Normal values for cardiovascular magnetic resonance in adults and children
Journal of Cardiovascular Magnetic Resonance volume 17, Article number: 29 (2015)
Abstract
Morphological and functional parameters such as chamber size and function, aortic diameters and distensibility, flow and T1 and T2* relaxation time can be assessed and quantified by cardiovascular magnetic resonance (CMR). Knowledge of normal values for quantitative CMR is crucial to interpretation of results and to distinguish normal from disease. In this review, we present normal reference values for morphological and functional CMR parameters of the cardiovascular system based on the peer-reviewed literature and current CMR techniques and sequences.
Introduction
Quantitative cardiovascular magnetic resonance (CMR) is able to provide a wealth of information to help distinguish health from disease. In addition to defining chamber sizes and function, CMR can also determine regional function of the heart as well as tissue composition (myocardial T1 and T2* relaxation time). Advantages of quantitative evaluation are objective differentiation between pathology and normal conditions, grading of disease severity, monitoring changes under therapy and evaluating prognosis.
Knowledge of normal values is required to interpret the disease state. Thus, the aim of this review is to provide normal reference values for morphological and functional CMR parameters of the cardiovascular system based on a systematic review of the literature using current CMR techniques and sequences. Technical factors such as sequence parameters are relevant for CMR, and these factors are provided as in relationship to the normal values. In addition, factors related to post processing will affect the CMR analysis, and these factors are also described. When multiple peer-reviewed manuscripts are available for normal values, we describe the criteria used to select data for inclusion into this review. When feasible, we provide weighted means based on these literature values. Finally, demographic factors (e.g. age, gender, and ethnicity) may have an influence on normal values and are specified in the review.
Statistical analysis
Results from multiple studies reporting normal values for the same CMR parameters were combined using a random effects meta-analysis model as implemented by the metan command [1]. This produced a weighted, pooled estimate of the population mean of the CMR parameters in the combined studies. Upper and lower limits were calculated as ±2SDp, where SDp is the pooled standard deviation calculated from the standard deviations reported in each study [2]. Statistical analyses were performed with the Stata software package (version 13.1, StataCorp, College Station, TX).
Left ventricular dimensions and functions in the adult
CMR acquisition parameters
The primary method used to assess the left ventricle is steady state free precession (SSFP) technique at 1.5 Tesla. Steady-state free precession (SSFP) technique yields significantly improved blood-myocardium contrast compared to conventional fast gradient echo (FGRE). However, at 3 Tesla, fast gradient echo CMR may also be used. To date however, no studies have presented normal data at 3 Tesla. The derived cardiac volumes and ventricular mass are known to differ for SSFP and FGRE CMR, so that normal ranges are different for each method [3].
Publications presenting reference values of the left ventricle based on the SSFP technique are listed in Table 1.
CMR analysis methods
Papillary muscle mass has been shown to significantly affect LV volumes and mass [6]. No uniformly accepted convention has been used for analyzing trabeculation and papillary muscle mass [7]. Papillary muscle mass has been noted to account for approximately 9% of total LV mass using FGRE technique [6]. Thus, tables of normal values should specify the status of the papillary muscles in the CMR analysis. Tables 2 and 3 provide normal values based on papillary muscle mass added to the remainder of the myocardial mass.
The majority of software approaches use a combination of semi-automated feature recognition combined with manual correction of contours. Short-axis images are most commonly analyzed on a per-slice bases by applying the Simpson’s method (“stack of disks”) [8]. An example of left ventricular contouring is shown in Figure 1.
Demographic parameters
Gender has been demonstrated to have significant independent influence on ventricular volumes and mass. All absolute and normalized volumes decrease in relationship to age in adults [5] in a continuous manner. When considering younger (e.g. <65 years) versus older adults (≥65 years), most studies have shown significant differences in normal values for mass and volumes. For convenience, both average, and younger/ older normal values are given in the tables as available in the literature. An age-related normal value may be useful for patients who are at the upper or lower limits of the values in Tables 2 and 4.
Studies included in this review
Multiple studies have presented cohorts of normal individuals for determining normal dimensions of the left ventricle. For the purpose of this review, only cohorts of 30 or more normal subjects by gender using SSFP CMR have been included. Only data at 1.5T is available for normal subjects using SSFP short axis imaging. Inclusion criteria for the tables below also included a full description of the subject cohort (including the analysis methods used), age and gender of subjects. One study used SSFP radial imaging, and is not included in this review [9].
Multiple studies (not shown in the tables) have used FGRE technique at 1.5T [9-13]. While FGRE is currently used at 3T in some settings, the relevance of FGRE technique at 1.5T to that at 3T is not known.
Because slice FGRE acquisition parameters at 3T are different than at 1.5T, adaptation of 1.5T FGRE normal parameters to 3T FGRE imaging is not recommended. Information on ethnicity in relationship to LV parameters is not available for SSFP technique. Finally, the studies in Table 1 were all conducted in European centers. Normal values for left ventricular dimensions and functions according to these studies are presented in Tables 2, 3 and 4. Age dependent normal values for men and women are also presented in Figures 2 and 3.
Left ventricular volumes, mass and function in systole and diastole normalized to age and body surface area for males according to reference [ 5 ].
Left ventricular volumes, mass and function in systole and diastole normalized to age and body surface area for females according to reference [ 5 ].
Additional LV function parameters
In addition to ejection fraction, Maceira et al. have provided additional functional parameters that may be useful in some settings [5]. For diastolic function, the derivative of the time/ volume filling curve expresses the peak filling rate (PFR). Both early (E) and active (A) filling rates are provided. In addition, longitudinal atrioventricular plane descent (AVPD) and sphericity index (volume observed/volume of sphere using long axis as diameter) at end diastole and end systole are given. These latter parameters are not routinely used for clinical diagnosis.
Right ventricular dimensions and functions in the adult
CMR acquisition parameters
For measurement of right ventricular volumes a stack of cine SSFP images acquired either horizontally or in short axis view can be used [7].
CMR analysis methods
Similar to the left ventricle, analysis of the right ventricle is usually performed on a per slice basis by manual contouring of the endocardial and epicardial borders. Volumes are calculated based on the Simpson’s method [8]. The right ventricular volumes and mass are significantly affected by inclusion or exclusion of trabeculations and papillary muscles [14,15]. For manual contouring, inclusion of trabeculations and papillary muscles as part of the right ventricular volume will achieve higher reproducibility [7,14,15]. However, semiautomatic software is increasingly used for volumetric analysis, enabling automatic delineation of papillary muscles [16]. Therefore normal values for both methods are provided. An example of right ventricular contouring using a semiautomatic software is shown in Figure 4.
Detailed recommendations for right ventricular acquisitions and post processing have been published [7].
Demographic parameters
BSA has been shown to have an independent influence on RV mass and volumes [16]. Absolute and normalized RV volumes are significantly larger in males compared to females [3,4,16]. Further, RV mass and volumes decrease with age [4,16].
Studies included in this review
Criteria regarding study inclusion are identical compared to the left ventricle. Three studies based on SSFP imaging were included (Table 5). In two studies, trabeculations and papillary muscles were included as part of the right ventricular cavity [3,4], and pooled weighted mean values of the two studies are presented in Table 6. In the third study papillary muscles were considered part of the right ventricular mass [16]. Similar to the left ventricle, data is presented as a younger age (<60 years) and an older age group (≥60 years) (Table 7). Further, age dependent normal values for men and women are presented in Figures 5 and 6.
Additional RV function parameters
Similar to the LV, Maceira et al. have provided additional functional parameters [16] (Table 8) that may have relevance to specific applications.
Left atrial dimensions and functions in the adult
CMR acquisition parameters
There is limited consensus in the literature about how to measure left atrial volumes. Therefore depending on the method that is used, SSFP sequences in different views are required. The most common methods to measure left atrial volume are the modified Simpson’s method analogous to the left and right ventricle and the biplane area-length method. Dedicated 3D-modeling software has also been used [17]. For evaluation by applying the Simpson’s method, a stack of cine SSFP images either in the short axis, the horizontal long axis or transverse view is required. For 3-dimensional modeling a stack of short axis images has been used [17]. Evaluation by the biplane area-length method is based on a 2 and 4 chamber view [4].
Left atrial longitudinal and transverse diameters and area have been measured on 2, 3, and 4 chamber cine SSFP images [17].
CMR analysis methods
Generally the left atrial appendage is included as part of the left atrial volume while the pulmonary veins are excluded [4,17,18].
The maximal left atrial volume is achieved during ventricular systole. Using cine images, the maximum volume can be defined as last image before opening of the mitral valve. Accordingly the minimal left atrial volume can be defined as first image after closure of the mitral valve [19].
