- Poster presentation
- Open Access
- Published:
Centric reordered echo planar imaging (EPI) for phase contrast MRI
Journal of Cardiovascular Magnetic Resonance volume 12, Article number: P94 (2010)
Introduction
Limited temporal resolution underestimates peak velocities assessed by phase contrast MRI (PC-MRI) [1]. Segmented EPI with partial Fourier as previously proposed [2] has relatively long TE, and is prone to chemical-shift artifact. We propose centric reordered water-excitation EPI for PC-MRI with short TE and fat suppression. It improves temporal resolution by twofold while halving scan time compared with turboFLASH.
Purpose
Design a centric reordered EPI sequence for PC-MRI, and compare it with turboFLASH in healthy volunteers.
Methods
Sequences
The prospective triggered EPI sequence used centric reordering [3], fast, flow-compensated water excitation [4] and TSENSE [5]. Concomitant gradient effects were corrected [6]. The sequence was implemented on a 1.5 T scanner (MAGNETOM Avanto, Siemens Healthcare, Germany) and its performance was compared to turboFLASH (imaging parameters in Table 1). The TR/TE/flip angle were 12.4 ms/2.4 ms (effective)/15o-25ofor EPI, and 5 ms/1.9 ms/25ofor turboFLASH. EPI echo spacing = 690 μs. In both cases, venc = 150-200 cm/s, pixel ~2.3 × 2 mm2, slice = 6-8 mm.
Imaging/Analysis
For each (n = 8) healthy volunteers in this IRB approved study, an imaging plane perpendicular to the ascending aorta was prescribed. Aortic flow was imaged twice using each of the two sequences. Argus (Siemens Healthcare, Germany) was used to find the peak velocity (PV) and flow volume (FV). PVs from EPI were obtained using 5-pixel spatial averaging to account for bandwidth related SNR differences between EPI and turboFLASH. PV and FV from each volunteer were averaged for one sided and two sided t-tests respectively
Results
All EPI images had good image quality. EPI related distortion or fat-water shift artifact was not observed (Fig. 1). The FV between the two techniques differ by 0.5% (p = 0.66). PVs from EPI were higher than those from turboFLASH by 7% (p < 0.01,), due most likely to improved temporal resolution (e.g., Fig. 2).
Conclusion
Centric reordered EPI provided short TE and reduced sensitivity to turbulence and minimized T2* effects. Fat suppression improved image quality. The volunteer study results suggested that the high temporal resolution of EPI improves peak velocity sampling, and the shortened scan time makes the technique clinically attractive.
References
Lotz J, et al: Radiographics. 22: 651-
McKinnon GC: MRM. 32: 263-
Ding S, et al: MRM. 39: 514-
Lin HY, et al: JCMR. 10: 22-
Kellman P, et al: MRM. 45: 846-
Bernstein MA, et al: MRM. 39: 300-
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Chung, YC., Ding, Y. & Simonetti, O. Centric reordered echo planar imaging (EPI) for phase contrast MRI. J Cardiovasc Magn Reson 12 (Suppl 1), P94 (2010). https://doi.org/10.1186/1532-429X-12-S1-P94
Published:
DOI: https://doi.org/10.1186/1532-429X-12-S1-P94
Keywords
- Flow Volume
- Peak Velocity
- Temporal Resolution
- Siemens Healthcare
- Echo Planar Imaging