Skip to main content
  • Poster presentation
  • Open access
  • Published:

Aortic Stenosis assessment with a 3-directional phase contrast magnetic resonance technique. Comparison to transthoracic echocardiography

Background

Transthoracic Doppler-echocardiography (TTE) is the standard clinical method for diagnosis and staging of aortic stenosis (AS). AS staging is based on measurement of aortic peak velocity, transvalvular gradient, and calculation of aortic valve area. Unidirectional through-plane phase-contrast magnetic resonance imaging (1DPC-MRI) has been widely applied in clinical imaging to quantify aortic peak velocities and flow. Nonetheless, 1DPC-MRI requires accurate positioning of imaging planes perpendicular to flow direction in order to avoid peak velocity underestimation, which can be challenging in patients with multiple or eccentric jets. Therefore PC techniques with multi-directional velocity quantification would likely improve the accuracy of velocity determination, and allow for more accurate grading of AS severity. The aim of this study is to determine whether a rapid technique that is able to capture 3 directions of velocity in a 2D image plane in a single breath-hold (3DPC-MRI) provides more accurate estimation of diagnostic parameters compared with the traditional 1DPC-MRI, using TTE as the reference standard.

Methods

We included 13 patients diagnosed with mild to severe AS by TTE (nine men, age range: 39-85 years, median age: 65 years). The average time elapsed between TTE and CMR was 24 days. Velocity-encoded CMR included breath-hold 1DPC-MRI and 3DPC-MRI. Acquisition parameters are listed in Table 1. After manual tracing of aortic valve contours, quantitative image analysis was performed offline using custom software developed in MATLAB (Mathworks, Natick, MA). The pixel with the highest average velocity within the valve contour was used to extract aortic peak velocities and peak and mean trans-valvular gradients, for comparison with TTE. Agreement between CMR and TTE parameters were explored using intraclass correlation coefficient (ICC). Statistical analyses were performed using SPSS, version 21 (IBM).

Results

3DPC-MRI peak velocities showed higher correlation with TTE (ICC 0.88, p<0.001), than 1DPC-MRI (ICC 0.82, p<0.001). 3DPC-MRI mean gradient estimation also showed better correlation with TTE results (ICC 0.72, p<0.001) than 1DPC-MRI (ICC 0.42, p=0.013). Since peak gradient estimations derive from peak velocity estimations, 3DPC-MRI peak gradients again showed better correlation with TTE (ICC 0.90, p<0.001) than 1DPC-MRI (ICC 0.80, p<0.001). Bland-Altman plots between TTE vs 1DPC-MRI (A-C), and TTE vs 3DPC-MRI parameters (D-F) are shown in Figure 1. Note less overall bias for the 3DPC-MRI technique.

Figure 1
figure 1

Bland-Altman plots of comparison between parameters derived from TTE versus 1DPC-MRI (A-C) and 3DPC-MRI (D-F).

Conclusions

Initial results in a small patient cohort support the hypothesis that 3DPC-MRI provides better estimation of hemodynamic parameters in AS patients in comparison to 1DPC-MRI.

Funding

Research grant from Siemens.

Table 1 Imaging parameters.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

da Silveira, J.S., Smyke, M., Jin, N. et al. Aortic Stenosis assessment with a 3-directional phase contrast magnetic resonance technique. Comparison to transthoracic echocardiography. J Cardiovasc Magn Reson 17 (Suppl 1), P381 (2015). https://doi.org/10.1186/1532-429X-17-S1-P381

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-17-S1-P381

Keywords