Demographic parameters
Body surface area (BSA) has been shown to have a significant independent influence on left atrial volume and most diameters [17]. Per Sievers et al. [20], age is not an independent predictor of left atrial maximal volume [17] nor diameter in normal individuals. Men have a larger maximal left atrial volume compared to women [4,17].
Studies included in this review
There are three publications for reference values of the left atrium (volume and/or diameter and/or area) based on SSFP imaging with a sufficient sample size [4,17,20] (Table 9). Hudsmith et al. [4] used the biplane area-length method (Figure 7), while Maceira et al. [17] used a 3D modeling technique (Figure 8). Since the results for left atrial maximal volume differ substantially between the two publications, probably based on the different methods, these data are presented separately (Tables 10 and 11, respectively). Maceira et al. provide reference values for maximum left atrial volume, longitudinal, transverse and anteroposterior diameters as well as area (Tables 11, 12 and 13; Figures 8 and 9) [17]. Hudsmith evaluated normal values of maximal and minimal left atrial volume and calculated left atrial ejection fraction and left atrial stroke volume (Table 10) [4]. Sievers et al. provide reference values for left atrial transverse diameters measured on the 2-, 3- and 4-chamber view at ventricular end-systole [20]. Maceira et al. provide both transverse and longitudinal diameters with different 3-chamber methodology than Sievers et al., thus only diameters of Maceira et al. are included (Table 13) [17,20].
Example of contouring for the biplane area-length method from reference [ 4 ]. The left atrial appendage was included in the atrial volume and the pulmonary veins were excluded.
Contouring of the left and right atrium using a 3D modeling method according to reference [ 17 ].
Measurement of left atrial area (A2C, A4C, A3C), longitudinal (L2C, L4C), transverse (T2C, T4C) and anteroposterior (APD) diameters on the 2-, 4- and 3-chamber views according to reference [ 17 ].
Right atrial dimensions and functions in the adult
CMR acquisition parameters
There is no consensus in the literature regarding acquisition and measurement method for the right atrium. Published methods for right atrial volume include the modified Simpson’s method, the biplane area-length method and 3D-modeling [21,22]. Comparing the Simpsons method and the biplane area length method results in different values for right atrial volume [21]. For Simpson’s method and 3D modeling, a stack of cine SSFP images in the short axis view are analyzed. For the biplane area-length method, a 4-chamber view and/or a right ventricular 2-chamber view are evaluated.
CMR analysis methods
Generally the right atrial appendage is included in the right atrial volume while the inferior and superior vena cava are excluded [21,22].
The maximal right atrial volume is achieved during ventricular systole and can be defined as last cine image before opening of the tricuspid valve. The minimal left atrial volume can be defined as first cine image after closure of the tricuspid valve.
Demographic parameters
Maceira et al. demonstrated a significant independent influence of BSA on most RA parameters [22]. There was no influence of age on atrial parameters and no influence of gender on atrial volumes [21,22].
Studies included in this review
There are two publications of reference values for the right atrium (volume and/or diameter) based on SSFP imaging with a sufficient sample size [21,22] (Table 14). For evaluation of volume, Maceira et al. [22] used a 3D modeling technique (Figure 8) while Sievers et al. [21] applied the Simpsons and the biplane area-length methods, respectively. Due to different methodology, no pooled mean values are provided. Normal values for right atrial volume and function, diameter and area are presented in Tables 15, 16, 17, 18 and 19.
Left and right ventricular dimensions and function in children
The presentation of normal values in children is different than in the adult population due to continuous changes in body weight and height as a function of age. These changes may also be asymmetrical. Normal data in children is frequently presented in percentiles and/or z-scores. While the use of percentiles is a daily routine for the paediatric radiologist, the use of percentile data might be unfamiliar to the general radiologist. Therefore in the current review, normal values are presented as mean ± standard deviation as well as in percentiles.
Demographic parameters
A linear correlation between ventricular volumes and BSA in children has been reported. Ventricular volumes also vary by gender [23-25]. Ejection fraction remains constant during somatic growth and does not appear to be gender specific [23-25]. Gender differences are more marked in older children, indicating that gender is more important after puberty and in adulthood.
Studies included in this review
Normal values published in studies based on older gradient echo sequences are not comparable to current SSFP techniques [26,27]. Literature values for normative SSFP values have been proposed by three different groups acquired with slightly different methods [23-25] (Table 20). A good agreement between the three studies regarding the dimensions for older children has been demonstrated [25].
In the studies of Robbers-Visser et al. and Sarikouch et al., normal values for older children of 8–17 years and 4–20 years, respectively, are presented [23,24], Buechel et al. also include younger children starting with an age of 7 months. In the studies of Robbers-Visser et al. and Sarikouch et al. papillary muscle mass was included as part of LV mass, while Buchel et al. included papillary muscles in the LV cavity and provide separate values for papillary muscle mass. Robbers-Visser et al. and Sarikouch et al. present normal values as mean ± SD for all female and male children and for children of different age groups and also as percentiles. In the study by Buechel et al. data is presented in percentiles only.
Due to similar study design and age range, Tables 21 and 22 show pooled mean values for male and female children calculated based on mean values presented by Robbers-Visser et al. and Sarikouch et al. Figures 10,11,12 show normal data in percentiles originally published by Buechel et al.
Percentiles for left ventricular parameters in children according to reference [ 25 ].
Percentiles for left ventricular papillary muscle mass in children according to reference [ 25 ].
Percentiles for right ventricular parameters in children according to reference [ 25 ].
Z-values can be calculated as z-value = (measurement – expected mean)/SD by using the values presented in Tables 21 and 22.
Left and right atrial dimensions and function in children
CMR acquisition parameters
Left and right atrial dimensions and function were evaluated using SSFP technique in a single publication [18], (Table 23). Measurements were obtained on a stack of transverse cine SSFP images with a slice thickness between 5 and 6 mm without interslice gap [18].
CMR analysis methods
In that study, the pulmonary veins, the superior and inferior vena cava and the coronary sinus were excluded from the left and right atrial volume, respectively, while the atrial appendages were included in the volume of the respective atrium. The maximal atrial volume was measured at ventricular end-systole and the minimal atrial volume at ventricular end-diastole.
Demographic parameters
Left and right atrial volumes show an increase with age with a plateau after the age of 14 for girls only. Absolute and indexed volumes have been shown to be significantly greater for boys compared to girls (except for the indexed maximal volumes for both atria) [18].
Studies included in this review
Sarikouch et al. evaluated atrial parameters of 115 healthy children (Table 23) [18] using SSFP imaging. Since the standard deviation is large for each parameter, lower and upper limits were not calculated (Tables 24 and 25). Theoretically calculation of lower limits by mean – (2*SD) would result in negative lower limits for certain parameters.
Normal left ventricular myocardial thickness
CMR acquisition parameters
Normal values of left ventricular myocardial thickness (LVMT) have been shown to vary by type of pulse sequence (FGRE versus SSFP) [3,28]. For the purposes of this review, only SSFP normal values are shown.
CMR analysis methods
Measures of LVMT vary by the plane of acquisition (short axis versus long axis) [29]. Measurements obtained on long axis images at the basal and mid-cavity level have been shown to be significantly greater compared to measurements on corresponding short axis images, whereas measurements obtained at the apical level of long axis images are significantly lower compared to short axis images. In recent publications, papillary muscles and trabeculations were excluded from measurements of left ventricular myocardial thickness [29,30].
Demographic parameters
LVMT is greater in men than women [29,30]. There are also small differences in LVMT in relationship to ethnicity and body size, but these variations are not likely to have clinical significance [29]. Regarding age, one study of 120 healthy volunteers age 20–80 years reported an increase in myocardial thickness with age—starting after the fourth decade [30]. In the study by Kawel el al. of 300 normal individuals without hypertension, smoking history or diabetes, there was no statistically significant difference in LVMT with age [29].
Studies included in this review
There are two publications of a systematic analysis of left ventricular myocardial thickness based on SSFP imaging at 1.5T [29,30]. In the study by Dawson et al., measurements were obtained on short axis images only. Kawel et al. published normal values of LVMT for long and short axis imaging (Tables 26, 27 and 28).
Cardiac valves and quantification of flow
CMR acquisition parameters
Prospectively and retrospectively ECG- gated phase contrast (PC) CMR sequences are available on most CMR machines. Prospectively-gated sequences use arrhythmia rejection and may be performed in a breath hold. Retrospectively gated techniques are mainly performed during free-breathing, often with higher spatial and temporal resolution compared to the breath hold techniques [31]. 4D PC flow quantification techniques show initial promising results, but 2D PC flow techniques are currently used in the daily clinical routine [32]. Apart from PC-CMR, valve planimetry—using ECG-gated SSFP CMR— can also be used to estimate stenosis or insufficiencies with good correlation to echocardiographic measurements [33].
