Improved high-resolution pediatric vascular cardiovascular magnetic resonance with gadofosveset-enhanced 3D respiratory navigated, inversion recovery prepared gradient echo readout imaging compared to 3D balanced steady-state free precession readout imaging
Journal of Cardiovascular Magnetic Resonance volume 18, Article number: 74 (2017)
Abstract
Background
Improved delineation of vascular structures is a common indication for cardiovascular magnetic resonance (CMR) in children and requires high spatial resolution. Currently, pre-contrast 3D, respiratory navigated, T2-prepared, fat saturated imaging with a bSSFP readout (3D bSSFP) is commonly used; however, these images can be limited by blood pool inhomogeneity and exaggeration of metal artifact. We compared image quality of pediatric vasculature obtained using standard 3D bSSFP to 3D, respiratory navigated, inversion recovery prepared imaging with a gradient echo readout (3D IR GRE) performed after administration of gadofosveset trisodium (GT), a blood pool contrast agent.
Methods
For both sequences, VCG triggering was used with acquisition during a quiescent period of the cardiac cycle. 3D bSSFP imaging was performed pre-contrast, and 3D IR GRE imaging was performed 5 min after GT administration. We devised a vascular imaging quality score (VIQS) with subscores for coronary arteries, pulmonary arteries and veins, blood pool homogeneity, and metal artifact. Scoring was performed on axial reconstructions of isotropic datasets by two independent readers and differences were adjudicated. Signal- and contrast-to-noise (SNR and CNR) calculations were performed on each dataset.
Results
Thirty-five patients had both 3D bSSFP and 3D IR GRE imaging performed. 3D IR GRE imaging showed improved overall vascular imaging compared to 3D bSSFP when comparing all-patient VIQS scores (n = 35, median 14 (IQR 11–15), vs 6 (4–10), p < 0.0001), and when analyzing the subset of patients with intrathoracic metal (n = 17, 16 (14–17) vs. 5 (2–9), p < 0.0001). 3D IR GRE showed significantly improved VIQS subscores for imaging the RCA, pulmonary arteries, pulmonary veins, and blood pool homogeneity. In addition, 3D IR GRE imaging showed reduced variability in both all-patient and metal VIQS scores compared to 3D bSSFP (p < 0.05). SNR and CNR were higher with 3D IR GRE in the left ventricle and left atrium, but not the pulmonary arteries.
Conclusions
Respiratory navigated 3D IR GRE imaging after GT administration provides improved vascular CMR in pediatric patients compared to pre-contrast 3D bSSFP imaging, as well as improved imaging in patients with intrathoracic metal. It is an excellent alternative in this challenging patient population when high spatial resolution vascular imaging is needed.
Background
The use of cardiovascular magnetic resonance (CMR) in the pediatric population continues to increase, aided by advancements in techniques that accommodate higher heart rates and shorter scanning time. CMR spatial resolution has improved and now approaches that of computed tomography (CT). Improved delineation of vascular structures, including coronary arteries and pulmonary veins, is a common indication for CMR imaging in children. However, this can be challenging given the need for high spatial resolution and the small size of the structures to be imaged.
Currently, pre-contrast 3D, respiratory navigated, T2 prepared, fat saturated imaging with a balanced steady-state free precession readout (3D bSSFP) is commonly used for imaging of pediatric vascular structures [1, 2]. Images generated by the 3D bSSFP technique, however, often demonstrate blood pool inhomogeneity and exaggerate metal artifacts, which is especially challenging in the congenital heart disease population. Imaging large fields of view using the 3D bSSFP technique also requires prolonged scan times due to navigator inefficiency [3]. The development and FDA approval of a protein-bound MR contrast agent, gadofosveset trisodium (GT), has allowed fundamental changes in the high spatial resolution, 3D pulse sequences. Protein binding allows for a longer duration of contrast effects, so that imaging during an equilibrium phase achieves minimal to no change in the degree of T1 shortening over the scan duration. A 3D, respiratory-navigated, inversion recovery prepared sequence with a gradient echo readout (3D IR GRE) leverages this prolonged equilibrium phase for vascular imaging.
