An efficient, free-breathing protocol for MR right heart catheterization
Journal of Cardiovascular Magnetic Resonance volume 16, Article number: T5 (2014)
Background
MR right heart catheterization is a valuable diagnostic tool in patients with cardiac and pulmonary vascular disease. Many patients are dyspneic and/or require moderate sedation so MR examinations with multiple breath holds are not feasible. We describe an efficient free-breathing MR imaging protocol designed to facilitate MR catheterization in these challenging patients.
Methods
The baseline MR examination begins with cardiac localizers. At this stage it is important to verify isocenter because if the patient is too deep in the bore then femoral vascular access sheaths are inaccessible to the operator. Through-plane phase contrast (PC) scans in the ascending aorta and main pulmonary artery follow for quantification of Qp and Qs. To avoid the need to repeat long PC scans (3 averages), we first perform brief PC 'localizer' scans with a single average to confirm the selected velocity-encoding factor is correct. For cardiac volumetric analysis, long and short axis views are obtained using a real-time SSFP sequence that sacrifices some spatial resolution for increased temporal resolution. Data is collected over one and a half heartbeats to ensure the entire cardiac cycle is captured. Because of the delay in image reconstruction, the short axis stack is acquired last while the operator prepares for catheterization. A dedicated user interface (Interactive Front End, Siemens; Erlangen, Germany) is used for real-time imaging during catheterization. This interface permits interactive manipulation of imaging planes, slice thickness, saturation pre-pulse, and cine loop recordings. To minimize catheterization time, reference images consisting of orthogonal views of the superior and inferior vena cavae, the right ventricular outflow tract, and the main and branch pulmonary arteries are prescribed prior to commencing catheterization. If additional physiological provocations during catheterization are indicated, then the baseline imaging sequences are repeated during each provocation via automatic propagation of the sequence parameters.
Results
Acquisition order was optimized to shorten scan times by enhancing scanner duty cycle. With practice we reduced the baseline MR examination time from 50 minutes to 30 minutes. With parameter duplication, subsequent scan time during physiological provocations was reduced to 8 minutes per provocation. We are now able to perform baseline MR examination, right heart catheterization and multiple physiological provocations.
Conclusions
This free-breathing MR examination protocol enables MR catheterization with multiple physiological provocations to be performed in patients with dyspnea and/or under moderate conscious sedation.
Funding
This work is supported by the Division of Intramural Research (Z01-HL005062-08, Z01-HL006039-01), National Heart Lung and Blood Institute, National Institutes of Health, USA.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Mazal, J.R., Rogers, T., Faranesh, A.Z. et al. An efficient, free-breathing protocol for MR right heart catheterization. J Cardiovasc Magn Reson 16 (Suppl 1), T5 (2014). https://doi.org/10.1186/1532-429X-16-S1-T5
Published:
DOI: https://doi.org/10.1186/1532-429X-16-S1-T5