A protocol for the measurement of myocardial blood volume and water exchange
Journal of Cardiovascular Magnetic Resonance volume 13, Article number: P140 (2011)
Objective
Absolute myocardial perfusion MR imaging (ml/min/100g tissue) has the potential to timely diagnose and reduce patient mortality from coronary artery disease.
Background
Organ perfusion can be quantified by direct calibration of relative perfusion images using absolute blood volume (ml/100g) [1]. It is well known that for an intravascular gadolinium-based T1 shortening contrast agent, the parenchymal T1 change reflects tissue blood volume [2].However, to accurately quantify blood flow from blood volume, we must describe the compartmentalization effects of intra- to extra-vascular water exchange [2, 3].
Materials and methods
Protocol
In an instrumented dog we measured T1 using a cardiac gated Modified Look Locker Inversion Recovery (MOLLI)(4) pulse sequence (slice thickness 8 mm, FOV 171 x 343 mm2, matrix 96 x192, TR 173 ms, effective TI 100 ms ). Images were acquired on a 1.5 T Espree scanner (Siemens Medical Systems, Erlangen, Germany), during a short breath-hold, 5 minutes after injections of 0.003 mmol/kg of MS-325 (Ablavar, Lantheus Medical Imaging, Billerica, MA).
Image processing
We estimated the myocardium and left ventricle blood pool T1 through fitting of MOLLI signal to the regrowth curves of the Look-Locker equation by an automatic image processing program developed in MATLAB R2009a (Mathworks, Natick, MA, USA).Myocardial blood volume (MBV) was calculated from the baseline to post-contrast change in T1 in the blood pool and myocardium.
Results
Low dose injections (1/10th of single dose for humans) of MS-325 effected significant changes in myocardial T1’s (Table 1). The measured MBV was 40% of total myocardial volume, or 28 ml/100g, a value that over-estimates those quoted in the literature [5]. Water exchange in the myocardium was shown to approach the slow or no-exchange limit (Figure 1).
Conclusions
We have established an imaging protocol to measure MBV and water exchange. Over-estimation of MBV may be caused by extravasation of MS-325, and to a lesser extent by T2 bias on the T1 measurements with the steady-state free precession MOLLI sequence. Future steps include measuring MBV with a strictly intravascular USPIO contrast agent, application of a more sophisticated fit that includes T2 effects, and determination of the water exchange constant by Monte-Carlo simulations.
References
Carroll TJ, Horowitz S, et al: Magn Reson Imaging. 2008, 26 (10): 1352-9. 10.1016/j.mri.2008.04.010.
Donahue KM, Weisskoff RM, et al: Magn Reson Med. 1996, 36 (6): 858-67. 10.1002/mrm.1910360608.
Hazlewood CF, Chang DC, et al: Biophys J. 1974, 14 (8): 583-606. 10.1016/S0006-3495(74)85937-0.
Messroghli DR, Radjenovic A, et al: Magn Reson Med. 2004, 52 (1): 141-6. 10.1002/mrm.20110.
Bjornerud A, Bjerner T, et al: Magn Reson Med. 2003, 49 (5): 828-37. 10.1002/mrm.10430.
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Biris, O., Chatterjee, N. & Carr, J. A protocol for the measurement of myocardial blood volume and water exchange. J Cardiovasc Magn Reson 13 (Suppl 1), P140 (2011). https://doi.org/10.1186/1532-429X-13-S1-P140
Published:
DOI: https://doi.org/10.1186/1532-429X-13-S1-P140
Keywords
- Water Exchange
- Compartmentalization Effect
- Myocardial Blood Volume
- Automatic Image Processing
- Quantify Blood Flow