Rat heart T2-mapping with full coverage of the left ventricle myocardium

Background

MRI cardiac T2-mapping visualizes edema and has been used for tissue characterization in settings such as myocarditis and ischemia. Clinically it is typically performed via spin-echo based techniques, but preclinically, the high heart rates of rats and mice make such approaches sensitive to motion. Successful implementations instead employ T2-preparation followed by a segmented gradient-echo readout (Beyers et al. MRM 2012, Coolen et al. MRM 2014). It is argued here that when full coverage of the left ventricle myocardium (LVM) is needed, such techniques benefit from applying phase encoding in the slice direction as opposed to using a multi-slice approach.

Methods

All work was approved by the local ethical committee. At day 0, 14 rats (male Clr: CD, Sprague-Dawley, 300 g) were subjected to sham operation (open chest surgery, non-occlusive suture in LVM). Scanning was performed at days 1, 4, and 8, under Isoflurane anaesthesia on a Bruker BioSpec 4.7T scanner equipped with a 72 mm quadrature coil. T2-preparation module: non-selective 90°-180°-90° excitation pulses with four echo times (5, 23, 41, 59 ms). Readout module: 3D gradient echo with phase encoding in the slice direction, FOV 40 × 40 mm2, slice thickness 1.5 mm, 128 × 126 × 8 matrix, FA 30°, TR 4.0 ms, TE 1.5 ms, bandwidth 50 kHz, 6 lines per segment synchronized to the systolic phase. The repetition time between segments was 2300 ms (minimum 2000 ms, but depending on the respiration rate). The acquisition time was 6 min. 30 s. per echo-time. T2 maps were calculated via the method described by Pei et al. MRM 2015. Regions-of-Interests (ROIs) were manually placed in the LVM and in the back muscle. ROI-level T2s were calculated as the harmonic mean of pixel-level T2s.

Results

T2 appeared homogeneous both in the LVM and in the back muscle (Figure 1). LVM T2 did not change over the eight-day period (Table 1) and remained fractionally higher than corresponding data in the back muscle at all-time points (p < 0.0001, two-sided t-test).

Figure 1
figure 1

T2 maps covering the full extent of the rat left ventricle myocardium. Gray-scale levels express T2-values in ms. Pixels with low SNRs are shown as black.

Table 1 T2 in rat left ventricle myocardium (LVM) and back muscle (mean ± standard deviation, n = 14) at days 1, 4, and 8 at 4.7T.

Conclusions

Precise T2-maps covering the full LVM were generated in sham operated rats with a signal acquisition technique earlier not used in this setting. T2 remained constant in the LV for eight days contrasting a previous report in healthy mice (Coolen et al. MRM 2014). When multi-slice imaging cannot be interleaved, as is the case for sequences relying on global T2 preparation, employing a phase encoding gradient in the slice direction is inherently a more efficient signal sampling strategy (Edelstein et al. MRM 1986). The signal-to-noise ratio (SNR) is increased by a factor √N for a given acquisition time, where N is the number of slices. Alternatively, for a fixed SNR, a potential reduction in acquisition time by a factor N can be reached. In the present study a standard volume coil was used. Further improvements in SNR or reductions in acquisition times can be obtained by combining the proposed imaging sequence with surface coils and imaging acceleration techniques.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Johansson, E., Albery, T., Palmér, M. et al. Rat heart T2-mapping with full coverage of the left ventricle myocardium. J Cardiovasc Magn Reson 18 (Suppl 1), P49 (2016). https://doi.org/10.1186/1532-429X-18-S1-P49

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-18-S1-P49

Keywords