3D T1-weighted self-gated cardiac MRI for assessing myocardial infarction in mouse models

Background

Mouse models with mycardial infarction (MI) have been intensively used to investigate the cardiac remodeling and functional change [1,2]. MRI with ECG triggering and respiratory gating (or breath-holding) is commonly used to assess MI due to its noninvasive nature. However, the manipulations of external ECG triggering and respiratory gating are cumbersome due to the small size of mouse. To address this issue, a 3D self-gating (SG) MR technique with stack-of-stars sampling trajectories was proposed for retrospectively cardiac and respiratory-gated MI imaging in mouse models [3].

Methods: Technical Design: MR data was acquired by a T1-weighted GRE sequence with stack-of-stars sampling trajectories and a partition-first golden-angle reordering (Figure 1a). The centers of k-space lines acquired at the same angle from individual partitions were aggregated and used as an SG time point. The periodic respiratory and cardiac motions were then detected through an iterative filtering process on the SG time series according to the cardiac rate of 300~500 per minute and respiration rate of 70~110 per minute. After resorting the imaging data into appropriate cardiac and respiratory phases, motion-artifact-free were finally reconstructed.

Experiments: The SG technique was preliminarily validated on 5 mice with MI induction and all MR scans were performed on a 3T scanner (Siemens Tim Trio, Germany) with a customized 4-channal mouse coil. Typical imaging parameters for the ungated GRE sequence included: flip angle = 18°, TR = 4.2 ms, TE = 2.4 ms, spatial resolution = 0.6 × 0.6 × 1.5 mm3, bandwidth = 620 Hz/Pixel, partition number = 12, and a total number of 3200 projections were continuously collected, corresponding to a fixed scan time of 3 min. An amount of 0.5 ml gadolinium contrast agent with concentration of 0.5 mmol/ml (Consun Pharmaceutical Group Limited, GuangZhou, China) was injected to enhance the MI. The mice were sacrificed immediately after MRI for histological analysis and comparison to the MR results.

Results

All MR scans were successfully conducted. The cardiac motion detected by the SG technique matched nicely to the ECG signals recorded by a small animal ventilator (Chengdu Taimeng Software Co.LTD, Chengdu, China) (Fig. 1b&1c). And the MI regions detected by the SG technique were also matched well to the histopathology analysis results (Fig. 2).

Figure 1
figure 1

Self-gating signals were compared with ECG signals. (a) k-space sampling trajectories, the dot (•) denotes the SG data; (b) ECG signals recorded by the small animal ventilator; (c&d) cardiac and respiratory motions detected by the SG technique.

Figure 2
figure 2

(a) Representative histopathology pictures (top row) and reconstructed MR images (bottom row) from a mouse with MI induction. The regions of MI (arrows) matched nicely between the histology and the proposed SG technique. (b) The correlation of the ratio of the infarction size and myocardial size between histology and the proposed method.

Conclusions

A 3D SG technique was developed for assessing MI in mouse model. The preliminary in vivo study has demonstrated that the technique can correctly detect the MI, which may outperform the conventional MR techniques with ECG-triggering and respiratory gating.

References

  1. Lutgens E, et al: Cardiovascular research. 1999, 41 (3): 586-593.

    Article  CAS  PubMed  Google Scholar 

  2. Hofmann U, et al: Circulation. 2012, 125 (13): 1652-1663.

    Article  CAS  PubMed  Google Scholar 

  3. Liu J, et al: MRM. 2010, 63 (5): 1230-1237.

    Article  PubMed Central  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

The Creative Commons Public Domain Dedication waiver (https://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhang, X., Chen, H., Wei, Z. et al. 3D T1-weighted self-gated cardiac MRI for assessing myocardial infarction in mouse models. J Cardiovasc Magn Reson 18 (Suppl 1), Q49 (2016). https://doi.org/10.1186/1532-429X-18-S1-Q49

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-18-S1-Q49

Keywords