Previous studies have compared MR angiography (MRA) with gadofosveset trisodium to traditional gadobenate dimeglumine contrast for coronary arteries in adults at 3.0 T [4]; coronary artery imaging with gadofosveset trisodium at 3.0 T compared to pre-contrast 3D SSFP imaging at 1.5 T [1]; and for other angiography of carotid [5], renal [6, 7], and iliofemoral arteries [6]. The use of gadofosveset trisodium has shown advantages in each of these situations. Some adult studies of gadofosveset vs. gadobenate show no advantage of gadofosveset, as long as the imaging is performed quickly after contrast administration [8]. Studies in adults with congenital heart disease have shown advantages of the use of gadofoseveset [9]. And, though a recent study shows that in adults 3D inversion recovery bSSFP shows improved image quality as compared to 3D IR GRE, our experience at our institution has shown improved imaging with 3D IR GRE [10]. Studies in the pediatric population are limited, showing improved pediatric vascular imaging with gadofosveset and combined respiratory and cardiac triggering compared to without triggering [3], and improved pediatric vascular imaging with gadofosveset and steady-state MRA compared to first-pass MRA [11]. We compared image quality of pediatric vasculature obtained using the standard 3D bSSFP technique to that obtained with 3D IR GRE imaging performed after administration of GT, including patients with repaired congenital heart disease.
Methods
Study population
The current study represents a retrospective review of all pediatric patients at our institution who underwent CMR using GT with both pre-contrast 3D bSSFP and post-contrast 3D IR GRE imaging during the same examination. Our institution began using GT for clinical CMR evaluations in August 2013, and the present study included patients from August, 2013 to June, 2014. Patients were excluded if either the 3D bSSFP or 3D IR GRE sequence was not completed or if there was a patient related factor that degraded one or both imaging sequences. The study was approved by our local Institutional Review Board.
Image acquisition
All studies were performed on a 1.5 T Siemens Aera magnet (Siemens Healthcare, Erlangen, Germany). For both sequences, vectorcardiographic triggering and respiratory navigation was used. For both 3D bSSFP and 3D IR GRE, a high temporal resolution image was used to select the acquisition period during a quiescent period of the cardiac cycle, field of view (FOV) and phase FOV were adjusted to avoid field wrap, voxel sizes were typically 1.2-1.6 mm, slice thickness was within 0.1 mm of the in-plane resolution to achieve isotropic voxel sizes, and shot duration and timing were similar between the two sequences. 3D bSSFP parameters included TE/TR = 1.2/2.8, flip angle = 90°; 3D IR GRE parameters included TE/TR = 1.3/3.3, flip angle = 18°; IT = 260. Both the 3D bSSFP and the 3D IR GRE sequences used had been optimized for use by the magnet manufacturer. For both sequences, respiratory navigation window was set to achieve 40–60 % image acquisition, and was roughly a 3 mm window for infants, 5 mm for smaller children, and 7 mm for adolescents and young adults. 3D bSSFP imaging was performed pre-contrast; 3D IR GRE imaging was performed 5 min after administration of 0.12 mL/kg (0.03 mmol/kg) GT (max 10 mL). Post-processing was performed using Siemens Argus software version VD 13.
Image analysis
To evaluate vascular imaging for both 3D bSSFP and 3D IR GRE sequences, a Vascular Imaging Quality Score (VIQS) system was developed (Table 1), which focused on key vascular structures for pediatric CMR, and common issues with image quality. For each patient, right coronary artery (RCA), left coronary artery (LCA), pulmonary arteries, pulmonary veins, and blood pool homogeneity were evaluated with a subscore from 0 (worst) to 3 (best), for a maximal “all-patient VIQS score” of 15. For patients with intrathoracic metal (including those with sternal wires), a sixth subscore indicating the amount of artifact caused by the metal was added, yielding a maximal “metal VIQS score” of 18. Scoring was performed on axial reconstructions of the images generated by each sequence irrespective of the native imaging plane of acquisition. Image scoring was performed on our institution’s PACS system on high resolution Barco monitors to ensure optimal visualization for fine details. The desire to view all images on the highest quality format did necessitate unblinding of the study ID’s for each patient to the reader. To minimize bias, each study was scored for each VIQS subscore by two independent readers (AT and TS) at different sittings separated by at least 24 h for 3D bSSFP and 3D IR GRE for each patient. Differences in subscores were adjudicated by a third experienced reader (WJP). Expected metal object diameter was determined by using other sequences in the study, as well as the pre-CMR chest radiograph.
