Can real-time cmr provide adequate morphologic or functional data? A real-time CMR/SSFP comparison trial
Journal of Cardiovascular Magnetic Resonance volume 13, Article number: P7 (2011)
Introduction
The objective of this trial was to assess the accuracy of cardiac magnetic resonance (CMR) images obtained via real-time CMR (RT-CMR) (GE: MR Echo) imaging sequences in comparison to steady-state free precession (SSFP).
Purpose
Diagnostically useful cardiac MR images are currently obtained using SSFP imaging sequences. Longer acquisition times and breath holding make this sequence unsuitable for use in certain patient populations (eg. - children and sick adults). Unfortunately, RT-CMR images have lower spatial and temporal resolution than those obtained using SSFP sequences and are not currently used clinically.
Methods
Forty nine consecutive patients (table 1) underwent both SSFP and RT-CMR images. Clinically relevant cardiac parameters (table 2) were measured in both diastole and systole utilizing SSFP and RT-CMR sequences. Measurements were made by considering the largest ventricular size to be diastole and the smallest to be systole. Measurements were made using common, consistent anatomic locations in both methods, were performed in random order, and were blinded to the results from the other sequence. Cardiac dimensions measured in each imaging modality were then compared using a paired t-test as well as a Passing and Bablock regression analysis.
Results
With the exception of LVEDA, our data show no statistical difference between the measurements obtained using either CMR sequence (figures 1 and 2). The slope and intercept for a Passing and Bablock regression are shown below and demonstrate no significant deviation from linearity (ALL P VALUES N >0.10). RT-CMR exhibits a consistent, though not statistically significant underestimation of LV areas in both systole and diastole. Furthermore, analysis of the relationship between the percentage difference of these methods showed this difference to be highly, significantly correlated.
Conclusions
RT-CMR provides clinically useful morphologic and functional LV and RV data compared to SSFP imaging. In this highly variable clinical population of pediatric and adults patients referred for CMR studies, there was no significant difference between SSFP and RT-CMR measurements with the exception of LV diastolic area.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Unger, W.D., Diller, M., Dewar, J.D. et al. Can real-time cmr provide adequate morphologic or functional data? A real-time CMR/SSFP comparison trial. J Cardiovasc Magn Reson 13 (Suppl 1), P7 (2011). https://doi.org/10.1186/1532-429X-13-S1-P7
Published:
DOI: https://doi.org/10.1186/1532-429X-13-S1-P7