- Poster presentation
- Open Access
- Published:
Detection of intramural scar by electroanatomic mapping versus MRI in patients with non-ischemic cardiomyopathy
Journal of Cardiovascular Magnetic Resonance volume 15, Article number: P65 (2013)
Background
Ventricular arrhythmias have been described to originate from intramural locations. Intramural scar can be assessed by delayed enhanced magnetic resonance imaging (DE-MRI ). The objective of this study was to determine whether endocardial voltage mapping by catheter can detect the presence and extent of scar deep within the myocardial wall thickness, using DE-MRI as gold standard.
Methods
In 15 consecutive patients with structural heart disease and without contraindication to MRI, intramural scar was detected by DE-MRI (Fig 1a). All patients underwent endocardial voltage mapping by catheter. Both unipolar and bipolar voltage maps were constructed from the endocardium. The scar on DE-MRI was registered with the electroanatomic map using fiducials (apex, aorta, mitral valve) and a polar representation of the corresponding endocardial voltage was generated (Fig 1b: red points: low voltage, purple points: high voltage, yellow and green points: fiducials).

Figure 1
Results
On MRI, scar volume was 11.7±8.0 cm3, endocardial scar area was 19.3±13.4 cm2. Scar thickness was 4.6±0.7 mm (39±7% of wall thickness), with a subendocardial rim thickness of 3.3±1.6 mm (26±13%) and a subepicardial rim thickness of 4.8±2.6 mm (37±15%). Endocardial bipolar voltage measurements were 1.62±1.73 mV over the scar, 2.12±2.15 mV in a 1cm neighborhood around the scar, and 2.83±2.34 mV in the more remote myocardium (p<0.0001). Endocardial unipolar voltage measurements were 4.94±3.25 mV over the scar, 6.59±3.81 mV in a 1cm neighborhood around the scar, and 8.32±3.39 mV in the more remote myocardium (p<0.0001). Using ROC curves (Fig 1c,d) of the voltage maps, a unipolar cut-off value of 6.78 mV (AUC 0.78) and a bipolar cut-off value of 1.55 mV (AUC 0.69) best separated endocardial measurements overlying scar from the endocardial measurements more remote to the scar.
Conclusions
Although scar deep within the myocardial wall is far from the tip of the endocardial measurement catheter, its presence and extent can still be detected by endocardial electroanatomic mapping using a combination of unipolar and bipolar voltage, with unipolar voltage being more useful.
Funding
NIH K23 EB006481
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Desjardins, B., Morady, F. & Bogun, F. Detection of intramural scar by electroanatomic mapping versus MRI in patients with non-ischemic cardiomyopathy. J Cardiovasc Magn Reson 15 (Suppl 1), P65 (2013). https://doi.org/10.1186/1532-429X-15-S1-P65
Published:
DOI: https://doi.org/10.1186/1532-429X-15-S1-P65
Keywords
- Myocardial Wall
- Remote Myocardium
- Electroanatomic Mapping
- Bipolar Voltage
- Myocardial Wall Thickness