For quantitative analysis, signal- and contrast-to-noise (SNR and CNR) calculations were performed using Syngo MultiModality Workplace (Syngo MMWP) version VE40A (Siemens Healthcare, Erlangen, Germany). Regions of interest were drawn on both the axial 3D bSSFP and 3D IR GRE images in the left ventricular blood pool and myocardium, left atrial blood pool and posterior wall, and proximal right pulmonary artery and vessel wall between the right pulmonary artery and the ascending aorta.
Statistical analysis
Descriptive statistics were calculated for all variables of interest and included means and standard deviations, and medians and interquartile ranges. Normality of continuous variables was assessed using histograms, normal probability plots, and the Anderson-Darling test for normality. VIQS scores for each subscore were compared using the test of symmetry. Image study duration was compared using paired t-tests. Overall VIQS scores, changes in image quality based on age, and SNR and CNR were compared using Wilcoxon signed-rank testing. Non-metal scores stratified by metal artifact scores and VIQS sores based on age were compared using Kruskal-Wallis testing. Score variability within sequences was compared using Brown-Forsythe testing, and inter-rater agreement was assessed using the weighted kappa statistic. Statistical analyses were performed using SAS 9.4 (Cary, NC) and statistical significance was assessed at the 0.05 level.
Results
Patient characteristics
A total of 35 patients completed both 3D bSSFP and 3D IR GRE imaging sequences as part of the same clinical CMR study during the study period, and comprise the all-patient cohort. Of the 35 patients, 17 had intrathoracic metal, and comprise the metal subset cohort. Three other patients were excluded from the study; one for failed ECG gating; one for patient movement, and one for contrast extravasation. Of the 35 patients, 9 underwent the study with general anesthesia and were mechanically ventilated for the CMR, 9 with conscious sedation who were free breathing, and 17 without sedation; the type of sedation used was based on clinical indications. The patients ranged in age from 0.2 to 20.2 years (median 10.7 years, interquartile range (IQR) 5.4–16.1 years). Patient median weight was 46.0 kg (IQR 18.7–63.2 kg). Patient median BSA was 1.43 m2 (IQR 0.77–1.65 m2). Patients who underwent the studies with general anesthesia had a median age of 4.6 years (IQR 3.1–7.9 years); those with sedation had median age 6.3 years (IQR 5.3–8.4 years), and those who had studies without sedation had median age 16.1 years (IQR 14.8–17.4 years). They represented a broad range of underlying clinical diagnoses, including 11 patients with aortic arch disease including bicuspid aortic valve, aortic dilation/aneurysm, and coarctation; 8 patients with concerns for coronary anomalies; 8 patients with single ventricle anatomy who had undergone Fontan palliation; 5 with repaired tetralogy of Fallot; 3 with inflammatory vasculopathies including Takayasu arteritis or Kawasaki disease; 2 with d-transposition of the great arteries after an arterial switch operation; and one patient with a recurrent left atrial appendage aneurysm (some patients had multiple clinical indications).
