Robust and fast SSFP for the evaluation of LV function at 3T
Journal of Cardiovascular Magnetic Resonance volume 15, Article number: O52 (2013)
Background
3T MRI is advantageous to CMR as it offers higher SNR and better T1 contrast. Its slow clinical adoption is partly due to the lack of robustness in SSFP cine imaging [1]. Banding artifact in SSFP caused by main field inhomogeneity is common at 3T. Wideband SSFP [2] or prescans (shimming, frequency scout, etc.) are needed to avoid/reduce the artifact, increasing patient scan time. We propose here to use shorter TR for SSFP through slightly reduced spatial resolution to improve the sequence's robustness and speed for cine imaging. Its ability for LV function assessment was evaluated against the standard SSFP acquisition protocol.
Methods
Protocol
A 600 µs (instead of 1000 µs) RF pulse and 192 (instead of 256) readout point in SSFP shortened the TR/TE from 3.4/1.5 ms to 2.7/1.2 ms. At 40 ms temporal resolution and 85% phase resolution, the segments/heart beat increased from 12 to 15, reducing patient's breath-hold time.
Imaging
The IRB approved study scanned 8 healthy volunteers (age 26±2) with a 3T MR scanner (TIM TRIO, Siemens, Germany). Cines of 10 short-axis slices covering the whole heart were acquired with retrogated breath-held SSFP (8mm thick, 2mm gap, FOV=340×287mm^2, iPAT=2, maximum bandwidth). The short TR sequence was used first. Frequency scout was then performed. After that, cine acquisition for the whole heart was repeated using the standard protocol. LV myocardial mass, end-diastolic volume (EDV), end-systole volume (ESV) and ejection fraction (EF) were analyzed by an experienced observer using QMass MR (Medis, Netherlands) (papillary muscles included). Correlation analysis was performed between paired measurements. Two-tailed paired t-test was conducted to assess the statistical differences with p<0.05 regarded as significant.
Results
The scan time for 1 slice of cine was about 6 and 9 heart beats for the protocols with short and standard TR respectively. Banding artifact was common in cine images obtained with the standard TR protocol, while images from the protocol using short TR scan were impervious to the artifact. Analysis results were shown in Table 1. Myocardial mass, EDV and ESV from the 2 protocols were very close. The short TR protocol yielded a 1% reduction in EF compared to the standard one.
Conclusions
The short TR SSFP cine is more robust to field inhomogeneity at 3T. It eliminates frequency scout and increases patient throughput. Initial results showed that short TR cines gave comparable myocardial mass, EDV and ESV. The measured EF was 1% lower than the standard protocol, and may be caused by the reduced image spatial resolution. Our preliminary study suggested that by slightly trading spatial resolution for shorter TR in SSFP, cine imaging at 3T can be more robust and efficient. The approach may greatly facilitate patient throughput in 3T CMR. Evaluation on more volunteers and patients will be done to fully validate the method.
References
Oshinski JN, et al: JCMR. 2010, 12: 55-
Lee H, et al: MRM. 2010, 63 (6):
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Wu, Y., Yang, F. & Chung, YC. Robust and fast SSFP for the evaluation of LV function at 3T. J Cardiovasc Magn Reson 15 (Suppl 1), O52 (2013). https://doi.org/10.1186/1532-429X-15-S1-O52
Published:
DOI: https://doi.org/10.1186/1532-429X-15-S1-O52
Keywords
- Field Inhomogeneity
- Cine Imaging
- Myocardial Mass
- Image Spatial Resolution
- Patient Throughput