Measurements of flow are most precise when a) the imaging plane is positioned perpendicular to the vessel of interest and b) the velocity encoded gradient echo (Venc) is encoded in a through plane direction [34]. The slice thickness should be <7 mm to minimize partial volume effects. Compared to aortic or pulmonary flow evaluation, quantification of mitral or tricuspid valves is more challenging using PC-CMR due to substantial through plane motion during the cardiac cycle [35].
Flow encoding velocity (Venc)
The Venc should be chosen close to the maximum expected flow velocity of the examined vessel for precise measurements. Setting the Venc below the peak velocity results in aliasing. For the normal aorta and main pulmonary artery, maximum velocities do not exceed 150 and 90 cm/sec, respectively.
Adequate temporal resolution is necessary to avoid temporal flow averaging, especially for the evaluation of short, fast, and turbulent jets within a vessel (e.g. aortic stenosis). For the clinical routine, 25–30 msec temporal resolution is usually sufficient. The minimum required spatial resolution should be less than one third of the vessel diameter to avoid partial volume effects with the adjacent vessel wall and surrounding stationary tissues for small arteries [34].
CMR analysis methods
For data analysis, dedicated flow software should be used. Most of the currently available flow software tools offer semi-automatic vessel contouring, which needs to be carefully checked by the examiner.
The modified Bernoulli equation (∆P = 4 × Vmax 2) is commonly used for calculation of pressure gradients using PC-CMR across the pulmonary or aortic valve [36,37].
It has to be considered that velocity measurements of a stenotic lesion with high jet velocities might be inaccurate due to partial volume effects in case of a small jet width and also the limited temporal resolution compared to the high velocity of the jet. Measurements are further affected by signal loss due to the high velocity that may lead to phase shift errors and dephasing. Misalignement of the slice relative to the direction of the jet may lead to an underestimation of the peak velocity [38].
Demographic parameters
To our knowledge, data of the association between normal values of flow and valve planimetry with demographic parameters has not been previously published.
Studies included in this review
There is good agreement between PC-CMR, SSFP CMR planimetry, and echocardiography measurements, American Heart Association (AHA) criteria for grading valve stenosis or insufficiency is suggested [33,39,40] (Table 29). To date, there is no publication of normal reference values of flow and valve planimetry based on CMR measurements.
Mitral valve flow velocities and deceleration time as for determination of diastolic left ventricular function measured by CMR showed a good correlation with measurements derived by transthoracic echocardiography but with a systematic underestimation [42] (Table 30).
Normal aortic dimensions in the adult
CMR acquisition parameters
Three- dimensional contrast enhanced MR Angiography (MRA) has gained broad acceptance and is widely used for assessment and follow-up of thoracic aortic diameter in clinical setting. The multi-planar reformation of MRA images leads to an accurate measurement perpendicular to the lumen of the vessel. However, the need of a contrast injection is a limitation for the use of this technique in patients who need multiple follow up examinations and in population based study settings [44]. Alternatively non-contrast techniques such as ECG gated non contrast 3D (2D) balanced steady state free precession (SSFP) CMR can be applied. The modulus image of phase contrast CMR has also been used to measure diameters of the aorta [45]. 2D Black blood CMR is used for a more detailed aortic wall assessment. In 2D acquisitions, the imaging plane needs to be acquired correctly at the time of the scan; thus any alterations in the imaging plane will result in a higher variability and lower accuracy of measurements. Another limitation for ascending thoracic aorta diameter measurement is the through plane motion during the cardiac cycle which can be minimized with ECG gating [44]. Potthast and colleagues compared the diameter of the ascending aorta obtained by different CMR sequences to ECG-triggered CT angiography as the gold standard and reported that ECG gated navigator triggered 3D SSFP sequence showed the best agreement with CT [44].
CMR analysis methods
It is important to indentify the anatomic locations of diameter measurements of the thoracic aorta. In the studies cited here, measurements were obtained at the following anatomic locations: 1. aortic root cusp-commissure and cusp-cusp measurements; 2. aortic valve annulus; 3. aortic sinus; 4. sinotubular junction; 5. ascending aorta and proximal descending aorta: measurements at the level of the right pulmonary artery; 6. abdominal aorta: 12 cm distal to the pulmonary artery (Figure 13).
The sagittal oblique view of the left ventricular outflow tract was used for measuring diameter at the level of the aortic annulus, the aortic sinus, and the sinotubular junction. Axial cross sectional images at predefined anatomic levels were used for measuring the ascending and descending aorta [46] as well as cusp-commissure and cusp-cusp diameters at the level of the aortic sinus [47] (Figures 13 and 14). There is no definite convention about measuring the luminal or outer to outer diameter of the aorta. Usually, measurement technique depends on the resolution and characteristics of the available MRI sequence. In the tables below, the method is specified.
Demographic parameters
Age, gender and body size are major determinants of physiologic variation in aortic size. In the Multi-Ethnic Study of Atherosclerosis, which included participants from four different ethnicities, the race/ethnicity were not clinically significant determinants of ascending aorta diameter [45].
Aortic diameter and ascending aorta length increase with age, leading to decreased curvature of the aortic arch [48,49]. The association of age with aortic diameter was more marked in the ascending aorta compared to the descending thoracic and abdominal aorta, respectively [50,51]. Additionally, the descending aorta did not demonstrate age associated lengthening [49].
Studies included in this review
Studies with normal values of aortic diameters including 50 or more subjects for both men and women and a range of ages (due to the age dependence of aortic diameters) have been included in this review (Tables 31, 32, 33, 34 and 35). There are three major publications regarding MR-based measurements of the thoracic aorta in adults: Davis et al. determined aortic diameter at three levels (ascending aorta, proximal descending aorta and abdominal aorta) by calculating the luminal diameter based on measurements of the cross sectional area obtained on cine SSFP images [46]. In the original publication normal age and gender specific absolute and indexed (for BSA) values are presented in a graph and absolute numbers are presented for different weight categories (Table 32).
Turkbey et al. measured the luminal diameter of the ascending aorta on magnitude images of a phase contrast sequence in a large number of healthy subjects [45] (Table 33).
Burmann et al. performed detailed measurements of the aortic root including cusp-commissure and cusp-cusp measurements at diastole and systole on cine SSFP images [47] (Tables 34 and 35).
Normal aortic dimensions in children
CMR acquisition parameters
There is no consensus regarding the sequence type used to measure aortic diameters and areas. In the three major publications (Table 36) measurements were obtained on 3 dimensional contrast enhanced MR angiography [52], gradient echo images [53] and phase contrast cine images [54].
CMR analysis methods
In order to reduce error in measurement, care has to be taken to obtain or reconstruct cross sectional images that are true perpendicular instead of oblique to the course of the vessel. Kaiser et al. demonstrated that aortic diameter measurements show a slight variation with measurement plane with a mean difference between measurements on cross-sectional and longitudinal images of 0.16 mm and a coefficient of variability of 2.1% [52].
Kutty et al. indicate that in their study the outer diameter of the vessel was measured [54] while Kaiser et al. and Voges et al. do not mention details in this respect [52,53].
Demographic parameters
Aortic diameters vary by BSA [52,54] but do not show gender differences [53,54]. Aortic area did also not show any gender differences [53].
Studies included in this review
There are three publications of a systematic evaluation of aortic dimensions (diameter and/or area) in children that vary by CMR-technique, measurement technique and data presentation (Table 36): In the study by Kaiser et al. aortic diameter was measured as the shortest diameter passing the center of the vessel at 9 levels of the thoracic aorta on reconstructed cross-sectional images of a contrast enhanced 3 dimensional MR angiography [52]. In the original publication data is presented as median and range as well as percentiles, z-scores and regression models incorporating BSA. Voges et al. present measurements obtained at four levels of the thoracic aorta obtained on cine GRE images at maximal aortic distension as mean ± standard deviation and as percentiles [53]. In the study by Kutty et al. aortic diameter and area was measured 1-2 cm distal to the sinotubular junction at systole on phase contrast cine images [54]. Data is presented as mean ± standard deviation, regression equation and z-scores.
In this review we present regression equations of normal aortic diameters measured at 9 different sites according to [52] (Table 37, Figure 15) and of normal area of the ascending aorta according to [54] (Table 38). Further reference percentiles of aortic area measured at 4 different locations obtained on cine GRE images are presented in Figure 16 according to [53]. The z-scores for each aortic diameter (D) can be calculated with the following equation:
Sites of measurement. AS = aortic sinus; STJ = sinotubular junction; AA = ascending aorta; BCA = proximal to the origin of the brachiocephalic artery; T1 = first transverse segment; T2 = second transverse segment; IR = isthmic region; DA = descending aorta; D = thoracoabdominal aorta at the level of the diaphragm.
