Coronary mra angiography at 3T: fat suppression versus water-fat separation
Journal of Cardiovascular Magnetic Resonance volume 16, Article number: P175 (2014)
Background
Suppression of lipid signal is a basic requirement in coronary magnetic resonance angiography (CMRA) because coronary arteries are embedded in epicardial fat and signal from fat can decrease coronary vessel conspicuity. Most CMRA scans are currently performed with fat suppression techniques such as Spectral Presaturation with Inversion Recovery (SPIR). However, methods based on spectrally-selective fat saturation are sensitive to B0 and B1 field inhomogeneities. Recent improvements in chemical shift based water fat separation methods such as Dixon[1, 2]provides an alternative to conventional spectrally-selective fat suppression techniques. The purpose of this study was to compare SPIR technique and Dixon water fat separation at 3.0 T for CMRA.
Methods
This work was performed on a 3T scanner (Achieva, Philips Healthcare, Best, The Netherland) equipped with a 32-element cardiac receiver coil. Data were acquired in eight healthy volunteers (six male, two female, mean age 36 ± 11) and eight patients with suspected coronary artery disease (five male, three female, mean age 60 ± 12). Two different scans were performed in each of them: 1) Conventional whole heart CMRA with SPIR fat suppression and 2) two-point Dixon CMRA. Two experts readers, blinded to the methods used, scored the image quality for each dataset. In addition, signal-to-noise ratio of blood, fat and myocardium, contrast-to-noise ratio (CNR) between blood, fat and myocardium, and right coronary artery sharpness and length were measured to compare these two techniques quantitatively. A Wilcoxon Signed-Rank was used for statistical analysis for comparison between images acquired with Dixon and SPIR.
Results
All scans were successfully performed and produced good quality images in all volunteers. Figure 1 shows representative CMRA images from two healthy volunteers and one patient acquired with SPIR and Dixon. In both volunteers fat suppression with Dixon was visually superior to the SPIR technique. Images acquired with the two-point Dixon method were scored higher than the ones acquired with SPIR in volunteers (4.4 ± 0.7 vs. 3.6 ± 0.6, p = 0.009) and in patients (2.35 ± 0.9 vs. 1.8 ± 1.2, p = 0.04). Vessel sharpness of the right coronary artery with Dixon acquisition (57 ± 0.1) was similar to SPIR (56 ± 0.1). Figure 2 demonstrates that the Dixon method leads to similar fat suppression but increased SNR and CNR compared to SPIR.
Conclusions
These findings demonstrate that Dixon water-fat separation leads to higher SNR of coronary blood and myocardium and improved image quality scores for coronary artery visualization at high field strengths. Furthermore, the additional fat data that is available with Dixon protocols may be an important biomarker and improve the diagnostic value of CMRA.
Funding
Wellcome Trust and ESPRC Medical Engineering Centre.
References
Eggers H, Brendel B, Duijndam A, Herigault G: Dual-echo Dixon imaging with flexible choice of echo times. Magn Reson Med. 2011, 65 (1): 96-107. 10.1002/mrm.22578.
Börnert P, Koken P, Nehrke K, Eggers H, Ostendorf P: Water/fat-resolved whole-heart Dixon coronary MRA: An initial comparison. Magn Reson Med. 2013
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Nezafat, M., Henningsson, M., Ripley, D.P. et al. Coronary mra angiography at 3T: fat suppression versus water-fat separation. J Cardiovasc Magn Reson 16 (Suppl 1), P175 (2014). https://doi.org/10.1186/1532-429X-16-S1-P175
Published:
DOI: https://doi.org/10.1186/1532-429X-16-S1-P175
Keywords
- Image Quality Score
- Coronary Magnetic Resonance Angiography
- Dixon Method
- Vessel Sharpness
- Spectral Presaturation With Inversion Recovery