Phantom validation of 4D flow: independent validation of vortex ring volume quantification using planar laser-induced fluorescence

Background

Previous studies suggest that vortex ring formation in the left ventricle of the human heart is a sensitive marker of cardiac diastolic function and overall cardiac health [1]. However, measurement of quantitative vortex ring parameters using 4D phase contrast magnetic resonance (4D PC-MR) has not previously been validated. Therefore, the purpose of this study was to validate measurement of vortex ring volume (VV) by 4D PC-MR using planar laser-induced fluorescence (PLIF) as the reference standard in a phantom setup.

Methods

We constructed a pulsatile pump and a water tank with a 25mm nozzle (Figure 1), and five different pump settings with different pulse volumes and velocities were used. PLIF was performed using the fluorescent dye Rhodamine 590 (Rhodamine 6G, Exciton Inc., Ohio, USA) excited using a 532 nm Nd:YAG laser in a 1 mm vertical laser sheet. Images were acquired 400 ms after pump initiation, i.e. after complete vortex ring formation. Nozzle stroke volumes were measured using 2D PC-MR (voxel size 2.4x2.4x6 mm, VENC 50 cm/s). 4D PC-MR was acquired at 1.5T with 3x3x3 mm voxels and 50 ms temporal resolution. Vortex ring volume (VV) was quantified in PLIF by manual delineation (Figure 2A), and in 4D PC-MR data using manual delination of Lagrangian Coherent Structures (Figure 2B), a method for analysis of flow data that enables detection of the vortex ring boundary [2].

Figure 1
figure 1

Design and construction of the vortex ring flow phantom. a) 2D view of the flow rectifier and vortex ring tank. All measures are in millimeters (mm). b) 3D view of the flow rectifier and vortex ring tank. c) Pump design. A servo motor powers a ball screw linear actuator, which in turn moves a piston-cylinder apparatus. A rubber membrane separates the flow medium from the pump to prevent abrasive particles from entering the piston/cylinder apparatus.

Figure 2
figure 2

Data analysis in PLIF and 4D PC-MR images and quantitative results for vortex ring volume (VV). Panel A: Analysis of PLIF images. Panel B: Analysis of 4D PC-MR data using Lagrangian Coherent Structures (LCS). Panel C: Comparison of vortex ring volume (VV) at the five pump settings.

Results

Stroke volumes ranged from 12-37 ml. Vortex ring volume (VV) showed excellent agreement between PLIF and 4D PC-MR (Figure 2C, R2 = 0.99, bias 2.4±1.5 ml).

Conclusions

This study shows that vortex ring volume (VV) can be reliably quantified using 4D PC-MR.

Funding

This study was supported by Swedish Research Council grants VR 621-2005-3129, VR 621-2008-2949, and VR K2009-65X-14599-07-3, National Visualization Program and Knowledge Foundation grant 2009-0080, the Medical Faculty at Lund University, Sweden, the Region of Scania, Sweden and the Swedish Heart-Lung Foundation.

References

  1. Gharib M, Rambod E, Kheradvar A, Sahn DJ, Dabiri JO: Optimal vortex formation as an index of cardiac health. Proc Natl Acad Sci. 2006, U S A, 103: 6305-8. 10.1073/pnas.0600520103.

    Google Scholar 

  2. Shadden SC, Dabiri JO, Marsden JE: Lagrangian Analysis of fluid transport in empirical vortex ring flows. Phys Fluids. 2006, 18: 047105-10.1063/1.2189885.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Töger, J., Bidhult, S.L., Revstedt, J. et al. Phantom validation of 4D flow: independent validation of vortex ring volume quantification using planar laser-induced fluorescence. J Cardiovasc Magn Reson 17 (Suppl 1), P38 (2015). https://doi.org/10.1186/1532-429X-17-S1-P38

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-17-S1-P38

Keywords