Reference percentiles for aortic areas measured at four different sites (ascending aorta [a], aortic arch [b], aortic isthmus [c] and descending aorta above the diaphragm [d]) on cine GRE images at maximal aortic distension according to reference [ 53 ].
on the base of the data provided in Table 37.
Due to the differences in acquisition and measurement technique as well as presentation of results, weighted mean values were not calculated.
Normal aortic distensibility/ pulse wave velocity in adults
CMR acquisition parameters
Pulse wave velocity (PWV) calculations using a velocity-encoded CMR with phase contrast sequences allow accurate assessment of aortic systolic flow wave and the blood flow velocity. The sequence should be acquired at the level of the bifurcation of the pulmonary trunk, perpendicular to both, the ascending and descending aorta. The distance between two aortic locations (aortic length) can be estimated from axial and coronal cine breath hold SSFP sequences covering the whole aortic arch [55]. Alternatively, sagittal oblique views of the aortic arch can be acquired using a black blood spin echo sequence [51].
Another measurement method of aortic stiffness is aortic distensibility. The cross sectional aortic area at different phases of the cardiac cycle is measured using ECG-gated SSFP cine imaging to assess aortic distensibility by CMR. Modulus images of cine phase contrast CMR can be used as well [56].
CMR analysis methods
PWV is the most validated method to quantify arterial stiffness using CMR. PWV is calculated by measuring the pulse transit time of the flow curves (Δt) and the distance (D) between the ascending and descending aortic locations of the phase contrast acquisition [51]: Aortic PWV = D/ Δt (Figure 17).
Measurement of pulse wave velocity according to reference [ 53 ] . Δx = length of the centerline between the sites of flow measurement in the ascending and descending aorta (A); Δt = time delay between the flow curves obtained in the descending aorta relative to the flow curve obtained in the ascending aorta calculated between the midpoint of the systolic up slope tails on the flow versus time curves of the ascending aorta (ta1) and the descending aorta (ta2) (B).
PWV increases with stiffening of arteries since the stiffened artery conducts the pulse wave faster compared to more distensible arteries.
Aortic distensibility is calculated with the fallowing formula after measuring the minimum and maximum aortic cross sectional area [57]:
Aortic Distensibility = (minimum area- maximum area)/ (minimum area x ΔP x 1000) where ΔP is the pulse pressure in mmHg.
Demographic parameters
Greater ascending aorta diameter and changes in aortic arch geometry by aging was significantly associated with increased regional stiffness of the aorta, especially the ascending portion. The relationship of age with measures of aortic stiffness is non –linear and the decrease of aortic distensibility is steeper before the fifth decade of life [51]. Males have stiffer aortas compared to females [58].
Studies included in this review
Two publications reported normal values of pulse wave velocity and aortic distensibility (Tables 39, 40 and 41).
Normal aortic distensibility/ pulse wave velocity in children
CMR acquisition parameters
In the only publication of aortic distensibility and pulse wave velocity in children, distensibility was measured on gradient echo cine CMR images and pulse wave velocity was measured on phase-contrast cine CMR [53].
CMR analysis methods
Distensibility was calculated as (Amax – Amin)/Amin x (Pmax – Pmin), where Amax and Amin represent the maximal and minimal cross-sectional areal of the aorta, and Pmax and Pmin represent the systolic and diastolic blood pressure measured with a sphygmomanometer cuff around the right arm.
Pulse wave velocity was calculated as Δx/Δt, where Δx is defined as the length of the centerline between the sites of flow measurement in the ascending and descending aorta and Δt represents the time delay between the flow curve obtained in the descending aorta relative to the flow curve obtained in the ascending aorta (Figure 17).
Demographic parameters
Aortic distensibility and pulse wave velocity did not vary by gender. Aortic distensibility decreases with age and correlates with height, body weight and BSA [53].
Studies included in this review
There is a single publication only of a systematic evaluation of normal aortic distensibility and pulse wave velocity in children (Table 42). Reference percentiles by age according to reference [53] are presented in Figures 18 and 19.
Reference percentiles for aortic distensibility measured at four different sites (ascending aorta [a], aortic arch [b], aortic isthmus [c] and descending aorta above the diaphragm [d]) on cine GRE images at maximal aortic distension according to reference [ 53 ].
Reference percentiles for aortic pulse wave velocity according to reference [ 53 ].
Normal values of myocardial T1 relaxation time and the extracellular volume (ECV)
CMR acquisition parameters
Most of the published myocardial T1 values have been acquired using a Modified Look-Locker Inversion Recovery (MOLLI) [59] or shortened-MOLLI (ShMOLLI) [60] method, combined with a balanced SSFP read out [59]. The MOLLI method acquires data over 17 heartbeats with a 3(3bt)3(3bt)5 sampling pattern, while ShMOLLI has been described with a 9 heart beat breath-hold and a 5(1bt)1(1bt)1 sampling pattern, although variations of these acquisition schemes have been proposed [60]. An alternative sampling method is saturation recovery single-shot acquisition (SASHA) in which a first single-shot bSSFP image is acquired without magnetization preparation followed by nine images prepared with variable saturation recovery times [61]. All methods usually acquire images at end diastole to limit cardiac motion artifacts [59] but acquisition of T1 maps at systole has been shown to be feasible [62]. Post contrast T1 values have been performed following a bolus or primed infusion (Equilibrium-EQCMR) with good agreement of ECV values up to 40% [63].
Factors affecting T1 and ECV
Field strength has a significant effect on T1 values; with 3T scans producing 28% higher native T1 and 14% higher post contrast T1 values when compared with 1.5T [62]. Post contrast T1 is also affected by the dose and relaxivity of the contrast agent used, contrast clearance, and the time between injection and measurement [62,64,65]. There is also greater heterogeneity for a T1 native normal range at 3 Tesla [62,66,67]. Further, it has been shown that T1 varies by cardiac phase (diastole versus systole) and region of measurement (septal versus non-septal) [62]. ECV values are relatively unaffected by field strength (3T versus 1.5T). Both native T1 and ECV values have been shown to be less reliable in the infero-lateral wall [62,68].
Flip-angle and pre-pulse can also affect normal values, with the adiabatic pre-pulse increasing native T1 values by approximately 25 ms compared with non-adiabatic pre-pulses. FLASH mapping sequences produce significantly lower native T1 values than bSSFP methods [69,70].
CMR analysis methods
T1 maps are based on pixel-wise quantification of longitudinal relaxation of the acquired images. Native T1 measures a composite signal from myocytes and interstitium and is expressed in ms [71]. Measurements that correlate pre and post contrast T1 myocardial values and blood T1 have been proposed, such as the partition coefficient or the extracellular volume fraction (ECV), expressed as a percantage [72].
Offline post-processing involves manually tracing endocardial and epicardial contours [65,73] (Figure 20) or placing a region of interest within the septal myocardium using a prototype tool [67]. Inclusion of blood pool or adjacent tissue should be carefully avoided. Motion correction is generally used to counter undesired breathing motion. However, motion correction can only correct for in-plane motion and not through-plane motion. All methods, therefore, are vulnerable to partial volume effects. Some investigators also corrected for this with smaller regions of interest and co-registration of images [74].
Demographic parameters
Increasing age has been shown to increase ECV in healthy volunteers in one publication [68] and females less than 45 years have been shown to have a higher precontrast T1 [74].
Studies included in this review
Several studies have shown a strong correlation between T1 values by CMR and diffuse myocardial fibrosis on myocardial biopsy [71,75,76], but the rapid evolution of acquisition methods over the past years has led to inconsistent T1 values reported in the literature.
In order to reflect the current literature, the normal values presented here are classified by field strength and acquisition pulse sequence and list pre contrast, post contrast, and ECV values where available.
It should be noted that a universal normal range for T1 cannot be determined given the heterogeneity of acquisition pulse sequences used in the existing literature and because no true reference value for in vivo T1 exists. Table 43 is a summary of publications over the last years presenting normal values based on ≥ 20 healthy subjects as available in December 2013.
For SASHA, only limited normal values are available. T1 estimates based on SASHA are higher than with MOLLI methods. One study reported SASHA derived T1 values in 39 normal subjects of 1170 ± 9 ms at 1.5T [61].
We conclude that at present, normal native T1 values are specific to pulse sequences and scanner manufacturer. For diagnostic purposes it is most important to use a method with a tight normal range, good reproducibility and sensitivity to disease.
