Comparison of white blood and black blood T2* cardiovascular magnetic resonance at 3 Tesla
Journal of Cardiovascular Magnetic Resonance volume 16, Article number: O86 (2014)
Background
T2* relaxometry offers a non-invasive method to accurately quantify myocardial iron levels, thus aiding the diagnosis and assessment of prognosis of patients. Two T2* sequences are used clinically: the single breath-hold multi-echo gradient echo white blood (WB) and black blood (BB) techniques. At 1.5T, BB T2* has superior reproducibility and fewer artefacts than WB T2*. We compared BB and WB T2* iron measurement at 3T and compared results with corresponding 1.5T measurements.
Methods
We recruited 100 patients (55 male, aged 14 to 81 years) referred for iron assessment (thalassemia major 43, sickle cell disease 15, hereditary haemochromatosis 11, other iron overload conditions 31). The same mid-ventricular short axis slices were used to generate both BB and WB T2* images at 1.5T (Avanto, Siemens, Erlangen, Germany), and similarly at 3T (Skyra). 20 patients underwent repeat studies on the same day for reproducibility. Septal regions of interest were analyzed using CMRtools, using the truncation algorithm. All patient images were ranked for artefact severity on an ordinal scale from 0 (uninterpretable images) to 5 (no septal artefact).
Results
Median BB T2* was 29.8 ms (range 3.12-50.5 ms) at 1.5T, and 21.4 ms (range 1.70-35.5 ms) at 3T. Median WB T2* values were 30.8 ms (range 3.76-49.2 ms) at 1.5T and 19.0 ms (range 2.82-34.7 ms) at 3T. Intra-observer, inter-observer and inter-study reproducibility measurements at 3T of BB T2* yielded coefficients of variation (CoV) of 3.1%, 4.0% and 6.5% respectively, which were superior to WB T2* whose CoV were 3.5%, 5.5% and 7.7% respectively (Figure 1). Mean artefact score was 4.11 for BB T2* at 3T, which was significantly better than WB T2* at 3T (3.45; p < 0.001). Applying the truncation algorithm to both 1.5T and 3T analyses, BB R2* (1000/T2*) correlated more strongly to 1.5T values (slope 1.91, intercept -15.6, Rsquared = 0.987; Figure 2) than WB R2* (slope 1.61, intercept +2.96, Rsquared = 0.947).
Conclusions
At 3T, BB T2* is superior to WB T2* for artefacts, noise and reproducibility, which accords with experience at 1.5T. The 3T BB T2* values correlate well with gold standard 1.5T T2* with values approximately halved. Considerable further experience with BB T2* is needed before robust clinical practice can be recommended
Funding
This research was supported by the NIHR Cardiovascular Biomedical Research Unit at Royal Brompton & Harefield NHS Foundation Trust and Imperial College London.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Alam, M.H., Smith, G.C., Carpenter, J.P. et al. Comparison of white blood and black blood T2* cardiovascular magnetic resonance at 3 Tesla. J Cardiovasc Magn Reson 16 (Suppl 1), O86 (2014). https://doi.org/10.1186/1532-429X-16-S1-O86
Published:
DOI: https://doi.org/10.1186/1532-429X-16-S1-O86
Keywords
- Cardiovascular Magnetic Resonance
- Thalassemia
- Black Blood
- Myocardial Iron
- Hereditary Haemochromatosis