Skip to main content
  • Poster presentation
  • Open access
  • Published:

Artifact suppression of SSFP cine sequences at 3T using a novel automatic 3D shimming algorithm

Background

Imaging the heart at 3T has some advantages, but magnetic susceptibility increases at higher field strengths, generating deleterious banding artifacts, especially when using SSFP-based sequence. We aimed to evaluate whether a recently developed advanced shimming algorithm reduces banding artifacts on SSFP cine images at 3T.

Methods

A total of 20 consecutive patients scheduled to undergo a clinically indicated CMR in normal sinus rhythm and able to perform apnea were prospectively enrolled to be scanned in a 3T system (MAGNETOM Verio, Siemens, Germany). Frequency scout images were acquired in two and four chamber views and the frequency offset associated with least banding artifacts involving the heart was manually set (Figure 1) for the acquisition of short and long axis segmented cine SSFP images covering the entire ventricular volumes (bandwidth of 700-750 Hz/pixel and echo spacing of 3.4 ms). The same sequences were then repeated with the offset reset to 0 Hz (only change made) but with the use of a novel 3D shimming algorithm restricted to the heart (instead of the entire FOV) based on a patient specific GRE fieldmap previously generated (WIP - Siemens, Germany).

Figure 1
figure 1

Frequency scout images of the four chamber view. Banding artifacts are seen at frequency offset of 0 Hz (a), but not at 150 Hz (b).

Images were blindly evaluated for segmental function (normal vs. abnormal), and the presence of artifacts within the heart (none, minor or important - the latter meaning that at least one myocardial segment could not be assessed). Ventricular volumes were also blindly measured according to guidelines. Categorical differences were assessed with chi-square, kappa was calculated for agreement and continuous variables were assessed using Student's T test.

Results

Mean LV and RV volumes and ejection fraction (EF) were not statistically different between acquisition modes, and had small biases between measurements (Table 1). Segmental wall motion analysis had good agreement with kappa of 0.76, p<0.001. Minor artifacts were seen in at least one segment in 4 (20%) patients using the new shimming and 7 (35%) using frequency offsets. Important artifacts were seen in at least one segment in 3 (15%) patients using the new shimming and 2 (10%) using frequency offsets.

Table 1 Mean LV and RV volumes and ejection fraction.

Conclusions

Cine SSFP at 3T with a new shimming algorithm targeted to the heart is similar to the standard practice of selecting the best frequency offset. This new shimming might simplify cardiac acquisitions at 3T.

Funding

Internal

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Rothstein, T., Camargo, G.C., Quintella, D.C. et al. Artifact suppression of SSFP cine sequences at 3T using a novel automatic 3D shimming algorithm. J Cardiovasc Magn Reson 15 (Suppl 1), P57 (2013). https://doi.org/10.1186/1532-429X-15-S1-P57

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-15-S1-P57

Keywords