Normal values of myocardial T2* relaxation time
CMR acquisition parameters
Quantification of the T2* relaxation time plays an important role for estimation of myocardial iron overload [81]. For quantification of the myocardial T2* time, the gradient-echo T2* technique with multiple increasing TEs is preferred over the spin-echo T2 technique due to a greater sensitivity to iron deposition [82-84]. Usually a single-breath hold technique is used. Normal values and a grading system for myocardial iron overload are available for 1.5T [84].
CMR analysis methods
Since the gradient-echo T2* technique is vulnerable to distortions of the local magnetic field e.g. by air-tissue interfaces, measurements are obtained by placing a region of interest on the interventricular septum of a midventricular short axis slice, since the septum is surrounded by blood on both sides [83] (Figure 21).
T2* times are frequently reported as relaxation rate, representing the reciprocal of the time constant and calculated as R2* = 1000/T2*. The unit of R2* is Hertz or s−1 [83].
Demographic parameters
It has been shown that T2* does not correlate with age [85]. To our knowledge the relationship between other demographic parameters and T2* has not been assessed yet.
Studies included in this review
Generally a T2* value - measured at the interventricular septum using a multiecho GRE sequence at 1.5T - of >20 ms is considered normal while the mean myocardial T2* is around 40 ms [81].
Examples of studies that used the current multiecho GRE technique with a sample size of >10 healthy subjects are presented in Table 44.
Depending on the risk to develop heart failure as a consequence of myocardial iron overload, a grading system for disease severity has been published (Table 45) [87].
Regional Measurements and Cardiac Strain
CMR acquisition parameters
A number of imaging methods have been developed to acquire cardiac strain information from CMR including cine CMR, tagged MR, phase-contrast CMR (PC-CMR), velocity encoded CMR, displacement encoding with stimulated echoes (DENSE), and strain-encoding (SENC) [88,89]. However, tagged CMR remains a widely validated reproducible tool for strain estimation. The method is used in clinical studies and is considered the reference standard for assessing regional function [90,91].
CMR analysis methods
Cardiac strain is a dimensionless measurement of the deformation that occurs in the myocardium. Cardiac strain can be reported as three normal strains (circumferential, radial, and longitudinal) and six shear strains—the angular change between two originally mutually orthogonal line elements, with the more clinically investigated shear strain in the circumferential-longitudinal shear strain (also known as torsion).
There are a number of different methods to quantify strain: registration methods, feature-based tracking methods, deformable models, Gabor Filter Banks, optic flow methods, harmonic phase analysis (HARP) [92], and local sine wave modeling (SinMod) [88]. Technical review papers for these methods can be found in the following literature [93-96]. HARP has become one of the most widely used methods for analyzing tagged MR images for cardiac strain, in part due to its large scale use in the Multi-Ethnic Study of Atherosclerosis (MESA) trial [92,97].
Strain patterns are reported according to the 16 and 17 segment model of the American College of Echocardiology. Consistent manual tracing of the endocardial and epicardial contours is necessary to reproducible strain results. With tagged CMR, midwall strain is preferred to epicardial and endocardial strain to maximize the amount of tagging data available for strain calculations [95,98]. With HARP analysis such as that used in the MESA trial [92], careful selection of the first harmonic is necessary.
Demographic parameters
With tagged CMR, it has been noted that age is associated with decrease in peak circumferential or longitudinal shortening [99,100]. In tagged CMR studies, gender also affects normal values. Cardiac strain values for women are higher than those of men [101,102].
Studies included in this review
Several studies have presented cohorts of normal individuals for determining normal strain of the left ventricle. For the purpose of this review, only cohorts of 30 or more normal subjects using SPAMM tagging have been included. (Feature tracking methods are being developed for strain, but are being validation in comparison to SPAMM tagging.) Inclusion criteria include a full description of the subject cohort (including the analysis methods used), age and gender of subjects. Table 46 represents a summary of publications reporting normal values for midwall strain that fit the criteria [89,90,98].
With tagged CMR, normal midwall circumferential [89,92,93,98,101,104] and longitudinal [89,96,98,104] strain values are relatively comparable between studies. For radiation strain, moderate differences between published results exist for reference values, probably due to low tag density by most methods in the radial direction [90,96-98,101,104].
Conclusions
Cardiovascular magnetic resonance enables quantification of various functional and morphological parameters of the cardiovascular system. This review lists reference values and their influencing factors of these parameters based on current CMR techniques and sequences.
Advantages of a quantitative evaluation are a better differentiation between pathology and normal conditions, grading of pathologies, monitoring changes under therapy, and evaluating prognosis and the possibility of comparing different groups of patients and normal subjects.
Abbreviations
- CMR:
-
Cardiovascular magnetic resonance
- SSFP:
-
Steady-state free precession
- LV:
-
Left ventricle
- Yrs:
-
Years
- FGRE:
-
Fast gradient echo
- EDV:
-
End-diastolic volume
- BSA:
-
Body surface area
- ESV:
-
End-systolic volume
- SV:
-
Stroke volume
- EF:
-
Ejection fraction
- SD:
-
Standard deviation
- SDp :
-
Pooled standard deviation
- Meanp :
-
Pooled weighted mean
- RV:
-
Right ventricle
- PFRE:
-
Peak filling rate early
- PFRA:
-
Peak filling rate active
- AVPD:
-
Atrioventricular plane descent
- 3D:
-
3-dimensional
- CI:
-
Confidence interval
- RA:
-
Right atrium
- LA:
-
Left atrium
- AP:
-
Anteroposterior
- 4ch:
-
4-chamber view
- 2ch:
-
2-chamber view
- 3ch:
-
3-chamber view
- Max.:
-
Maximal
- Min.:
-
Minimal
- CI:
-
Cardiac index
- LVMT:
-
Left ventricular myocardial thickness
- SVC:
-
Superior vena cava
- IVC:
-
Inferior vena cava
- Vol.:
-
Volume
- PC:
-
Phase contrast
- 4D:
-
4-dimensional
- Venc :
-
Flow encoding velocity
- AHA:
-
American Heart Association
- MRA:
-
Magnetic resonance angiography
- MDT:
-
Mitral deceleration time
- E/A ratio:
-
Ratio of the mitral early (E) and atrial (A) components of the mitral inflow velocity profile
- CT:
-
Computed tomography
- BMI:
-
Body mass index
- Prox.:
-
Proximal
- Decs.:
-
Descending
- AS:
-
Aortic sinus
- STJ:
-
Sinotubular junction
- AA:
-
Ascending aorta
- BCA:
-
Proximal to the origin of the brachiocephalic artery
- IR:
-
Isthmic region
- DA:
-
Descending aorta
- PWV:
-
Pulse wave velocity
- ECV:
-
Extracellular volume
- MOLLI:
-
Modified Look-Locker Inversion Recovery
- ShMOLLI:
-
Shortened Modified Look-Locker Inversion Recovery
- SASHA:
-
Saturation Recovery Single-Shot Acquisition
- EQ-CMR:
-
Equilibrium cardiovascular magnetic resonance
- FLASH:
-
Fast low angle shot
- G(R)E:
-
Gradient echo
- BH:
-
Breath hold
- VAST:
-
Variable sampling of k-space
- LL:
-
Look-Locker
- Hb:
-
Heart beat
- SAX:
-
Short axis
- DENSE:
-
Displacement encoding with stimulated echoes
- SENC:
-
Strain encoding
- HARP:
-
Harmonic phase analysis
- SinMod:
-
Sine wave modeling
- MESA:
-
Multi-Ethinic Study of Atherosclerosis
- Resol:
-
Resolution
- SPAMM:
-
Spatial modulation of the magnetization
- mid:
-
Mid cavity
- ap:
-
Apical level
- sept:
-
Septal
- ant:
-
Anterior
- lat:
-
Lateral
- inf:
-
Inferior
- FS:
-
Field strength
References
Harris RJ BM, Deeks JJ, Harbord RM, Altman DG, Sterne JAC. Metan: fixed- and random-effects meta-analysis. Stata J. 2008;8:3–28.
Selvin S. Statistical analysis of epidemiological data. 3rd ed. New York, USA: Oxford University Press; 2004.
Alfakih K, Plein S, Thiele H, Jones T, Ridgway JP, Sivananthan MU. Normal human left and right ventricular dimensions for MRI as assessed by turbo gradient echo and steady-state free precession imaging sequences. J Magn Reson Imaging. 2003;17:323–9.
Hudsmith LE, Petersen SE, Francis JM, Robson MD, Neubauer S. Normal human left and right ventricular and left atrial dimensions using steady state free precession magnetic resonance imaging. J Cardiovasc Magn Reson. 2005;7:775–82.
