Whole heart coronary MRA at 3.0T: a comparison between multi and single- RF transmission

Introduction

On high field MRI scanners uniform radio frequency (RF) excitation over the entire field-of-view is often challenging with single-channel RF transmit coils. This may cause a reduction of image quality (e.g. shadowing artifacts). This problem is most pronounced in sequences which heavily rely on a homogenous magnetic field, such as steady state free precession (SSFP) sequences, despite having been shown to yield best results for MR coronary angiography at 1.5T. Multi-channel RF transmission (Tx) has been shown to significantly improve the RF (B1 field) uniformity in high-field MRI. More accurate knowledge of the local B1 field also allows to better predicting the local specific absorption rate (SAR), thereby allowing for shorter repetition times (TR). Altogether, this may allow for the use of SSFP sequences at 3T and thus to combine the high CNR benefits of 3D SSFP with the high SNR advantages of 3T for whole heart coronary MR angiography (WHCMRA).

Purpose

The purpose of this study was to evaluate if the use of a multi-channel RF transmit coil can improve image quality of WHCMRA at 3.0T compared to a conventional single-channel RF transmit coil.

Methods

Ten healthy volunteers were scanned (Achieva, 3T, 32-channel coil, Philips Healthcare, Best Netherlands). After preparatory scans for localization of the heart, B1 calibration and identification of the coronary resting period, 3D transverse WHCMRA images were acquired using a fast segmented gradient echo (TFE) and a SSFP sequence with T2 preparation and fat saturation (SENSE factor = 2; full Fourier encoding; acquisition duration per cardiac cycle <120ms; navigator gating window = 5mm; resolution = 1.5x1.5x1.5mm, k-space encoding = Cartesian). The following 3 settings were compared: TFE, SSFP, SSFP+TX (TR/TE/flip = 4.0/1.16ms/20o, 5.0/2.5ms/70o, 4.5/2.3ms/70o, respectively). The acquisition order was randomized in each subject. Objective values such as SNR, CNR and vessel sharpness were determined for the right and left coronary artery systems. Two blinded, expert reviewers assessed a subjective quality score.

Results

Mean actual scan time was between 9-11 minutes. Objective and subjective image quality for TFE and SSFP+Tx was comparable and substantially higher than for SSFP (Table 1, 2; Representative image, Figure 1). Banding artifacts were observed in all SSFP sequences.

Table 1 Objective evaluation of WHCMRA image quality.
Table 2 Subjective evaluation of WHCMRA image quality
Figure 1
figure 1

Right coronary scan with TFE (A), SSFP (B) and SSFP+Tx (C)

Conclusion

SSFP WHCMRA benefits from multi-transmit RF transmission. This may be explained by the added benefit from an improved SAR model (shorter TR) and improved B1 homogeneity. However, banding artifacts still affect SSFP WHCMRA image quality when compared to the conventional TFE sequence.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Ishida, M., Chiribiri, A., Hussain, S. et al. Whole heart coronary MRA at 3.0T: a comparison between multi and single- RF transmission. J Cardiovasc Magn Reson 13 (Suppl 1), P233 (2011). https://doi.org/10.1186/1532-429X-13-S1-P233

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-13-S1-P233

Keywords