Accelerated cine DENSE MRI using compressed sensing and parallel imaging

Background

Cine DENSE (Displacement Encoding with Stimulated Echoes) provides accurate and high-resolution displacement and strain imaging of the heart; however, image acquisition times are relatively long and, due to properties inherent to stimulated echoes, signal-to-noise ratio (SNR) is relatively low. Accelerated cine DENSE could substantially shorten scan times and/or provide improved spatial and/or temporal resolution; however, the low SNR, requirement to preserve phase information, and cardiac motion present technical challenges. We aimed to develop acceleration methods that overcome these challenges.

Methods

A variable-density spiral cine DENSE sequence was implemented on a 1.5T scanner (Avanto, Siemens) where the center of k-space was fully sampled and the outer portion of k-space was undersampled. Spiral interleaves were distributed uniformly within each cardiac phase and rotated by the golden angle through different cardiac phases. Cine DENSE datasets were collected from 5 healthy volunteers (age 25-28) using a 5-channel RF coil. To make comparisons, fully-sampled cine DENSE datasets with 6 to 8 spiral interleaves per image were acquired within one breathhold (20 to 26 heartbeats), and these datasets were retrospectively rate-2 undersampled. Also, prospectively rate-2 undersampled data with 4 spiral interleaves per image were acquired within a shorter breathhold (14 heartbeats). Images had spatial resolution of 1.8-2.2 × 1.8-2.2 × 8 mm and temporal resolution of 19.6 ms. All undersampled data were reconstructed using a recently-developed compressed sensing (CS) method called Block LOw-rank Sparsity with Motion-guidance (BLOSM), combined with SENSE parallel imaging. The BLOSM-SENSE algorithm exploits matrix low-rank sparsity within motion-tracked regions of SENSE-combined images. Complex-valued images were reconstructed to preserve the phase information used for displacement encoding.

Results

In Figure 1, example end-systole magnitude- (A-C) and phase-reconstructed (D-F) fully-sampled (A,D), retrospectively-undersampled (B,E), and prospectively-accelerated (C,F) DENSE images and strain maps (G-I) are shown, demonstrating excellent performance of the BLOSM-SENSE reconstruction method. Figure 2 shows that quantitation of myocardial displacement by retrospectively rate-2 accelerated cine DENSE is accurate compared to fully-sampled data. Using prospectively-accelerated cine DENSE, circumferential strain curves from all 5 volunteers demonstrated typical values and patterns for healthy subjects (C).

Figure 1
figure 1

Example end-systolic DENSE images (A-F) and strain maps (G-I). Magnitude- (A-C) and phase-reconstructed (D-F) images from fully-sampled (A,D), retrospectively-undersampled (B,E), and prospectively-accelerated (C,F) studies are shown. The retrospectively-undersampled results (B,E,H) closely resembled the fully-sampled references (A,D,G). The prospectively-accelerated results (C,F,I) presented high quality images and strain maps from data acquired within one breath-hold using BLOSM-SENSE.

Figure 2
figure 2

Quantitation of myocardial displacement (A,B) and circumferential strain (C). Example scatter plots of myocardial displacements in × (A) and y (B) directions from one volunteer. Linear regression showed good correlations of retrospectively rate-2 undersampled and fully-sampled DENSE data. The averaged correlation for all 5 volunteers is v=(1.01 ± 0.03)*u+(0.03 ± 0.009) with R2=0.993 ± 0.005. Circumferential strain (Ecc) from all 5 volunteers (C) showed typical values and patterns for healthy subjects.

Conclusions

Rate-2 accelerated spiral cine DENSE with the BLOSM-SENSE reconstruction method provides high-quality complex images. Retrospective undersampling of fully-sampled datasets demonstrated accurate displacement and strain measurements. Prospectively undersampled cine DENSE demonstrated typical myocardial strain measurements, using approximately half the data acquisition time of conventional cine DENSE.

Funding

R01 EB 001763, R01 HL 115225, AHA Award 12PRE12040059 and Siemens Healthcare.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Cite this article

Chen, X., Yang, Y., Salerno, M. et al. Accelerated cine DENSE MRI using compressed sensing and parallel imaging. J Cardiovasc Magn Reson 16 (Suppl 1), W16 (2014). https://doi.org/10.1186/1532-429X-16-S1-W16

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-16-S1-W16

Keywords