Maceira AM, Prasad SK, Khan M, Pennell DJ. Normalized left ventricular systolic and diastolic function by steady state free precession cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2006;8:417–26.
Vogel-Claussen J, Finn JP, Gomes AS, Hundley GW, Jerosch-Herold M, Pearson G, et al. Left ventricular papillary muscle mass: relationship to left ventricular mass and volumes by magnetic resonance imaging. J Comput Assist Tomogr. 2006;30:426–32.
Schulz-Menger J, Bluemke DA, Bremerich J, Flamm SD, Fogel MA, Friedrich MG, et al. Standardized image interpretation and post processing in cardiovascular magnetic resonance: Society for Cardiovascular Magnetic Resonance (SCMR) board of trustees task force on standardized post processing. J Cardiovasc Magn Reson. 2013;15:35.
Reiter G, Reiter U, Rienmuller R, Gagarina N, Ryabikin A. On the value of geometry-based models for left ventricular volumetry in magnetic resonance imaging and electron beam tomography: a Bland-Altman analysis. Eur J Radiol. 2004;52:110–8.
Clay S, Alfakih K, Radjenovic A, Jones T, Ridgway JP, Sinvananthan MU. Normal range of human left ventricular volumes and mass using steady state free precession MRI in the radial long axis orientation. MAGMA. 2006;19:41–5.
Cain PA, Ahl R, Hedstrom E, Ugander M, Allansdotter-Johnsson A, Friberg P, et al. Age and gender specific normal values of left ventricular mass, volume and function for gradient echo magnetic resonance imaging: a cross sectional study. BMC Med Imaging. 2009;9:2.
Natori S, Lai S, Finn JP, Gomes AS, Hundley WG, Jerosch-Herold M, et al. Cardiovascular function in multi-ethnic study of atherosclerosis: normal values by age, sex, and ethnicity. AJR Am J Roentgenol. 2006;186:S357–365.
Salton CJ, Chuang ML, O'Donnell CJ, Kupka MJ, Larson MG, Kissinger KV, et al. Gender differences and normal left ventricular anatomy in an adult population free of hypertension. A cardiovascular magnetic resonance study of the Framingham Heart Study Offspring cohort. J Am Coll Cardiol. 2002;39:1055–60.
Lorenz CH, Walker ES, Morgan VL, Klein SS, Graham Jr TP. Normal human right and left ventricular mass, systolic function, and gender differences by cine magnetic resonance imaging. J Cardiovasc Magn Reson. 1999;1:7–21.
Sievers B, Kirchberg S, Bakan A, Franken U, Trappe HJ. Impact of papillary muscles in ventricular volume and ejection fraction assessment by cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2004;6:9–16.
Winter MM, Bernink FJ, Groenink M, Bouma BJ, van Dijk AP, Helbing WA, et al. Evaluating the systemic right ventricle by CMR: the importance of consistent and reproducible delineation of the cavity. J Cardiovasc Magn Reson. 2008;10:40.
Maceira AM, Prasad SK, Khan M, Pennell DJ. Reference right ventricular systolic and diastolic function normalized to age, gender and body surface area from steady-state free precession cardiovascular magnetic resonance. Eur Heart J. 2006;27:2879–88.
Maceira AM, Cosin-Sales J, Roughton M, Prasad SK, Pennell DJ. Reference left atrial dimensions and volumes by steady state free precession cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2010;12:65.
Sarikouch S, Koerperich H, Boethig D, Peters B, Lotz J, Gutberlet M, et al. Reference values for atrial size and function in children and young adults by cardiac MR: a study of the German competence network congenital heart defects. J Magn Reson Imaging. 2011;33:1028–39.
Rohner A, Brinkert M, Kawel N, Buechel RR, Leibundgut G, Grize L, et al. Functional assessment of the left atrium by real-time three-dimensional echocardiography using a novel dedicated analysis tool: initial validation studies in comparison with computed tomography. Eur J Echocardiogr. 2011;12:497–505.
Sievers B, Kirchberg S, Franken U, Bakan A, Addo M, John-Puthenveettil B, et al. Determination of normal gender-specific left atrial dimensions by cardiovascular magnetic resonance imaging. J Cardiovasc Magn Reson. 2005;7:677–83.
Sievers B, Addo M, Breuckmann F, Barkhausen J, Erbel R. Reference right atrial function determined by steady-state free precession cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2007;9:807–14.
Maceira AM, Cosin-Sales J, Roughton M, Prasad SK, Pennell DJ. Reference right atrial dimensions and volume estimation by steady state free precession cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2013;15:29.
Robbers-Visser D, Boersma E, Helbing WA. Normal biventricular function, volumes, and mass in children aged 8 to 17 years. J Magn Reson Imaging. 2009;29:552–9.
Sarikouch S, Peters B, Gutberlet M, Leismann B, Kelter-Kloepping A, Koerperich H, et al. Sex-specific pediatric percentiles for ventricular size and mass as reference values for cardiac MRI: assessment by steady-state free-precession and phase-contrast MRI flow. Circ Cardiovasc Imaging. 2010;3:65–76.
Buechel EV, Kaiser T, Jackson C, Schmitz A, Kellenberger CJ. Normal right- and left ventricular volumes and myocardial mass in children measured by steady state free precession cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2009;11:19.
Helbing WA, Rebergen SA, Maliepaard C, Hansen B, Ottenkamp J, Reiber JH, et al. Quantification of right ventricular function with magnetic resonance imaging in children with normal hearts and with congenital heart disease. Am Heart J. 1995;130:828–37.
Lorenz CH. The range of normal values of cardiovascular structures in infants, children, and adolescents measured by magnetic resonance imaging. Pediatr Cardiol. 2000;21:37–46.
Malayeri AA, Johnson WC, Macedo R, Bathon J, Lima JA, Bluemke DA. Cardiac cine MRI: Quantification of the relationship between fast gradient echo and steady-state free precession for determination of myocardial mass and volumes. J Magn Reson Imaging. 2008;28:60–6.
Kawel N, Turkbey EB, Carr JJ, Eng J, Gomes AS, Hundley WG, et al. Normal left ventricular myocardial thickness for middle-aged and older subjects with steady-state free precession cardiac magnetic resonance: the multi-ethnic study of atherosclerosis. Circ Cardiovasc Imaging. 2012;5:500–8.
Dawson DK, Maceira AM, Raj VJ, Graham C, Pennell DJ, Kilner PJ. Regional thicknesses and thickening of compacted and trabeculated myocardial layers of the normal left ventricle studied by cardiovascular magnetic resonance. Circ Cardiovasc Imaging. 2011;4:139–46.
Sondergaard L, Stahlberg F, Thomsen C, Spraggins TA, Gymoese E, Malmgren L, et al. Comparison between retrospective gating and ECG triggering in magnetic resonance velocity mapping. Magn Reson Imaging. 1993;11:533–7.
Allen BD, Barker AJ, Carr JC, Silverberg RA, Markl M. Time-resolved three-dimensional phase contrast MRI evaluation of bicuspid aortic valve and coarctation of the aorta. Eur Heart J Cardiovasc Imaging. 2013;14:399.
Kupfahl C, Honold M, Meinhardt G, Vogelsberg H, Wagner A, Mahrholdt H, et al. Evaluation of aortic stenosis by cardiovascular magnetic resonance imaging: comparison with established routine clinical techniques. Heart. 2004;90:893–901.
Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics. 2002;22:651–71.
Srichai MB, Lim RP, Wong S, Lee VS. Cardiovascular applications of phase-contrast MRI. AJR Am J Roentgenol. 2009;192:662–75.
Caruthers SD, Lin SJ, Brown P, Watkins MP, Williams TA, Lehr KA, et al. Practical value of cardiac magnetic resonance imaging for clinical quantification of aortic valve stenosis: comparison with echocardiography. Circulation. 2003;108:2236–43.
Kilner PJ, Manzara CC, Mohiaddin RH, Pennell DJ, Sutton MG, Firmin DN, et al. Magnetic resonance jet velocity mapping in mitral and aortic valve stenosis. Circulation. 1993;87:1239–48.
Myerson SG. Heart valve disease: investigation by cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2012;14:7.
Bonow RO, Carabello BA, Chatterjee K, de Leon Jr AC, Faxon DP, Freed MD, et al. 2008 Focused update incorporated into the ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the 1998 Guidelines for the Management of Patients With Valvular Heart Disease): endorsed by the Society of Cardiovascular Anesthesiologists, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons. Circulation. 2008;118:e523–661.
