Whole-heart first-pass myocardial perfusion imaging with non-rigid respiratory motion correction
Journal of Cardiovascular Magnetic Resonance volume 15, Article number: P176 (2013)
Background
Assessment of myocardial perfusion plays an important role in the evaluation of coronary artery and ischemic heart disease [1, 2]. Scan acceleration methods such as k-t PCA [3] have proven invaluable for increasing coverage and have facilitated 3D perfusion imaging [4]. However, the sensitivity to respiratory motion for high acceleration factors remains challenging [4]. Especially in clinical routine, long breathhold durations are not always feasible.
In this work, an iterative k-t PCA algorithm is proposed which corrects for nonrigid frame-to-frame motion based on regularization with motion corrected k-t training data and motion operators thereof. It is demonstrated that this approach permits artifact-free reconstruction of 3D perfusion data acquired in non-compliant patients.
Methods
Three motion corrupted 3D data sets were taken from a study of patients with suspected or known coronary artery disease. Data was acquired with a 10-fold k-t undersampled, saturation recovery gradient-echo sequence on a 3T Philips Ingenia system (Philips Healthcare, Best, The Netherlands). Scan parameters were: TR/TE: 2.2/1 ms, flip angle = 15°, spatial resolution: 2.3x2.3x10 mm3, saturation prepulse delay: 150 ms, acquisition window: 210 ms, #dynamics: 30.
Two out of the three data sets presented here were obtained under pharmacological stress (adenosine: 140 μg/kg/min over 6 min). For reconstruction, an iterative k-t PCA algorithm was implemented with all dynamic frames being warped to a reference respiratory position by an additional spatial transformation before applying Fourier and PCA transform in the time domain. The deformation fields for warping were estimated by image registration of frame-by-frame SENSE reconstructed images. To avoid misregistration due to contrast uptake, a b-spline transformation model was constrained to affine motion for several predefined image regions, including heart, liver, chest wall, and back. These regions were defined by an atlas based on a segmented multi-slice survey. Motion corrected images were compared to standard k-t PCA images without motion correction.
Results
Still frame images and profiles over time are shown for standard and motion corrected reconstruction in Figure 1. Image artifacts were successfully suppressed when using the motion-corrected k-t PCA implementation. In addition the approach allowed warping of all dynamic frames to a single respiratory position hence greatly simplifying image segmentation and analysis. Signal intensity profiles are shown for myocardium and left ventricular blood pool in Figure 2 demonstrating reduction of respiratory motion induced signal fluctuations with motion correction.
Conclusions
Iterative k-t PCA with nonrigid motion correction enables correction of respiratory motion artifacts in dynamic 3D myocardial perfusion imaging. Thereby 3D perfusion data obtained during interrupted breathholding or even during free breathing can be successfully reconstructed and analyzed.
Funding
The authors acknowledge funding by the Swiss National Science Foundation, grant #CR3213_132671/1 and by Bayer (Switzerland) AG.
References
Schwitter J, et al: . European heart journal. 2011, 32 (7): 799-U162. 10.1093/eurheartj/ehq481.
Greenwood JP, et al: . Lancet. 2012, 379 (9814): 453-60. 10.1016/S0140-6736(11)61335-4.
Pedersen H, et al: . MRM. 2009, 62 (3): 706-16.
Vitanis V, et al: . MRM. 2011, 65 (2): 575-587.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Schmidt, J., Wissmann, L., Manka, R. et al. Whole-heart first-pass myocardial perfusion imaging with non-rigid respiratory motion correction. J Cardiovasc Magn Reson 15 (Suppl 1), P176 (2013). https://doi.org/10.1186/1532-429X-15-S1-P176
Published:
DOI: https://doi.org/10.1186/1532-429X-15-S1-P176
Keywords
- Myocardial Perfusion Imaging
- Respiratory Motion
- Dynamic Frame
- Respiratory Motion Artifact
- High Acceleration Factor