Background
X-ray magnetic resonance fusion (XMRF) is used to enhance fluoroscopically guided catheterization procedures. We present a new method for biplane XMRF involving an augmented fluoroscopy prototype (Siemens, Forchheim, Germany). With this software it is possible to register 3D MRI data to biplane X-ray projections based on internal markers, without the need for a C-arm CT to perform 3D/3D registration. The software supports overlaying volume rendered data as well as multiple surface models. Visualization techniques, such as contour or solid rendering and surface carving are supported to allow for clear presentation of complex 3D structures.
Methods
We reviewed data obtained with different visualization methods on 20 patients that underwent clinical XMRF procedures. Surface models were generated by threshold based segmentation from high resolution MRA (syngo Twist or Navigator gated 3D flash IR sequence) of structures of interest using Mimics (Leuven, Belgium). Initial registration was achieved through planar alignment of the volume in anterior-posterior (AP) and lateral projections matching anatomical landmarks, such as the heart and vessel borders. In addition, the registration accuracy of the prototype was assessed using a phantom by measuring the maximum distance between a single point and the 3D model boundaries in comparison to the boundaries seen on fluoroscopy.
Results
The maximum segmentation time was 10 min and the initial registration required less than 30 sec. Registration was performed without the need for contrast injection or additional radiation exposure. The registration error based on phantom measurements was 2.2 ± 1.1 mm in the AP projection and 1.36 ± 0.7 mm in the lateral projection. For surface rendered data, solid rendering with carving provided the optimal display of complex 3D data. In contrast to volume rendering, solid surface rendering provides delineation of the 3D relation between objects and clear visualization of internal structures such as ostia of the vessels and muscle bundles using the carving feature, see Figure 1. Contour rendering offered reasonable visualization for smooth, uncomplicated structures like the atria, but it was not ideal for complex overlapping structures due to missing 3D depth information. When multiple surfaces are loaded, contour rendering was, however, found to be useful in combination with solid rendered structures to provide a see-through 3D relation between multiple structures, see Figure 2.
Conclusions
The prototype demonstrated a high level of accuracy for fluoroscopic overlays. Biplane internal marker based registration can be performed quickly, without the need for additional radiation or contrast. We found solid rendering of surface models in combination with carving techniques to be most useful for visualization. This new biplane XMRF technique has the potential to provide enhanced guidance under fluoroscopy by integrating 3D information and to reduce radiation during complicated catheterization procedures.
Funding
This project is funded by a research grant from Siemens, AG, Healthcare, Forchheim, Germany. The concepts and information presented in this paper are based on research and are not commercially available.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Kurzendorfer, T., Girard, E., Gralewski, K. et al. New biplane x-ray magnetic resonance image fusion prototype for 3D enhanced cardiac catheterization in congenital heart diseases. J Cardiovasc Magn Reson 16 (Suppl 1), O103 (2014). https://doi.org/10.1186/1532-429X-16-S1-O103
Published:
DOI: https://doi.org/10.1186/1532-429X-16-S1-O103