Integrated quantitative first-pass cardiac perfusion MRI protocol

Introduction

The proposed contrast-enhanced first-pass perfusion cardiovascular MR (CMR) protocol integrates some recent technical MRI advances towards quantitative analysis of perfusion CMR images: fast multi-slice pulse sequence [1], robust saturation pulse [2], accurate dedicated AIF imaging [3], signal-to-concentration modeling [4], and higher SNR at 3T.

Purpose

To evaluate an integrated first-pass perfusion cardiovascular MR (CMR) protocol designed to determine absolute contrast-agent concentrations in blood and tissues.

Methods

A multi-slice saturation recovery (SR) pulse sequence with sequential SR time delays (TD) after a non-selective saturation pulse [2] was implemented at 3T (Fig.1). The rationale for this acquisition scheme was to acquire a dedicated arterial input function (AIF) image with a short TD (50ms) in the aortic root and short-axis myocardial images with longer TD values (~150-400ms), to allow for the different amounts of T1 shortening expected in blood and wall. First-pass perfusion CMR was performed in 7 volunteers (0.05mmol/kg, Gd-DTPA). A signal-to-concentration model was applied to calculate Gd-DTPA concentrations in blood and tissues [4, 5]. A proton density-weighted (PDw) image was acquired in the first heartbeat, without the saturation pulse, in order to normalize the image signal, and obtain a theoretical signal-to-T1 relationship based on Bloch equation in the center of k-space. Gd-DTPA concentrations were calculated assuming: fast water exchange condition [6], longitudinal relaxivity r1=3.8L.mmol-1.s-1[7], and baseline T1 measured with a multi-point SR fit. TurboFLASH imaging parameters included: FOV=350mm×315mm, slice thickness=8mm, matrix=160×144, in-plane resolution=2.2mm×2.2mm, TE/TR=1.2/2.4ms, flip angle 10°, temporal resolution=114ms, tSENSEx3, centric k-space trajectory, and receiver bandwidth=1008Hz/pix. Total image acquisition time was 523ms for the acquisition of 4 slices, namely aortic root and SA base, mid, and apex levels, with respective TD values 50-164-278-393ms. Contours for the blood and left ventricle were drawn manually, and the myocardium was divided into 6 (base-mid) or 4 (apex) standard segments.

Figure 1
figure 1

Schematic diagram of the mulit-slice with sequential TD first-pass perfusion CMR pulse sequence

Results

Representative images at peak contrast in blood and myocardium are shown Fig.2, along with representative AIF and myocardial segment time-responses. The peak blood signal was not clipped in the short TD=50ms AIF images. Normalized signal in the myocardium increased along with TD; however similar [Gd-DTPA] were measured in all 3 short-axis images. First-pass perfusion peak Gd-DTPA concentrations were 3.95±0.080, 0.26±0.07mM in the blood and myocardium, respectively.

Figure 2
figure 2

Representative images at peak blood and peak myocardium concentrations.

Figure 3
figure 3

Representative [Gd-DTPA] times-curves in blood in the aortic root and in 6 cardiac segments in cardiac short-axis images.

Conclusions

The proposed integrated first-pass perfusion CMR protocol at 3T produced AIF and myocardial wall Gd-DTPA concentrations consistent with previously published results. Future work includes evaluation of the integrated protocol in cardiac patients.

References

  1. Nagel E: Circulation. 2003, 108: 432-437. 10.1161/01.CIR.0000080915.35024.A9.

    Article  PubMed  Google Scholar 

  2. Kim D: Magn Reson Med. 2008, 62: 1368-1378. 10.1002/mrm.22140.

    Article  Google Scholar 

  3. Kim D, Axel LJ: Magn Reson Imaging. 2006, 23: 81-86. 10.1002/jmri.20471.

    Article  Google Scholar 

  4. Cernicau A, Axel L: Acad Radiol. 2006, 13: 686-693. 10.1016/j.acra.2006.02.040.

    Article  Google Scholar 

  5. Breton E: SCMR. 2010

    Google Scholar 

  6. Dobahue KM: J Magn Reson Imaging. 1997, 7: 102-110. 10.1002/jmri.1880070114.

    Article  Google Scholar 

  7. Rohrer M: invest Radiol. 2005, 40: 715-724. 10.1097/01.rli.0000184756.66360.d3.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Breton, E., Kim, D., Chung, S. et al. Integrated quantitative first-pass cardiac perfusion MRI protocol. J Cardiovasc Magn Reson 13 (Suppl 1), P60 (2011). https://doi.org/10.1186/1532-429X-13-S1-P60

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-13-S1-P60

Keywords