- Workshop presentation
- Open Access
- Published:
Right-ventricular assessment using a segmented cine acquisition employing iterative SENSE reconstruction with spatio-temporal L1 regularization: Initial clinical experience
Journal of Cardiovascular Magnetic Resonance volume 18, Article number: W18 (2016)
Background
Cardiac MR (CMR) has emerged as the gold standard for assessing right-ventricular (RV) size and function with cine acquisitions. However, dyspnea often accompanies RV dysfunction, limiting patient ability to breath-hold and consequently image quality at CMR. The application of a novel iterative reconstruction technique to segmented CMR cine acquisitions may enable higher acceleration factors, shortening image acquisitions, while maintaining image quality for RV assessment. The purpose of this study is to evaluate the clinical utility of a prototype iterative reconstruction algorithm utilizing spatio-temporal L1-regularization (iteratively reconstructed Sparse-SENSE) in the quantitative assessment of RV systolic function.
Methods
9 healthy volunteers (44.3 ± 13.5 yrs) and 29 patients (54.3 ± 13.8 yrs) were scanned on a 1.5T scanner. Cine images were acquired in the 4-chamber and short-axis orientations using conventional generalized auto-calibrating partially parallel acquisition (GRAPPA) factor 2 acceleration ("GRAPPA 2"), a spatio-temporal undersampled TSENSE acquisition with factor 4 acceleration ("TSENSE 4"), and iteratively reconstructed Sparse SENSE with an acceleration factor of 4 ("IS-SENSE 4"). Quantitative assessment of RV function was performed by a single reviewer. A subset of 14 subjects was re-analyzed by the initial reviewer and a second reviewer to assess intraobserver and interobserver agreement. Two independent reviewers scored images for image quality, noise, and artifacts on a 5-point Likert scale. All continuous variables were analyzed using linear regression with associated R2 values. Comparison of acquisition techniques was performed using univariate analysis of variance (ANOVA), and an intraclass correlation coefficient (ICC) was calculated to assess both intraobserver and interobserver agreement.
Results
Among all three acquisition types, differences in RV ejection fraction (EF) and BSA-indexed end-diastolic volume (EDVi) were shown to be statistically insignificant. The R2 values for linear regression of TSENSE 4 and IS-SENSE 4 versus GRAPPA 2 were 0.34 and 0.72 for RV EF, and 0.61 and 0.76 for RV EDVi, respectively (Figure 1). ICC results for both intraobserver and interobserver consistency between GRAPPA 2 and IS-SENSE 4 yielded excellent agreement. Qualitative review yielded small, but statistically significant differences (p < 0.05) in image quality and noise between both TSENSE 4 and IS-SENSE 4 compared to GRAPPA 2 (Figure 2). All three techniques were rated as nearly artifact free.
A: linear regression analysis of TSENSE 4 versus GRAPPA 2 for RV EF B: linear regression analysis of IS-SENSE 4 versus GRAPPA 2 for RV EF C: linear regression analysis of TSENSE 4 versus GRAPPA 2 for BSA indexed RV EDV D: linear regression analysis of IS-SENSE 4 versus GRAPPA 2 for BSA indexed RV EDV E: Bland-Altman plot comparing TSENSE 4 and GRAPPA 2 RV EF F: Bland-Altman plot comparing IS-SENSE 4 and GRAPPA 2 RV EF G: Bland-Altman plot comparing TSENSE 4 and GRAPPA 2 BSA indexed RV EDV H: Bland-Altman plot comparing IS-SENSE 4 and GRAPPA 2 BSA indexed RV EDV. Red points indicate averages outside of 1.96 SD. EF = ejection fraction; EDV = end diastolic volume
Conclusions
2D CINE imaging acquisitions with prototype IR Sparse-SENSE reconstruction and an acceleration factor of 4 accurately and reliably quantitates RV systolic function parameters, while maintaining image quality comparable to segmented cine sequences with GRAPPA factor 2 acceleration. IR Sparse-SENSE has the potential to improve imaging options in patients with intermittent arrhythmias or difficulties with breath holding.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Bogachkov, A., Ayache, J.B., Allen, B.D. et al. Right-ventricular assessment using a segmented cine acquisition employing iterative SENSE reconstruction with spatio-temporal L1 regularization: Initial clinical experience. J Cardiovasc Magn Reson 18 (Suppl 1), W18 (2016). https://doi.org/10.1186/1532-429X-18-S1-W18
Published:
DOI: https://doi.org/10.1186/1532-429X-18-S1-W18
Keywords
- Iterative Reconstruction
- Acceleration Factor
- Iterative Reconstruction Technique
- Maintain Image Quality
- Cine Acquisition