Comparing 3DQRS and VCG approaches for ECG QRS detection within 1.5T, 3T and 7T MRI
Journal of Cardiovascular Magnetic Resonance volume 16, Article number: P148 (2014)
Background
Electrocardiograms (ECGs) obtained within High-field MRIs are distorted due to the Magneto-hydrodynamic (MHD) effect. Blood plasma electrolytes ejected into the aorta during early systole interact with the strong magnetic field of the MR scanner to produce a MHD-induced voltage (VMHD) [1]. The VMHD overlay on ECG traces can result in intermittent QRS detection. Vectorcardiogram (VCG) based gating approaches have been conventionally adapted in most MRI scanners [2], but may fail at high field strengths [3]. Recently, a multiple ECG channel cross-correlation based algorithm, 3DQRS, has been developed to provide increased sensitivity levels in these environments [4]. The 3DQRS approach constructs a 3-D ECG representation, where the third dimension, in addition to voltage and time, is deemed a channels axis, formed from concurrent viewing of the precordial leads V1-V6. This study quantitatively compares 3DQRS and VCG approaches at a variety of MRI field strengths to assess the robustness of these methods.
Methods
12-lead ECG data was recorded using a prototype MRI-conditional 12-lead ECG recorder [5] from 2 Premature Ventricular Contraction (PVC) patients, 2 Atrial Fibrillation (AF) patients, and a healthy exercising athlete at 1.5T and 3T [4]. A Halter recorder was used in 2 healthy volunteers at 7T [6] (Figure 1a-f). QRS detection was performed using a VCG-based approach (V1-V6, I, II) [2] and 3DQRS (V1-V6) using standard 12-lead ECG chest positioning [4]. Assessments of 3DQRS robustness relative to variations in field strength and cross-correlation kernel temporal length were performed (Figure 1g-h). False Positive (FP) and False Negative (FN) counts were recorded (total of 1,262 beats) in order to assess the sensitivity for QRS detection for each method at 1.5T, 3T, and 7T.
Results
Table 1 shows the gating results for 3DQRS and VCG at 1.5T, 3T, and 7T in various subjects. 3DQRS subject-averaged accuracy levels in QRS detection (False Negative), relative to VCG, were: 1.5T (100% vs. 96.6%), 3T (98.1% vs. 87%), 7T (96% vs. 71.2%). In PVC patients at 1.5T, 3DQRS separated between the SR and PVC beats with 100% accuracy, whereas VCG falsely detected PVC beats, which were of similar length and magnitude to the sinus rhythm beats, with only 37.3% accuracy (Table 1).
Conclusions
The 3DQRS method represented a higher sensitivity in QRS detection than the VCG based approach, which resulted in decreased error levels in high field MRI.
Funding
NIH U41-RR019703, NIH R03 EB013873-01A1, SBIR-1 R43 HL110427-01.
References
Gupta : IEEE Trans Bio Med Eng. 2008
Fischer : MRM. 1999
Krug : JCMR. 2013
Tse : JCMR. 2013
Tse : MRM. 2013
Krug : CompInCardiol. 2012
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Gregory, T.S., Schmidt, E.J., Zhang, S.H. et al. Comparing 3DQRS and VCG approaches for ECG QRS detection within 1.5T, 3T and 7T MRI. J Cardiovasc Magn Reson 16 (Suppl 1), P148 (2014). https://doi.org/10.1186/1532-429X-16-S1-P148
Published:
DOI: https://doi.org/10.1186/1532-429X-16-S1-P148
Keywords
- Premature Ventricular Contraction
- Rhythm Beat
- Premature Ventricular Contraction Beat
- Concurrent Viewing
- Decrease Error Level