Bioprosthetic mitral valve effective orifice area using 4D flow cardiac magnetic resonance derived time velocity integral. An in vitro comparison with Doppler Echocardiography

Background

4D Flow Cardiac Magnetic Resonance (CMR) is a novel imaging modality to assess bioprosthetic mitral valve (BMV) function. We describe a new, 4D Flow derived velocity time integral (TVI) based method to assess effective orifice area (EOA) for BMVs.

Methods

In our MRI-compatible circulatory loop 4 stented porcine BMVs (27, 29, 31, 33mm) underwent CMR with a 1.5T Siemens scanner. The valves were evaluated at forward stroke volumes of 70, 90 and 110ml at a beat rate of 70bpm. We plotted instantaneous peak velocities and calculated TVI for each scenario. 4D CMR-EOA was calculated by dividing the forward stroke volume measured by our high fidelity flow transducers with the calculated TVI. Doppler derived EOA was calculated identically using the Doppler derived TVI.

Results

4D CMR-EOA was successfully calculated in a total number of 12 tests. 4D Flow CMR-EOA revealed a strong correlation when compared with Doppler derived EOA (r=0.985, p<0.001) and the mean difference was -0.1±0.05cm2 between the two methods. Calculated CMR-TVI by our method correlated well with Doppler derived TVI (r=0.99, p<0.001 and mean difference 3.2±1.97cm).

Conclusions

Our study demonstrated the feasibility of 4D Flow CMR to evaluate bioprosthetic valve EOA with the use of velocity encoded CMR data from 4D Flow Software analysis. Our data showed high agreement with Doppler EOA calculations, demonstrating a potential for 4D Flow CMR in bioprosthetic valve evaluation.

Funding

American Heart Association grant #11BGIA5840008.

figure 1

Figure 1

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Maragiannis, D., Jackson, M., Igo, S. et al. Bioprosthetic mitral valve effective orifice area using 4D flow cardiac magnetic resonance derived time velocity integral. An in vitro comparison with Doppler Echocardiography. J Cardiovasc Magn Reson 17 (Suppl 1), P336 (2015). https://doi.org/10.1186/1532-429X-17-S1-P336

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-17-S1-P336

Keywords