Sequence comparisons
3D IR GRE imaging showed improved overall vascular imaging compared to 3D bSSFP for all patients when comparing all-patient VIQS scores (median 14 (IQR 11–15), vs 6 (4–10), p < 0.0001, Fig. 1). When analyzing patients with intrathoracic metal, 3D IR GRE again showed improved VIQS (16 (14–17) vs. 5 (2–9), p < 0.0001, Fig. 1). 3D IR GRE showed significantly improved VIQS subscores for imaging the RCA, pulmonary arteries, pulmonary veins, and blood pool homogeneity (Fig. 2 and Table 2). Of the 140 total vascular structures evaluated, only 6 had a lower VIQS score on any of structures on the 3D IR GRE sequence compared to the 3D bSSFP sequence (2 RCA, 4 LCA), while 64 structures (46 %) showed improvement from a 0 or 1 score in the 3D bSSFP to 2 or 3 score on 3D IR GRE (Table 3); this improvement was seen for at least a single vascular structure in 34 of 35 patients. To assess for effects of metal artifact on other scoring, we stratified the non-metal VIQS scores by metal artifact score in the metal cohort, and there was no significant change in the non-metal scores based on the metal score. In addition, 3D IR GRE imaging showed reduced variability in both all-patient and metal VIQS scores compared to 3D bSSFP (both p < 0.05). Typical example images are shown in Fig. 3. SNR and CNR were significantly higher for 3D IR GRE compared to 3D bSSFP in the left ventricle and left atrium (p < 0.0001), but not the pulmonary arteries (Fig. 4).
3D IR GRE studies had a longer average duration (7:57 ± 4:06 min vs. 6:02 ± 2:26, p < 0.01). There was no correlation between heart rate and VIQS for either sequence. Inter-rater agreement assessed by the weighted kappa statistic was considered good on all VIQS subscores except metal artifact (Table 2).
To assess whether the imaging differences were related to age, patients were stratified based on being older or younger than 10 years old. Younger patients had significantly higher 3D IR GRE VIQS scores than 3D bSSFP scores for all but the LCA, whereas older patients had higher VIQS scores for all but RCA, pulmonary arteries, and metal artifacts (Table 4).
Discussion
We present a method of vascular CMR, post-GT 3D IR GRE, which we demonstrate to be superior to the standard pre-contrast 3D bSSFP technique in pediatric patients. Since vascular imaging in children remains challenging due to small structure sizes and increased heart rates, uncovering methods of improved vascular imaging is paramount for our field. In addition, delineation of anatomy and size of vascular structures in children is a common indication for CMR, therefore achieving improved vascular imaging is clinically relevant.
We found improved overall imaging with 3D IR GRE in patients with and without metal artifact compared to 3D bSSFP, though the individual metal artifact VIQS subscore was not significantly improved with 3D IR GRE. This finding may be due to our metal artifact VIQS scores having higher inter-rater variability. In some cases, the imaging quality was so poor, particularly for 3D SSFP imaging with intrathoracic metal, that it was difficult to discern the specific causes of image distortion (e.g. whether the poor imaging was due to metal artifact or other reasons). However, given that many patients who are referred for CMR are post-surgical patients with intrathoracic metal, the improved all-patient VIQS with the 3D IR GRE technique was reassuring, even if the differences in the metal artifact subscore were not statistically significant.
We did have a high rate of VIQS scores of 0 for pulmonary veins on the 3D bSSFP imaging. The VIQS scoring system we developed requires the ostium of the vein to be seen clearly entering the left atrium. Given the high rate of signal inhomogeneity/inflow artifacts in the pulmonary veins in 3D bSSFP imaging, this resulted in a number of 0 VIQS scores for pulmonary veins. 3D IR GRE reduces this artifact, therefore improving pulmonary vein VIQS scores (Fig. 3e-f).
There are some limitations to the 3D IR GRE technique. The 3D IR GRE technique did have longer acquisitions on average compared to the 3D bSSFP technique; however, we believe that the improved vascular imaging justifies the approximately 2 min longer sequence duration for this cohort. In addition, the inversion pulse precludes the performance of dual phase imaging due to insufficient time for two inversion pulses in a single RR interval. Though two 3D IR GRE sequences, one in systole and one in diastole, could be run in tandem, we did not investigate this possibility. We also did not investigate the use of different inversion times, as the sequence was used as prescribed by the manufacturer.
Gadofosveset trisodium shows a similar safety profile to standard extracellular gadolinium contrast agents [12], and can be used similarly to extracellular agents for first-pass arterial imaging, though GT offers the advantage of higher relaxivity [13]. Gadofosveset trisodium’s primary advantage is its increased duration of intravascular residence, offering the opportunity to obtain equilibrium images (up to 45–60 min after injection in adults) [13]. GT’s higher relaxivity also allows a smaller total dose of gadolinium to be given (standard dose is 0.03 mmol/kg), and even lower doses of GT have been reported as well [14]. No cases of nephrogenic systemic fibrosis have been reported with GT to date, but given renal clearance, caution should be used in patients with renal impairment, similar to extracellular gadolinium contrast agents [12, 13].
