Skip to main content
  • Poster presentation
  • Open access
  • Published:

A new method to suppress ghosting artifacts arising from long-T1 species in segmented inversion recovery (IR) sequences

Objective

To evaluate a new long-T1 artifact suppression method and compare its effectiveness with previously reported techniques.

Background

Bright ghosting due to long-T1 species (e.g. cerebrospinal fluid, pericardial or pleural effusion, fig.1) often hampers the identification of infarcted myocardium on delayed enhancement CMR. The artifacts are caused by signal oscillations of the long-T1 species due to the periodic application of IR pulses. Previous methods use a dummy heartbeat (+HB, sequence run without data acquisition for leading heartbeat) or various pre-pulses (SR or IR). Both techniques may result in insufficient suppression or may be cumbersome to use. We developed a new suppression method utilizing saturation pulses (followed by “crusher” gradient pulses) after each readout (+Po) in combination with a dummy heartbeat. This method does not require user input. We tested it against four other suppression methods by quantifying the long-T1 ghosting seen in the respective images.

Figure 1
figure 1

Example of long T1 ghosting artifact in a patient with a left pleural effusion.

Methods

A phantom containing five compartments with different T1-values (figure 2) was imaged (MAGNETOM Avanto, Siemens) with IR GRE and SSFP sequences with and without PSIR at heart rates (HR) 60 and 100 bpm. No suppression (NO) and five suppression methods were applied: 1) +HB, 2) -HB+Pre, 3) +HB+Pre, 4) -HB+Po, 5) +HB+Po. The inversion time (TI) was set to null the background at each HR. The images were quantified in ImageJ (NIH) as relative artifact signal by dividing the mean signal in the ghosting region of interest (ROI1) by the mean signal of the short T1 (brightest, control) compartment (ROI2, fig.2b). Additionally, artifacts were visually scored by three blinded readers, ordering them by quality (1 worst, 6 best). Signal-to-noise ratio (SNR) was measured in ROI2.

Figure 2
figure 2

a) Example of long T1 ghosting artifacts using a phantom with different T1. b) Phantom diagram and annotations. c) Artifact suppression using our best suppression method.

Results

Table demonstrates a significantly lower relative artifact signal of method 5 compared to all other suppression techniques (p<0.05), for GRE and SSFP with and without PSIR, and independent of HR. Qualitative analysis the techniques from best to worst as 5, 3, 1, 2, 4, NO, where again method 5 performed best. SNR in control region ROI2 for method 5 was not significantly different from that of method 2 (p<0.05).

Conclusions

The new post-suppression method (5) consistently showed the best ghosting suppression by qualitative and quantitative tests in both GRE and SSFP sequences, while preserving SNR. Due to its improved artifact suppression, simple implementation and heart rate independence, this method is currently used clinically at our site for delayed enhancement imaging.

Table 1 Artifact to maximal signal ratios (x100) for suppression method imaging type and heart rate. Underlined numbers have the least artifact to signal ratio

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Chaptini, N., Rehwald, W.G., Jenista, E.R. et al. A new method to suppress ghosting artifacts arising from long-T1 species in segmented inversion recovery (IR) sequences. J Cardiovasc Magn Reson 13 (Suppl 1), P9 (2011). https://doi.org/10.1186/1532-429X-13-S1-P9

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-13-S1-P9

Keywords