Beat to beat volumetric analysis in arrhythmia using real time CMR
Journal of Cardiovascular Magnetic Resonance volume 17, Article number: O37 (2015)
Background
ECG-gated cardiac MRI is the gold standard for volumetric evaluation of patients, and clinically, ejection fraction is used as a surrogate for function. We hypothesized that the use of arrhythmia rejection in the presence of ectopic beats compromises the accuracy of hemodynamic measurements. Real-time MRI, coupled with ECG telemetry and semi-automated LV endocardial segmentation, can be used to identify multiple beat morphologies and derive global hemodynamic measurements for each beat.
Methods
Short-axis golden angle radial bSSFP projections (8000 projections/slice) were reconstructed using Gadgetron (non-Cartesian, iterative SENSE) with 34 projections per frame, and slice volume was measured via segmentation of LV endocardial contour with ITK-SNAP. ECG was synchronously recorded with the acquisition, and each image frame was assigned a time point on the ECG (~22 seconds/slice). QRS detection was used to identify cardiac cycles, and similar beats (across slices) were grouped via RR-duration of the beat of interest as well as the previous (loading) beat. The use of the loading beat duration allowed for accurate grouping of premature ventricular contractions (PVCs). For each beat morphology, global volume was obtained by summation of slice volume curves allowing for hemodynamic evaluation via EDV, ESV, SV, and EF.
Results
10 patients were imaged: 5 in sinus rhythm, 3 with PVCs with 19%, 24%, and 33% prevalence, 1 in regular bigeminy, and 1 in irregular bigeminy. All beats detected beats were assigned a morphology, excluding the first beat of each acquisition since the loading conditions of these beats were unknown. A single beat morphology was observed in patients in sinus rhythm while patients with rhythm disturbances demonstrated multiple beat morphologies with large variations in measured volumes and EF. Figure 1 shows real-time and composite hemodynamic measurements in a patient with frequent PVCs (33%). In this patient, the SV and EF varied from 7.2 - 69.9 mL and 14.7 - 69.7% respectively, depending on the beat morphology. This variation suggests that cardiac cine, even with perfect arrhythmia rejection, misses important aspects of ventricular function in the presence of arrhythmias.
Conclusions
ECG-gated cardiac MRI is unable to capture the hemodynamics associated with arrhythmic events. As a result, values such as EF are acquisition - dependent (desired RR-duration determines arrhythmia rejection). By combining real-time volume measurements with ECG recordings, we categorized beat morphologies and provided a more comprehensive evaluation of ventricular function during arrhythmia.
Funding
We would like to acknowledge the National Institutes of Health for support through grants: F31-HL120580, R00-HL108157, R01-EB014346, R01-HL103723, R01-HL63954, T32-HL007954, T32-EB009384.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Contijoch, F., Rears, H., Rogers, K. et al. Beat to beat volumetric analysis in arrhythmia using real time CMR. J Cardiovasc Magn Reson 17 (Suppl 1), O37 (2015). https://doi.org/10.1186/1532-429X-17-S1-O37
Published:
DOI: https://doi.org/10.1186/1532-429X-17-S1-O37