Gadofosveset trisodium’s protein binding and long equilibrium state, however, do confer some limitations to its use. Fibrosis imaging using late gadolinium enhancement (LGE) techniques have not been fully established with GT. While the principals behind LGE imaging hold regardless of the specific gadolinium agent utilized, inversion times and optimal delay between administration and LGE imaging have not been well studied with GT [15]. Additionally, the emerging field of diffuse fibrosis imaging with T1 mapping and extra-cellular volume (ECV) calculation have fundamental assumptions which are violated when using blood pool contrast agents such as GT. Finally, GT is currently unavailable for sale. These limitations notwithstanding, our center had moved toward using GT-enhanced 3D IR GRE imaging for many of our pediatric CMR cases due to its improved vascular imaging in our population, and will do so again when GT is available.
Limitations
This study does have limitations. We did not randomize the patients to be included in the study. Patient conditions may have been slightly different for the pre-contrast 3D bSSFP and the post-contrast 3D IR GRE, and due to the nature of the sequences, we were unable to alter the fact that the 3D bSSFP had to be run first. Finally, though we did rate the 3D bSSFP and 3D IR GRE datasets in different sittings, there may have been some bias introduced by evaluation of the 3D IR GRE images after the 3D bSSFP.
Conclusion
In this study, we demonstrate improved imaging of pediatric vascular structures using the post-GT 3D IR GRE technique compared to the pre-contrast 3D bSSFP technique. We recommend its use for many common indications for pediatric CMR. The kinetics of GT do preclude its use for some applications, such as ECV calculation with T1 mapping, but if high quality cardiac and vascular anatomy is the primary clinical question for the study, GT-enhanced 3D IR GRE imaging should be considered.
Abbreviations
- 3D bSSFP:
-
Pre-contrast 3D, respiratory navigated, T2 prepared, fat saturated imaging with a balanced steady-state free precession readout
- 3D IR GRE:
-
3D, Respiratory navigated, inversion recovery prepared imaging with a gradient echo readout
- CMR:
-
Cardiovascular magnetic resonance
- CT:
-
Computed tomography
- FOV:
-
Field of view
- GT:
-
Gadofosveset trisodium
- IQR:
-
Interquartile range
- LCA:
-
Left coronary artery
- LGE:
-
Late gadolinium enhancement
- MRA:
-
MRI angiography
- MRI:
-
Magnetic resonance imaging
- RCA:
-
Right coronary artery
- VIQS:
-
Vascular imaging quality score
References
Prompona M, Cyran C, Nikolaou K, Bauner K, Reiser M, Huber A. Contrast-enhanced whole-heart MR coronary angiography at 3.0 T using the intravascular contrast agent gadofosveset. Invest Radiol. 2009;44:369–74.
Fratz S, Chung T, Greil GF, Samyn MM, Taylor AM, Valsangiacomo Buechel ER, Yoo SJ, Powell AJ. Guidelines and protocols for cardiovascular magnetic resonance in children and adults with congenital heart disease: SCMR expert consensus group on congenital heart disease. J Cardiovasc Magn Reson. 2013;15:51.
Vasanawala SS, Chan FP, Newman B, Alley MT. Combined respiratory and cardiac triggering improves blood pool contrast-enhanced pediatric cardiovascular MRI. Pediatr Radiol. 2011;41:1536–44.
Raman FS, Nacif MS, Cater G, Gai N, Jones J, Li D, Sibley CT, Liu S, Bluemke DA. 3.0-T whole-heart coronary magnetic resonance angiography: comparison of gadobenate dimeglumine and gadofosveset trisodium. Int J Cardiovasc Imaging. 2013;29:1085–94.
Amarteifio E, Essig M, Bockler D, Attigah N, Schuster L, Demirel S. Comparison of gadofosveset (Vasovist) with gadobenate dimeglumine (Multihance)-enhanced MR angiography for high-grade carotid artery stenosis. J Neuroradiol. 2015;42:236–44.
