Improved real-time blood flow velocity quantification via application of the Karhunen-Loeve transform for increased signal-to-noise ratio
Journal of Cardiovascular Magnetic Resonance volume 13, Article number: P38 (2011)
Objective
Evaluate the application of temporal filtering via the Karhunen-Loeve Transform for increasing signal-to-noise ratio in real-time blood flow quantification.
Background
Real-time blood flow velocity quantification requires fast acquisition time and both high spatial and temporal resolution in order to capture clinically relevant features. Multi-coil parallel imaging techniques trade fast real-time cine acquisition with high temporal resolution for lower signal quality. Methods for increasing signal-to-noise ratio (SNR) will allow higher acceleration rates in parallel imaging, leading to faster acquisition times in the context of exercise stress while maintaining signal quality.
Increased SNR via the Karhunen-Loeve Transform [1] (KLT) is achieved by decomposing a temporal cine sequence into a series of eigenmodes where the signal of interest is localized in a subset of eigenmodes and eliminating noise-only eigenmodes.
Methods
KLT was applied retrospectively to ten through-plane real-time velocity data sets of the aorta at rest acquired at 1.5T (Avanto, Siemens, Malvern, PA). Optimal identification of noise-only eigenmodes was determined using the Marcenko-Pastur Law [2], and SNR gain was calculated as the square root of the ratio of noise-only eigenmodes to total number of eigenmodes.
Results
Optimal selection of SNR gain via the Marcenko-Pastur Law results in SNR increase of 61.5 ± 3.7% (mean ± standard deviation) with no statistically significant difference in peak velocity measurement (Difference: 1.45 ± 2.98%, p-value: 0.2794). Figure 1 illustrates the minimal difference in peak velocity measurement due to KLT filtering.
Conclusions
Temporal filtering of real-time velocity images via KLT results in significant gains in SNR with minimal velocity profile smoothing and minimal effect on peak velocity measurement and may be useful in improved blood flow velocity quantification.
References
Phys. Med. Biol. 2009, 54: 3909-10.1088/0031-9155/54/12/020.
Mag. Res. Med. 2010, 63: 782-789. 10.1002/mrm.22258.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Ting, S., Ding, Y. & Simonetti, O.P. Improved real-time blood flow velocity quantification via application of the Karhunen-Loeve transform for increased signal-to-noise ratio. J Cardiovasc Magn Reson 13 (Suppl 1), P38 (2011). https://doi.org/10.1186/1532-429X-13-S1-P38
Published:
DOI: https://doi.org/10.1186/1532-429X-13-S1-P38
Keywords
- Parallel Imaging
- Improve Blood Flow
- Parallel Imaging Technique
- Cine Sequence
- Blood Flow Quantification