Accuracy and reproducibility of T1rho mapping sequences

Background

Previous studies have shown that T1rho-weighted imaging using long spin locking pulses enables high discrimination between infarct and myocardium without the need for exogenous contrast agents. To optimize imaging protocols, it's beneficial to understand the factors that influence the measurement. We measured the accuracy and reproducibility of spin echo-spin lock (SESL) T1rho-prepared balanced-steady-state free precession (bSSFP) sequences for myocardial relaxation time mapping. We subsequently performed in vivo studies and determined the average T1rho values in normal myocardium.

Methods

A phantom containing 8 homogeneous H2O cylinders with MnCl2 concentrations of 0.01- 0.15 g/mL was used for the comparison of 24 T1rho mapping sequences with various repetition times (TR = 1.8-10 sec), flip angles (0-70 degrees), and numbers of segments (Nseg = 10-70). The scans were performed on a 1.5 T MRI (Avanto model, Siemens), and reference T1rho measurements were obtained by using gradient echo acquisition with TR = 10 sec, flip angle = 90 degrees, Nseg = 1. The overall scan time for this exam was approximately 1.5 hours. Three fitting methods were used to estimate T1rho recovering times,

linear model: log[S(t)] = -TSL/A + B;

2-parameter model: S(t) = B*exp(-TSL/A); and,

3-parameter model: S(t) = B*exp(-TSL/A) + C,

where t is the inversion time, A is the T1rho value, B is the initial magnetization M0 and C is the rms noise level (Figure 1a).

Figure 1
figure 1

(a) The calculated T1rho values were 34.2, 55.0, and 58.6ms for 3-parameter, 2-parameter and linear fitting models respectively. (b) Accuracy of the 24 tested sequences assessed by the difference between each sequence and the gold standard weighted by reference T1rho values. Long TR and small number of segments produce more accurate results. (c) Reproducibility of the 24 tested sequences was assessed by the coefficient of variance. In contrary to results shown in the error maps, TR doesn't have a significant effect on reproducibility. Higher segments and higher flip angles are more reproducible.

In vivo data were acquired on 10 healthy subjects on a 1.5 T scanner (Siemens). Two sequences were applied on each subject (TR = 1.8 sec, flip angle = 70 degrees, Nseg = 31).

Results

For T1rho lower than 50 ms, we found that the 3-parameter model was the most accurate and reproducible. 2-parameter and linear models are preferred for higher T1rho values. As shown in Figure 1-a, the signal intensity decayed to the rms noise level (C, a positive nonzero value) and the 3-parameter fitting model is the only model that captured this behavior.

As shown in Figure 1-b and 1-c, substantial variations were observed in T1rho estimates measured and computed using preferred fitting models, from the 24 sequences. We observed that lower flip angle and higher Nseg increases T1rho underestimation, while higher flip angle and lower Nseg were more accurate and reproducible. SNR is related to the component of amplitude of the transverse magnetization which also varies with flip angle, there's a thus trade-off between T1rho measurement precision and accuracy. Although long TR and small Nseg allows the full relaxation of magnetization producing more accurate results, the resulting long scan time was not very practical in clinical settings.

Figure 2 shows an example of in vivo T1rho and its error map obtained in a 24-year-old healthy male subjects.

Figure 2
figure 2

T1rho map of a normal subject, and the corresponding coefficient of determination map.

Conclusions

We described a number of adjustable factors in T1rho mapping. These factors depend on protocol adjustments, and can potentially influence the accuracy and reproducibility.

Funding

None.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Han, J.Q., Han, Y. & Witschey, W.R. Accuracy and reproducibility of T1rho mapping sequences. J Cardiovasc Magn Reson 17 (Suppl 1), P22 (2015). https://doi.org/10.1186/1532-429X-17-S1-P22

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-17-S1-P22

Keywords