Iezzi R, Soulez G, Thurnher S, Schneider G, Kirchin MA, Shen N, Pirovano G, Spinazzi A. Contrast-enhanced MRA of the renal and aorto-iliac-femoral arteries: comparison of gadobenate dimeglumine and gadofosveset trisodium. Eur J Radiol. 2011;77:358–68.
Schneider G, Pasowicz M, Vymazal J, Seidl Z, Aschauer M, Konopka M, Bilecen D, Iezzi R, Ballarati C. Gadobenate dimeglumine and gadofosveset trisodium for MR angiography of the renal arteries: Multicenter intraindividual crossover comparison. AJR Am J Roentgenol. 2010;195:476–85.
Camren GP, Wilson GJ, Bamra VR, Nguyen KQ, Hippe DS, Maki JH. A comparison between gadofosveset trisodium and gadobenate dimeglumine for steady state MRA of the thoracic vasculature. Biomed Res Int. 2014;2014:625614.
Makowski MR, Wiethoff AJ, Uribe S, Parish V, Botnar RM, Bell A, Kiesewetter C, Beerbaum P, Jansen CH, Razavi R, et al. Congenital heart disease: cardiovascular MR imaging by using an intravascular blood pool contrast agent. Radiology. 2011;260:680–8.
Febbo JA, Galizia MS, Murphy IG, Popescu A, Bi X, Turin A, Collins J, Markl M, Edelman RR, Carr JC. Congenital heart disease in adults: quantitative and qualitative evaluation of IR FLASH and IR SSFP MRA techniques using a blood pool contrast agent in the steady state and comparison to first pass MRA. Eur J Radiol. 2015;84:1921–9.
Naehle CP, Kaestner M, Muller A, Willinek WW, Gieseke J, Schild HH, Thomas D. First-pass and steady-state MR angiography of thoracic vasculature in children and adolescents. JACC Cardiovasc Imaging. 2010;3:504–13.
Sabach AS, Bruno M, Kim D, Mulholland T, Lee L, Kaura S, Lim RP. Gadofosveset trisodium: abdominal and peripheral vascular applications. AJR Am J Roentgenol. 2013;200:1378–86.
Goyen M. Gadofosveset-enhanced magnetic resonance angiography. Vasc Health Risk Manag. 2008;4:1–9.
Kim CY, Heye T, Bashir MR, Gebhard TA, Merkle EM. Gadofosveset-enhanced magnetic resonance angiography of the thoracic vasculature in the equilibrium phase: feasibility and impact of dose. J Comput Assist Tomogr. 2013;37:732–6.
Thouet T, Schnackenburg B, Kokocinski T, Fleck E, Nagel E, Kelle S. Visualization of chronic myocardial infarction using the intravascular contrast agent MS-325 (gadofosveset) in patients. ScientificWorldJournal. 2012;2012:236401.
Acknowledgements
We acknowledge the Emory + Children’s Pediatric Biostatistics Core for their help with analysis of the data.
Funding
Not applicable.
Availability of data and materials
All data will be available for review in an anonymized way if a request is sent to the corresponding author.
Authors’ contributions
All authors made substantial contributions to conception and design of the study. AT, SH, WJP, and TCS acquired the data. AT, WJP, MSK, and TCS analyzed the data. All authors were involved in drafting and revising the manuscript. All authors read and approved the final manuscript.
Competing interests
The authors declare that they have no competing interests.
Consent for publication
Not applicable for this retrospective review.
Ethics approval and consent to participate
The study was approved by the Emory University School of Medicine and Children’s Healthcare of Atlanta Institutional Review Board. There was a waiver of consent for this retrospective study.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Tandon, A., Hashemi, S., Parks, W.J. et al. Improved high-resolution pediatric vascular cardiovascular magnetic resonance with gadofosveset-enhanced 3D respiratory navigated, inversion recovery prepared gradient echo readout imaging compared to 3D balanced steady-state free precession readout imaging. J Cardiovasc Magn Reson 18, 74 (2017). https://doi.org/10.1186/s12968-016-0296-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12968-016-0296-4