Bonow RO, Carabello BA, Kanu C, de Leon Jr AC, Faxon DP, Freed MD, et al. ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (writing committee to revise the 1998 Guidelines for the Management of Patients With Valvular Heart Disease): developed in collaboration with the Society of Cardiovascular Anesthesiologists: endorsed by the Society for Cardiovascular Angiography and Interventions and the Society of Thoracic Surgeons. Circulation. 2006;114:e84–231.
Manning WJ, Pennell DJ. Cardiovascular magnetic resonance. In: Book Cardiovascular magnetic resonance. 2nd ed. Philadelphia, USA: Saunders/Elsevier; 2010.
Rathi VK, Doyle M, Yamrozik J, Williams RB, Caruppannan K, Truman C, et al. Routine evaluation of left ventricular diastolic function by cardiovascular magnetic resonance: a practical approach. J Cardiovasc Magn Reson. 2008;10:36.
Caudron J, Fares J, Bauer F, Dacher JN. Evaluation of left ventricular diastolic function with cardiac MR imaging. Radiographics. 2011;31:239–59.
Potthast S, Mitsumori L, Stanescu LA, Richardson ML, Branch K, Dubinsky TJ, et al. Measuring aortic diameter with different MR techniques: comparison of three-dimensional (3D) navigated steady-state free-precession (SSFP), 3D contrast-enhanced magnetic resonance angiography (CE-MRA), 2D T2 black blood, and 2D cine SSFP. J Magn Reson Imaging. 2010;31:177–84.
Turkbey EB, Jain A, Johnson C, Redheuil A, Arai AE, Gomes AS, et al. Determinants and normal values of ascending aortic diameter by age, gender, and race/ethnicity in the Multi-Ethnic Study of Atherosclerosis (MESA). J Magn Reson Imaging. 2014;39:360–8.
Davis AE, Lewandowski AJ, Holloway CJ, Ntusi NA, Banerjee R, Nethononda R, et al. Observational study of regional aortic size referenced to body size: production of a cardiovascular magnetic resonance nomogram. J Cardiovasc Magn Reson. 2014;16:9.
Burman ED, Keegan J, Kilner PJ. Aortic root measurement by cardiovascular magnetic resonance: specification of planes and lines of measurement and corresponding normal values. Circ Cardiovasc Imaging. 2008;1:104–13.
Redheuil A, Yu WC, Mousseaux E, Harouni AA, Kachenoura N, Wu CO, et al. Age-related changes in aortic arch geometry: relationship with proximal aortic function and left ventricular mass and remodeling. J Am Coll Cardiol. 2011;58:1262–70.
Sugawara J, Hayashi K, Yokoi T, Tanaka H. Age-associated elongation of the ascending aorta in adults. JACC Cardiovasc Imaging. 2008;1:739–48.
Lederle FA, Johnson GR, Wilson SE, Chute EP, Littooy FN, Bandyk D, et al. Prevalence and associations of abdominal aortic aneurysm detected through screening. Aneurysm Detection and Management (ADAM) Veterans Affairs Cooperative Study Group. Ann Intern Med. 1997;126:441–9.
Redheuil A, Yu WC, Wu CO, Mousseaux E, de Cesare A, Yan R, et al. Reduced ascending aortic strain and distensibility: earliest manifestations of vascular aging in humans. Hypertension. 2010;55:319–26.
Kaiser T, Kellenberger CJ, Albisetti M, Bergstrasser E, Valsangiacomo Buechel ER. Normal values for aortic diameters in children and adolescents–assessment in vivo by contrast-enhanced CMR-angiography. J Cardiovasc Magn Reson. 2008;10:56.
Voges I, Jerosch-Herold M, Hedderich J, Pardun E, Hart C, Gabbert DD, et al. Normal values of aortic dimensions, distensibility, and pulse wave velocity in children and young adults: a cross-sectional study. J Cardiovasc Magn Reson. 2012;14:77.
Kutty S, Kuehne T, Gribben P, Reed E, Li L, Danford DA, et al. Ascending aortic and main pulmonary artery areas derived from cardiovascular magnetic resonance as reference values for normal subjects and repaired tetralogy of Fallot. Circ Cardiovasc Imaging. 2012;5:644–51.
Dogui A, Redheuil A, Lefort M, DeCesare A, Kachenoura N, Herment A, et al. Measurement of aortic arch pulse wave velocity in cardiovascular MR: comparison of transit time estimators and description of a new approach. J Magn Reson Imaging. 2011;33:1321–9.
Turkbey EB, Redheuil A, Backlund JY, Small AC, Cleary PA, Lachin JM, et al. Complications Trial/Epidemiology of Diabetes I, Complications Research G: Aortic distensibility in type 1 diabetes. Diabetes Care. 2013;36:2380–7.
Cavalcante JL, Lima JA, Redheuil A, Al-Mallah MH. Aortic stiffness: current understanding and future directions. J Am Coll Cardiol. 2011;57:1511–22.
Rose JL, Lalande A, Bouchot O, el Bourennane B, Walker PM, Ugolini P, et al. Influence of age and sex on aortic distensibility assessed by MRI in healthy subjects. Magn Reson Imaging. 2010;28:255–63.
Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP. Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med. 2004;52:141–6.
Piechnik SK, Ferreira VM, Dall'Armellina E, Cochlin LE, Greiser A, Neubauer S, et al. Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson. 2010;12:69.
Chow K, Flewitt JA, Green JD, Pagano JJ, Friedrich MG, Thompson RB. Saturation recovery single-shot acquisition (SASHA) for myocardial T(1) mapping. Magn Reson Med. 2014;71:2082–95.
Kawel N, Nacif M, Zavodni A, Jones J, Liu S, Sibley CT, et al. T1 mapping of the myocardium: intra-individual assessment of the effect of field strength, cardiac cycle and variation by myocardial region. J Cardiovasc Magn Reson. 2012;14:27.
Schelbert EB, Testa SM, Meier CG, Ceyrolles WJ, Levenson JE, Blair AJ, et al. Myocardial extravascular extracellular volume fraction measurement by gadolinium cardiovascular magnetic resonance in humans: slow infusion versus bolus. J Cardiovasc Magn Reson. 2011;13:16.
Gai N, Turkbey EB, Nazarian S, van der Geest RJ, Liu CY, Lima JA, et al. T1 mapping of the gadolinium-enhanced myocardium: adjustment for factors affecting interpatient comparison. Magn Reson Med. 2011;65:1407–15.
Kawel N, Nacif M, Zavodni A, Jones J, Liu S, Sibley CT, et al. T1 mapping of the myocardium: intra-individual assessment of post-contrast T1 time evolution and extracellular volume fraction at 3T for Gd-DTPA and Gd-BOPTA. J Cardiovasc Magn Reson. 2012;14:26.
Lee JJ, Liu S, Nacif MS, Ugander M, Han J, Kawel N, et al. Myocardial T1 and extracellular volume fraction mapping at 3 tesla. J Cardiovasc Magn Reson. 2011;13:75.
Puntmann VO, D'Cruz D, Smith Z, Pastor A, Choong P, Voigt T, et al. Native myocardial T1 mapping by cardiovascular magnetic resonance imaging in subclinical cardiomyopathy in patients with systemic lupus erythematosus. Circ Cardiovasc Imaging. 2013;6:295–301.
Ugander M, Oki AJ, Hsu LY, Kellman P, Greiser A, Aletras AH, et al. Extracellular volume imaging by magnetic resonance imaging provides insights into overt and sub-clinical myocardial pathology. Eur Heart J. 2012;33:1268–78.
Flett AS, Sado DM, Quarta G, Mirabel M, Pellerin D, Herrey AS, et al. Diffuse myocardial fibrosis in severe aortic stenosis: an equilibrium contrast cardiovascular magnetic resonance study. Eur Heart J Cardiovasc Imaging. 2012;13:819–26.
Sado DM, Flett AS, Banypersad SM, White SK, Maestrini V, Quarta G, et al. Cardiovascular magnetic resonance measurement of myocardial extracellular volume in health and disease. Heart. 2012;98:1436–41.
White SK, Sado DM, Flett AS, Moon JC. Characterising the myocardial interstitial space: the clinical relevance of non-invasive imaging. Heart. 2012;98:773–9.
Arheden H, Saeed M, Higgins CB, Gao DW, Ursell PC, Bremerich J, et al. Reperfused rat myocardium subjected to various durations of ischemia: estimation of the distribution volume of contrast material with echo-planar MR imaging. Radiology. 2000;215:520–8.
Kellman P, Wilson JR, Xue H, Ugander M, Arai AE. Extracellular volume fraction mapping in the myocardium, part 1: evaluation of an automated method. J Cardiovasc Magn Reson. 2012;14:63.
