth noting that magnetization transfer effects [4] were not considered in this study.

Results

Figure 1 contains data from one representative volunteer including a comparison of normalized myocardial signal (measured) and three signal models (conventional, A+B, and A+B+C) as a function of the prescribed flip angle. The case A+B+C shows an excellent agreement with the experimentally measured values. Figure 2 contains the mean and standard deviations of the relative error averaged over all five subjects. The relative error decreases when considering more practical factors while the case A and A+C show similar performance compared to the case A+B+C.

Figure 1
figure 1

The appearance of myocardium in balanced SSFP images is influenced by imperfect slice profiles, off-resonance, and flip angle variation. We demonstrate accurate myocardial signal prediction using models that consider these three practical factors. (a-d) Cardiac cine images with different prescribed flip angles. (e) The normalized myocardial signals (measured, conventional way, A+B, A+B+C) as a function of the prescribed flip angle.

Figure 2
figure 2

Relative error (mean and standard deviation) with different signal models.

Conclusion

Accurate myocardial signal prediction can facilitate optimization of SNR and CNR in bSSFP, by guiding the selection of the imaging flip angle and RF pulse. The relative error between measured and theoretically predicted myocardial signal was reduced by 85% by considering three practical factors: non-ideal slice profile, off-resonance, and B1+ inhomogeneity.

References

  1. Schar M ,et al.: MRM. 2004, 51: 799-806.

    Article  PubMed  Google Scholar 

  2. Sung K, Nayak KS , et al.: ISMRM. 2007, 355-

    Google Scholar 

  3. Noeske R , et al.: MRM. 2000, 44: 978-982.

    Article  CAS  PubMed  Google Scholar 

  4. Bieri O, Scheffler K: MRM. 2006, 56: 1067-1074.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Sung, K., Nayak, K.S. 50 Myocardial signal behaviors of balanced SSFP imaging at 3 T. J Cardiovasc Magn Reson 10 (Suppl 1), A51 (2008). https://doi.org/10.1186/1532-429X-10-S1-A51

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-10-S1-A51

Keywords