Free-breathing 3D phase-sensitive inversion recovery late gadolinium enhancement at 3.0 Tesla: reliability and image quality in ischemic and non-ischemic cardiomyopathy in comparison with multiple breath-hold 3D imaging
Journal of Cardiovascular Magnetic Resonance volume 17, Article number: P97 (2015)
Background
In both ischemic (ICM) and non-ischemic (NICM) cardiomyopathy late gadolinium enhancement (LGE) is an important cardiovascular magnetic resonance (CMR) technique. LGE CMR, traditionally performed in 2D during multiple breath-holds (MB), is challenging for vulnerable patients and subject to slice misregistration. Therefore, LGE CMR during free-breathing (FB) is more robust and enables to perform 3D acquisition, increase resolution and apply phase-sensitive inversion recovery (PSIR). We developed and clinically tested a high spatial resolution PSIR LGE sequence using a respiratory navigator in both ICM and NICM patients. Reliability and image quality were compared to a multiple breath-hold 3D approach (MB-3D).
Methods
Approval was waived by the institutional ethical committee. Between May 2012 and November 2013 FB-3D and MB-3D sequences were acquired in patients suspected of ICM and NICM. Data of 48 patients (36 men; mean age ± standard deviation: 60.8 years ± 10.9; 34 ICM) were retrospectively collected. LGE sequences were acquired using a 3.0 Tesla MR system. For both FB-3D (1.68 x 1.68 x 3.4 mm) and MB-3D (1.86 x 2.8 x 10 mm) a PSIR sequence was used. Multiplanar reformats with high resolution (FB-3D-HR; 0.91 mm isotropic voxels) and low resolution (FB-3D-LR; 1.86 x 2.8 x 10 mm) were constructed. LGE mass and image quality (SNR, CNR and edge sharpness) were compared using the Friedman test, Wilcoxon signed rank test, spearman correlation and Bland-Altman analysis.
Results
34 ICM patients (23 chronic MI; 4 sub-acute MI; 7 acute MI) and 14 NICM were included (5 dilated cardiomyopathy; 4 hypertrophic cardiomyopathy; 3 peri-/myocarditis; 2 other). In figures 1 and 2 3D reconstructions of two patients are shown.In the complete dataset of 48 patients, there were no significant differences between FB-3D-HR, FB-3D-LR and MB-3D datasets in terms of LGE mass (FB-3D-HR: (median [interquartile range]): 8.5 g [4.4-19.9]; FB-3D-LR: 10.7 g [4.5-22.9]; MB-3D: 9.8 g [5.1-19.8]; p = 0.099). LGE mass correlated well (FB-3D-HR vs MB-3D: r=0.922, p=0.01; FB-3D-LR vs MB3D: r=0.922, p=0.01) with Bland Altman analysis indicating good agreement for FB-3D-HR vs MB-3D (mean difference ± SD: 1.15 ± 5.45 gram) and FB-3D-LR vs MB-3D (mean difference ± SD: 0.00 ± 5.47 gram). SNR was higher in both FB-3D-HR (median [interquartile range]: 244 [192-326], p=0.001) and FB-3D-LR (825 [498-1417], P<0.001) as compared to MB-3D (200 [146-263]). CNR was also higher in both FB-3D-HR (226 [164-277], p<0.001) and FB-3D-LR (691 [480-1086], p<0.001) as compared to MB-3D (216 [162-260]). Edge sharpness was 25% higher in FB-3D-HR as compared to MB-3D (0.21 [0.19-0.23] vs 0.17 [0.15-0.20], p<0.001).
Conclusions
Free-breathing 3D phase-sensitive inversion recovery LGE CMR enables reliable myocardial scar tissue assessment with significantly improved image quality as compared with multiple breath-hold 3D imaging in ischemic and non-ischemic cardiomyopathy.
Funding
No funding.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Bizino, M., Amersfoort, J., Tao, Q. et al. Free-breathing 3D phase-sensitive inversion recovery late gadolinium enhancement at 3.0 Tesla: reliability and image quality in ischemic and non-ischemic cardiomyopathy in comparison with multiple breath-hold 3D imaging. J Cardiovasc Magn Reson 17 (Suppl 1), P97 (2015). https://doi.org/10.1186/1532-429X-17-S1-P97
Published:
DOI: https://doi.org/10.1186/1532-429X-17-S1-P97