Piechnik SK, Ferreira VM, Lewandowski AJ, Ntusi NA, Banerjee R, Holloway C, et al. Normal variation of magnetic resonance T1 relaxation times in the human population at 1.5T using ShMOLLI. J Cardiovasc Magn Reson. 2013;15:13.
Bull S, White SK, Piechnik SK, Flett AS, Ferreira VM, Loudon M, et al. Human non-contrast T1 values and correlation with histology in diffuse fibrosis. Heart. 2013;99:932–7.
Fontana M, White SK, Banypersad SM, Sado DM, Maestrini V, Flett AS, et al. Comparison of T1 mapping techniques for ECV quantification. Histological validation and reproducibility of ShMOLLI versus multibreath-hold T1 quantification equilibrium contrast CMR. J Cardiovasc Magn Reson. 2012;14:88.
Iles L, Pfluger H, Phrommintikul A, Cherayath J, Aksit P, Gupta SN, et al. Evaluation of diffuse myocardial fibrosis in heart failure with cardiac magnetic resonance contrast-enhanced T1 mapping. J Am Coll Cardiol. 2008;52:1574–80.
Kellman P, Wilson JR, Xue H, Bandettini WP, Shanbhag SM, Druey KM, et al. Extracellular volume fraction mapping in the myocardium, part 2: initial clinical experience. J Cardiovasc Magn Reson. 2012;14:64.
Puntmann VO, Voigt T, Chen Z, Mayr M, Karim R, Rhode K, et al. Native T1 mapping in differentiation of normal myocardium from diffuse disease in hypertrophic and dilated cardiomyopathy. JACC Cardiovasc Imaging. 2013;6:475–84.
Liu S, Han J, Nacif MS, Jones J, Kawel N, Kellman P, et al. Diffuse myocardial fibrosis evaluation using cardiac magnetic resonance T1 mapping: sample size considerations for clinical trials. J Cardiovasc Magn Reson. 2012;14:90.
Pennell DJ. T2* magnetic resonance: iron and gold. JACC Cardiovasc Imaging. 2008;1:579–81.
Anderson LJ, Holden S, Davis B, Prescott E, Charrier CC, Bunce NH, et al. Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload. Eur Heart J. 2001;22:2171–9.
Wood JC, Ghugre N. Magnetic resonance imaging assessment of excess iron in thalassemia, sickle cell disease and other iron overload diseases. Hemoglobin. 2008;32:85–96.
Pennell DJ, Udelson JE, Arai AE, Bozkurt B, Cohen AR, Galanello R, et al. Cardiovascular function and treatment in beta-thalassemia major: a consensus statement from the American Heart Association. Circulation. 2013;128:281–308.
Kirk P, Smith GC, Roughton M, He T, Pennell DJ. Myocardial T2* is not affected by ageing, myocardial fibrosis, or impaired left ventricular function. J Magn Reson Imaging. 2010;32:1095–8.
Pepe A, Positano V, Santarelli MF, Sorrentino F, Cracolici E, De Marchi D, et al. Multislice multiecho T2* cardiovascular magnetic resonance for detection of the heterogeneous distribution of myocardial iron overload. J Magn Reson Imaging. 2006;23:662–8.
Kirk P, Roughton M, Porter JB, Walker JM, Tanner MA, Patel J, et al. Cardiac T2* magnetic resonance for prediction of cardiac complications in thalassemia major. Circulation. 2009;120:1961–8.
Arts T, Prinzen FW, Delhaas T, Milles JR, Rossi AC, Clarysse P. Mapping displacement and deformation of the heart with local sine-wave modeling. IEEE Trans Med Imaging. 2010;29:1114–23.
Cupps BP, Taggar AK, Reynolds LM, Lawton JS, Pasque MK. Regional myocardial contractile function: multiparametric strain mapping. Interact Cardiovasc Thorac Surg. 2010;10:953–7.
Del-Canto I, Lopez-Lereu MP, Monmeneu JV, Croisille P, Clarysse P, Chorro FJ, et al. Characterization of normal regional myocardial function by MRI cardiac tagging. J Magn Reson Imaging. 2015;41:83–92.
el Ibrahim SH. Myocardial tagging by cardiovascular magnetic resonance: evolution of techniques–pulse sequences, analysis algorithms, and applications. J Cardiovasc Magn Reson. 2011;13:36.
Miller CA, Borg A, Clark D, Steadman CD, McCann GP, Clarysse P, et al. Comparison of local sine wave modeling with harmonic phase analysis for the assessment of myocardial strain. J Magn Reson Imaging. 2013;38:320–8.
Bogaert J, Rademakers FE. Regional nonuniformity of normal adult human left ventricle. Am J Physiol Heart Circ Physiol. 2001;280:H610–620.
Jeung MY, Germain P, Croisille P, El Ghannudi S, Roy C, Gangi A. Myocardial tagging with MR imaging: overview of normal and pathologic findings. Radiographics. 2012;32:1381–98.
Piella G, De Craene M, Bijnens BH, Tobon-Gomez C, Huguet M, Avegliano G, et al. Characterizing myocardial deformation in patients with left ventricular hypertrophy of different etiologies using the strain distribution obtained by magnetic resonance imaging. Rev Esp Cardiol. 2010;63:1281–91.
Young AA, Kramer CM, Ferrari VA, Axel L, Reichek N. Three-dimensional left ventricular deformation in hypertrophic cardiomyopathy. Circulation. 1994;90:854–67.
Castillo E, Osman NF, Rosen BD, El-Shehaby I, Pan L, Jerosch-Herold M, et al. Quantitative assessment of regional myocardial function with MR-tagging in a multi-center study: interobserver and intraobserver agreement of fast strain analysis with Harmonic Phase (HARP) MRI. J Cardiovasc Magn Reson. 2005;7:783–91.
Moore CC, Lugo-Olivieri CH, McVeigh ER, Zerhouni EA. Three-dimensional systolic strain patterns in the normal human left ventricle: characterization with tagged MR imaging. Radiology. 2000;214:453–66.
Augustine D, Lewandowski AJ, Lazdam M, Rai A, Francis J, Myerson S, et al. Global and regional left ventricular myocardial deformation measures by magnetic resonance feature tracking in healthy volunteers: comparison with tagging and relevance of gender. J Cardiovasc Magn Reson. 2013;15:8.
Oxenham HC, Young AA, Cowan BR, Gentles TL, Occleshaw CJ, Fonseca CG, et al. Age-related changes in myocardial relaxation using three-dimensional tagged magnetic resonance imaging. J Cardiovasc Magn Reson. 2003;5:421–30.
Lawton JS, Cupps BP, Knutsen AK, Ma N, Brady BD, Reynolds LM, et al. Magnetic resonance imaging detects significant sex differences in human myocardial strain. Biomed Eng Oxnline. 2011;10:76.
Shehata ML, Cheng S, Osman NF, Bluemke DA, Lima JA. Myocardial tissue tagging with cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2009;11:55.
Venkatesh BA, Donekal S, Yoneyama K, Wu C, Fernandes VR, Rosen BD, et al: Regional myocardial functional patterns: Quantitative tagged magnetic resonance imaging in an adult population free of cardiovascular risk factors: The multi-ethnic study of atherosclerosis (MESA). J Magn Reson Imaging 2014 [Epub ahead of print].
Kuijer JP, Marcus JT, Gotte MJ, van Rossum AC, Heethaar RM. Three-dimensional myocardial strains at end-systole and during diastole in the left ventricle of normal humans. J Cardiovasc Magn Reson. 2002;4:341–51.
Acknowledgements
This work was funded in part by the National Institutes of Health Intramural research program.
Author information
Authors and Affiliations
Corresponding author
Additional information
Competing interests
The authors declare they have no competing interests.
Authors’ contributions
NKB: manuscript design, manuscript drafting, manuscript revision. AM: manuscript drafting. EVB: manuscript drafting. JVC: manuscript drafting. ET: manuscript drafting. RW: manuscript drafting. SP: manuscript drafting. MT: manuscript drafting. JE: manuscript design, manuscript drafting, data analysis. DAB: manuscript design, manuscript drafting, manuscript revision. All authors read and approved the final manuscript.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.
The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.
To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.
The Creative Commons Public Domain Dedication waiver (https://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Kawel-Boehm, N., Maceira, A., Valsangiacomo-Buechel, E.R. et al. Normal values for cardiovascular magnetic resonance in adults and children. J Cardiovasc Magn Reson 17, 29 (2015). https://doi.org/10.1186/s12968-015-0111-7
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12968-015-0111-7
Keywords
- Normal values
- Reference values
- Cardiovascular magnetic resonance
Comments
View archived comments (1)