T2-mapping- influence of arrhythmia and heart rate: A phantom experiment
Journal of Cardiovascular Magnetic Resonance volume 18, Article number: W36 (2016)
Background
Myocardial tissue differentiation applying Cardiovascular magnetic resonance (CMR) offers the capability to detect myocardial injury already in preserved ejection fraction. Detection of edema is a unique capability of CMR. Conventional T2 imaging suffers from technical challenges related to diseases specific issues like tachycardia, atrial fibrillation and pericardial effusion. T2-mapping seems to be more stable based on visual assessment. Little is known about its dependency of disease relate abnormalities. We aimed to investigate on the influence of heart rhythm abnormalities on T2 mapping.
Methods
We performed phantom experiments using a 1.5 Tesla Siemen Avanto Fit (Siemens Healthcare, Erlangen, Germany) equipped with a 32-channel cardiac coil. The T2 phantom consisted of five tubes (figure 1A). They were filled with different concentrations of manganese(II) chloride, to achieve various T2-times (60-80 ms). Tube 1 was set to 65 ms, tube 2 (70 ms), tube3 (60 ms), tube3 (80 ms) and tube 4 (control tube 65 ms) and the solution around the tubes ("blood" =150 ms).The phantom was surrounded by two NISO4 + NaCL doped water bottles to achieve enough load for MRI. We applied a T2 prepared SSFP- based T2 mapping technique with the following parameters: voxel size: 1.6 × 1.6 × 6 mm³, TR 2.7 ms, flip angle 70°, TE 1,15 ms; GRAPPA reduction factor 2. Sinus rhythm heart rates of 60, 80,100 and 120 bpm were simulated using an ECG simulator (Siemens Healthcare). Arrhythmia was simulated using a commercial ECG simulation device (ES 300, SPI electronics). To assess repeatability of the T2 times, the T2 maps were acquired 6 times for each heart rate and rhythm. T2 times were quantified based on the inline T2 maps by drawing a ROI covering the entire phantom intersection using cvi42 5.0 (Circle Cardiovascular Imaging Calgary,Canada).
Results
T2 times were evaluable in all settings. T2 times did not show significant differences depending on heart rate variability within each tube (tube1 p = 0.541; tube2 p = 0.661; tube3 p = 0.51; tube4 p = 0.813; tube5 p = 0.729). Details are given in table 1. Figure 1B shows the T2 phantom using a simulated sinus rhythm at 60 bpm.
Conclusions
The influence of heart rhythm abnormalities and heart rate on T2-times seems to be not relevant. That gives the possibility to identify myocardial edema also in patients with arrhythmia. Further patient studies are needed.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Prothmann, M., Dieringer, M.A. & Schulz-Menger, J. T2-mapping- influence of arrhythmia and heart rate: A phantom experiment. J Cardiovasc Magn Reson 18 (Suppl 1), W36 (2016). https://doi.org/10.1186/1532-429X-18-S1-W36
Published:
DOI: https://doi.org/10.1186/1532-429X-18-S1-W36
Keywords
- Heart Rate Variability
- Cardiovascular Magnetic Resonance
- Pericardial Effusion
- Siemens Healthcare
- Preserve Ejection Fraction