Virtual Dye Angiography: flow visualization for MRI-guided interventions using endogenous contrast

Objective

To efficiently visualize blood flow in real-time without exogenous contrast agents.

Background

Visualizing blood flow is important in MR-guided interventions (e.g. repairing structural heart defects). We introduce a method, called Virtual Dye Angiography (VDA), to visualize flow in a manner inspired by contrast-enhanced X-ray angiography. Slightly similar to arterial spin labeling, VDA saturates a localized region using multidimensional RF pulses. Unlike phase-contrast velocity mapping, which has unsuitably long acquisition times, VDA can be integrated into existing high-contrast, high-SNR SSFP imaging sequences with minimal modification.

Methods

Porcine experiments were performed on a 1.5T scanner (Siemens;Espree) using a real-time multi-slice SSFP sequence that was modified to include VDA. When VDA was enabled (in real-time) a cylindrical column (whose size and location were interactively controlled) was continuously saturated. The VDA module consisted of a shaped RF pulse played concurrently with a spiral gradient[1] (to flip spins in the cylindrical column into the transverse magnetization plane) followed by a spoiler gradient in the slice direction (to saturate the column). Two slices (1 and 2) were imaged in real-time with the VDA saturation module (13 ms) played before the acquisition of, and with the saturation column perpendicular to, slice 1.

Results

VDA was demonstrated in multiple locations in the pig heart. Figure 1 shows flow from the LV to the aorta; without VDA (top row) and with VDA (bottom row). The dashed white line in (e) denotes the relative position of slice 2 to slice 1. The saturation column is placed through the LV. The white arrows track the saturated spins: in initial saturation in the LV (d) and into the aorta (e and f). Figure 2 shows flow from the RV to the pulmonary vessels using a saturation column placed through the RV. The white arrows track the saturated spins: in initial saturation in the RV (d), then exiting the RV outflow tract (e), and in the pulmonary arteries (f). Figure 3 shows the corresponding subtraction images.

Figure 1
figure 1

LV to Ao flow

Figure 2
figure 2

RV to MPA flow

Figure 3
figure 3

Subtraction images corresponding to Figure 2

Conclusions

We have introduced a real-time method of visualizing blood flow. Its modularity allows for minimal modification of the existing sequence. Future refinements we envision include cardiac synchronization and motion compensation for consistent placement of the saturation column and the use of DSA-like image subtraction for enhanced visualization.

References

  1. Börnert : MAGMA. 1998, 7 (3): 166-178. 10.1007/BF02591334.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

George, A.K., Faranesh, A.Z., Ratnayaka, K. et al. Virtual Dye Angiography: flow visualization for MRI-guided interventions using endogenous contrast. J Cardiovasc Magn Reson 13 (Suppl 1), O25 (2011). https://doi.org/10.1186/1532-429X-13-S1-O25

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-13-S1-O25

Keywords