MAPS; acute safety data of the St Jude accent - tendril IPG system during prolonged max power CMR scanning
Journal of Cardiovascular Magnetic Resonance volume 17, Article number: M6 (2015)
Background
Until recently, the use of MRI in patients with PPM's was prohibited. The lifetime probability of a patient with a cardiac device requiring an MRI is 50-75%. Serious adverse events occurring during MRI of patients with cardiac devices are rare, but include asystole, VF and death. There is a clinical need to develop knowledge of MR safe devices and safe scanning protocols.
This study specifically address these issues in the SJM Accent MR PPM System, undergoing a dedicated CMR scan at 1.5 T, with a long scan duration at max power.
Methods
Patients were recruited into the MAPS trial and implanted with a SJM Accent ppm and 2 tendril MR leads. All patients were PPM dependent. CMR was performed more than 6 weeks following implant. Pacing capture thresholds, impedances and battery voltages were assessed prior to, between and immediately following the CMR scan. The scans were performed on a Siemens Avanto 1.5T scanner. All patients were placed in an MR pacing mode. Each scan duration was over 90 mins.
Results
Between February 2012 and February 2014, 50 patients were implanted with the SJM MR ppm. The mean age of the patients was 67.3±8.1 years, 30 male. All 50 patients had at least 1 CMR.
There were no significant adverse events reported during any of the scans and no loss of capture was seen in any scan.
Pacing thresholds
The mean pacing threshold for RVOT lead at implant was 0.67±0.22V and at 2 week check was 0.73±0.21V. Pacing thresholds prior to the 1st CMR scan, between the lead switch over and following the scan were 0.66±0.16V, 0.66±0.16V and 0.69±0.27V respectively, p=0.34.
The mean pacing threshold for the apical lead at implant was 0.71±0.29V and at 2 week check was 0.74±0.26V. Pacing thresholds prior to the 1st CMR scan, between the lead switch over and following the scan were 0.69±0.17V, 0.69±0.16V and 0.69±0.16V respectively, p= 1.
Impedance
The mean pacing impedance for the RVOT lead at implant was 739±168Ω and at 2 week check was 655±251Ω. Pacing impedances prior to the CMR scan, between the lead switch over and following the scan were 601±123Ω, 595±114Ω and 579±141Ω respectively, p=0.008.
The mean pacing impedance for Apical lead at implant was 631±130Ω and at 2 week check was 616±81Ω. Pacing impedances prior to the CMR scan, between the lead switch over and following the scan were 612±81Ω, 611±80Ω and 574±69Ω respectively, p=0.004.
Battery
The mean battery voltage prior to, between and following every CMR scan did not alter acutely. CMR scan 1 was at 2.99±0.03V.
Specific absorption rate
The max SAR of 4 w/kg was never exceeded. See Fig 1.
Tendril leads 5 control patients had the MR system but no CMR scans. Table 1 compares the parameters between cohorts over 12 months. A similar trend in parameter changes was observed between the CMR and non-CMR pacing cohort.
Conclusions
Prolonged max power CMR scanning of the St Jude Accent - Tendril IPG system at 1.5 T is safe and has no clinically relevant effects on PCT, voltage and Battery power.
Funding
British Heart Foundation St Jude Fellowship Grant.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Ainslie, M.P., Reid, A., Miller, C.A. et al. MAPS; acute safety data of the St Jude accent - tendril IPG system during prolonged max power CMR scanning. J Cardiovasc Magn Reson 17 (Suppl 1), M6 (2015). https://doi.org/10.1186/1532-429X-17-S1-M6
Published:
DOI: https://doi.org/10.1186/1532-429X-17-S1-M6
Keywords
- Battery Power
- Scanning Protocol
- Specific Absorption Rate
- Safe Device
- Battery Voltage