Peak velocity estimation in aortic stenosis patients using a fast three-directional two-dimensional phase contrast technique in a single breath-hold: comparison to unidirectional phase contrast MRI and transthoracic echocardiography
Journal of Cardiovascular Magnetic Resonance volume 18, Article number: P335 (2016)
Background
Assessment of aortic valve stenosis (AVS) severity is crucial for valve replacement indication and is typically performed by transthoracic Doppler-echocardiography (TTE). However, TTE may be suboptimal in up to 30% of patients. Unidirectional through-plane phase-contrast magnetic resonance imaging (1Dir PC-MRI) is the most common MRI technique used to quantify peak velocities (Vpeak) and flow (Figure 1A). Nonetheless, 1Dir PC-MRI has been shown to underestimate aortic velocities if imaging planes are not prescribed exactly perpendicular to flow direction. Thus, multi-directional velocity quantification would likely improve the accuracy of peak velocity measurements, and allow for more accurate grading of AVS severity. We sought to determine whether a PC technique capable of measuring 3 directions of velocity in a 2D image plane in a single breath-hold (3Dir PC-MRI) (Figure 1B) provides more accurate estimation of Vpeak compared to the traditional 1Dir PC-MRI, using TTE as the reference standard.
Methods
Patients with variable degrees of aortic valvular disease were prospectively included, and assessed with both TTE and CMR. 1Dir (TR/TE = 49/2.3 ms, α = 250, BW = 420Hz/px, segmented GRE) and 3Dir PC-MRI (TR/TE = 49/2.8 ms, α = 150, BW = 1860 Hz/px, segmented EPI) data were acquired at 3 levels above the aortic valve using a 1.5T Siemens Avanto. Imaging parameters were: 6 mm slice thickness, FOV: 380 × 300 mm2, matrix = 192 × 140, Venc 200-550 cm/s, prospective ECG triggering, GRAPPA r = 2. Quantitative image analysis was performed offline using Matlab (Mathworks, Natick, MA). 3Dir PC-MRI Vpeak was calculated pixel by pixel using the root sum square of the three orthogonal velocities (i.e., direction independent speed). After magnitude and flow thresholding to eliminate noise, the pixel with the highest velocity within the valve contour was used for comparison to TTE. Stroke volumes (SV) were also estimated from through-plane 1Dir and 3Dir PC-MRI and compared to left ventricular volumes from SSFP cine imaging.
Results
Forty-one patients were enrolled (25 males, median age 68 years [range 27-85 years]). The average interval between TTE and CMR was 33 ± 23 days. 1Dir PC-MRI tended to underestimate Vpeak while 3Dir PC-MRI measured a higher Vpeak than TTE. Bland-Altman Plots in Figure 1 C/D illustrate a mean difference of -0.1 m/s and +0.2 m/s for 1Dir and 3Dir PC-MRI, respectively. Good correlation was observed between both 1Dir and 3Dir PC-MRI SV versus cine SV at all levels above the aortic valve (ρc = 0.85 to 0.89), with a slight tendency of SV overestimation by 1Dir PC-MRI and underestimation by 3Dir PC_MRI (Table 1).
(A/B) Differences in velocity estimation between 1Dir and 3Dir PC-MRI. While 1Dir PC-MRI computes velocity in one direction (Z), 3Dir PC-MRI simultaneously computes velocities in 3 directions (X, Y and Z) in a single breath-hold. (C/D) Bland-Altman plots of comparison between peak velocities from TTE versus 1Dir PC-MRI, and TTE versus 3Dir PC-MRI.
Conclusions
The higher Vpeak by 3Dir PC-MRI may be explained by its directional independence, as opposed to 1Dir PC-MRI and TTE, which can only accurately measure velocity perpendicular or parallel to the stenotic jet, respectively. 3Dir PC-MRI may therefore offer an advantage over both 1Dir PC-MRI and TTE in the clinical assessment of AVS.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
da Silveira, J.S., Smyke, M.E., Ahmad, R. et al. Peak velocity estimation in aortic stenosis patients using a fast three-directional two-dimensional phase contrast technique in a single breath-hold: comparison to unidirectional phase contrast MRI and transthoracic echocardiography. J Cardiovasc Magn Reson 18 (Suppl 1), P335 (2016). https://doi.org/10.1186/1532-429X-18-S1-P335
Published:
DOI: https://doi.org/10.1186/1532-429X-18-S1-P335
Keywords
- Stroke Volume
- Peak Velocity
- Aortic Valve Stenosis
- Aortic Valve Stenosis Severity
- Phase Contrast Technique