- Oral presentation
- Open Access
- Published:
Ultrafast in-line computation of ejection fraction from cardiac cine steady-state free precession (SSFP) images
Journal of Cardiovascular Magnetic Resonance volume 12, Article number: O47 (2010)
Introduction
In clinical practice estimation of global LV function from cine-SSFP images involves at least two workstations, i.e., one to acquire (scanner), and another to process. These workstations are often present at distant locations, and therefore the physicians and technologists at the scanner console often do not get a real-time, quantitative feedback regarding patients LV function. A fast automated in-line processing tool can eliminate this gap, and integrate the process of data acquisition and post-processing on one console.
Purpose
The purpose of this abstract is to describe our computationally inexpensive, in-line post-processing algorithm that can calculate EF on the scanner console immediately upon the completion of data acquisition and to investigate its clinical feasibility.
Methods
The EF computation program automatically runs in the background as a part of the acquisition protocol on the console of a 1.5 T Philips Achieva clinical scanner (Intel Xeon, 3.20 GHz, 6 GB RAM). The program seamlessly retrieves, processs, and saves resultant contoured LV images back into the patient database.
LV Segmentation
The algorithm steps: 1) automatically recognize LV; 2) classify partial volumed pixels as blood or muscle using periodic intensity variation intrinsic to the cardiac cine-MR; and 3) incorporate smoothness of the LV shape through convex, closed, and piecewise parametric curve to delineate endocardial boundary (details in [2]).
Study Population
12 (11 m/1 f, age 52 ± 17 yrs) clinical patients participated in this study. A stack of contiguous short-axis cine SSFP slices were acquired.
Results
total of 221 slices were analyzed using our algorithm (A) and manually by a clinical expert (R). The only user input to the processing algorithm was to specify the basal and apical slice number. The processing time was 80-120 ms/slice. The manual contours for slice with LV outflow tract were used for accurate volumetric and LVEF measurement. Some representative images describing the segmentation process and BA comparisons of EDV, ESV, and EF are shown in Fig. 1.
The mean bias computed using Bland-Altman (BA) analysis between expert manual contours and automated contours for EF (-6%), EDV (-15 ml, -8%), and ESV (5 ml, 5%) were comparable to that of the typical inter- and intra-observer variability (<5% EDV, <8% ESV, <5% EF) [1].
Conclusion
It is feasible to integrate the computation of global LV functional indices such as EDV, ESV, and EF as a part of acquisition in real-time on the scanner console.
References
J Magn Reson Imaging. 2008, 28: 39-50. 10.1002/jmri.21363.
ISMRM. 09: 3749-
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Pednekar, A., Muthupillai, R., Arena, C. et al. Ultrafast in-line computation of ejection fraction from cardiac cine steady-state free precession (SSFP) images. J Cardiovasc Magn Reson 12 (Suppl 1), O47 (2010). https://doi.org/10.1186/1532-429X-12-S1-O47
Published:
DOI: https://doi.org/10.1186/1532-429X-12-S1-O47
Keywords
- Manual Contour
- Apical Slice
- Quantitative Feedback
- Cine SSFP
- Cardiac Cine