Skip to main content

Table 2 Pulse sequence parameters

From: Three-dimensional balanced steady state free precession myocardial perfusion cardiovascular magnetic resonance at 3T using dual-source parallel RF transmission: initial experience

 

TGRE

bSSFP

TR

1.8 ms

2.2 ms

TE

0.70 ms

1.04 ms

Flip angle

15°

35°

RF Shimming

Yes (local)

Yes (local)

Partial Echo, Halfscan (Partial Fourier) in x-y and z planes

Yes 0.75/0.75

Yes 0.75/0.75

Spatial resolution

2.3 × 2.3 × 10 mm3

2.3 × 2.3 × 10 mm3

No slices

12

12

Acquisition

191 ms

211 ms

SAR

<30%

<88%